CLASS II Device recall · Reported 9 Mar 2016 · FDA Enforcement Report
Sense Body Coil 1.5T with identification recalled
Philips Electronics North America is recalling Sense Body Coil 1.5T with identification, distributed nationwide in the US. The recall was initiated on 6 Jul 2015 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1024-2016
- FDA event ID
- 73333
- Recalling firm
- Philips Electronics North America, Andover, MA
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 6 Jul 2015
- FDA classified
- 26 Feb 2016
- Reported
- 9 Mar 2016
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Temperature Abuse
- Device type
- Other Medical Devices
Timeline
| 6 Jul 2015 | Recall initiated by company | |
| 26 Feb 2016 | Classified by FDA | +235 days |
| 9 Mar 2016 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Wrong positioning of the coil cables can create unintended resonances via the patient or elevated coil cable temperature due to coupling to the QBC. The combined use of the Sense coil, high SAR level scanning and placement of the cables contrary to instructions for use can lead to patient burns.
Foods that need refrigeration and were stored or shipped too warm can grow harmful bacteria without any change in look or smell.
Product as listed by the FDA
SENSE Body Coil 1.5T with identification; used in the ACS NT, Intera 1.5T and Achieva 1.5T MR system
Where it was distributed
US: Nationwide OUS: Canada, Afghanistan Albania Algeria Andorra Argentina Australia Austria Azerbaijan Bahamas Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Botswana Brazil Burkina Faso Chile China Colombia Columbia Congo, Democratic Costa Rica Croatia Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt Estonia Ethiopia Finland France Georgia Germany Ghana Greece Guadeloupe Hong Kong Hungary Iceland India Indonesia Iran Iraq Ireland Israel Italy Japan Jersey Jordan Kazakhstan Kenya Korea, Republic of Kuwait Kyrgyzstan Latvia Lebanon Libya Lithuania Luxembourg Malaysia Malta Martinique Mauritius Mexico Monaco Mongolia Morocco Mozambique Nepal Netherlands New Caledonia New Zealand Norway Oman Pakistan Palestinian territory, occupied Panama Papua New Guinea Peru Philippines Poland Portugal Puerto Rico Romania Russian Federation Rwanda Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Sudan Sweden Switzerland Syrian Arab Republic Taiwan Tanzania Thailand Tunisia Turkey Uganda Ukraine United Arab Emirates United Kingdom Venezuela Viet Nam Yemen Zimbabwe
Questions about this recall
Why was it recalled?
Wrong positioning of the coil cables can create unintended resonances via the patient or elevated coil cable temperature due to coupling to the QBC. The combined use of the Sense coil, high SAR level scanning and placement of the cables contrary to instructions for use can lead to patient burns.
Which lots are affected?
4522 131 5575x 4522 132 1985x 4522 132 7619x 4598 000 5187x 4598 000 5189x 4598 003 59051
Where was it sold?
US: Nationwide OUS: Canada, Afghanistan Albania Algeria Andorra Argentina Australia Austria Azerbaijan Bahamas Bangladesh Belarus Belgium Bolivia Bosnia and Herzegovina Botswana Brazil Burkina Faso Chile China Colombia Columbia Congo, Democratic Costa Rica Croatia Cyprus Czech Republic Denmark Dominican Republic Ecuador Egypt Estonia Ethiopia Finland France Georgia Germany Ghana Greece Guadeloupe Hong Kong Hungary Iceland India Indonesia Iran Iraq Ireland Israel Italy Japan Jersey Jordan Kazakhstan Kenya Korea, Republic of Kuwait Kyrgyzstan Latvia Lebanon Libya Lithuania Luxembourg Malaysia Malta Martinique Mauritius Mexico Monaco Mongolia Morocco Mozambique Nepal Netherlands New Caledonia New Zealand Norway Oman Pakistan Palestinian territory, occupied Panama Papua New Guinea Peru Philippines Poland Portugal Puerto Rico Romania Russian Federation Rwanda Saudi Arabia Serbia Singapore Slovakia Slovenia South Africa Spain Sri Lanka Sudan Sweden Switzerland Syrian Arab Republic Taiwan Tanzania Thailand Tunisia Turkey Uganda Ukraine United Arab Emirates United Kingdom Venezuela Viet Nam Yemen Zimbabwe
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Philips Electronics North America
All 160| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | IntelliVue MX40 Patient Monitor recalled by Philips Electronics North | Philips Electronics North America | Device Malfunction | Nationwide |
| Device | CLASS II | Philips HeartStart FR3 Defibrillator recalled | Philips Electronics North America | Device Malfunction | Nationwide |
| Device | CLASS II | Allura Xper F010, System recalled by Philips Electronics North | Philips Electronics North America | Other | Nationwide |
| Device | CLASS II | Allura Xper FD2O Biplane, System recalled by Philips Electronics North | Philips Electronics North America | Other | Nationwide |
| Device | CLASS II | AIIuraXperFDlO, System recalled by Philips Electronics North | Philips Electronics North America | Other | Nationwide |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | AMS Convenience Kit: Dsaek Pack - RX recalled | Windstone Medical Packaging | Device Malfunction | Nationwide |