CLASS II ONGOING Device recall · Reported 12 Nov 2025 · FDA Enforcement Report

Sentinel V11 recalled by Spacelabs Healthcare

Spacelabs Healthcare is recalling Sentinel V11, distributed nationwide in the US. The recall was initiated on 16 Sep 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCatalog Number: 92810 UDI-DI code; (01)10841522125751 Software version: 11.6.0
Quantity recalled142

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0416-2026
FDA event ID
97686
Recalling firm
Spacelabs Healthcare, Hertford, N/A
Classification
Class II
Status
Ongoing
Recall initiated
16 Sep 2025
FDA classified
31 Oct 2025
Reported
12 Nov 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

16 Sep 2025Recall initiated by company
31 Oct 2025Classified by FDA+45 days
12 Nov 2025Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Due to two distinct issues: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific navigation conditions, test data intended for one patient may be saved under another patient's record

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Brand Name: Sentinel V11 Product Name: Sentinel V11 Model/Catalog Number: 92810 Software Version: 11.6.0 Product Description: Cardiology Information Management System Component: No The Sentinel Cardiology Information Management System is intended to connect to supported medical devices and analyzers in order to download, store, access and manage cardiovascular information and to manage patient and facility information.

Where it was distributed

Worldwide - US Nationwide distribution in the state of MN and the countries of CA, DE, FR, GB, IT, NC.

Nationwide CaliforniaDelawareMinnesotaNorth Carolina

Questions about this recall

Why was it recalled?

Due to two distinct issues: 1. During patient admission, patient demographic fields may default to those of a previously viewed patient. 2. Systems configured with Resting/Rhythm ECG functionality, under specific navigation conditions, test data intended for one patient may be saved under another patient's record

Which lots are affected?

Catalog Number: 92810 UDI-DI code; (01)10841522125751 Software version: 11.6.0

Where was it sold?

Worldwide - US Nationwide distribution in the state of MN and the countries of CA, DE, FR, GB, IT, NC.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Spacelabs Healthcare

All 47

Other Heart & Vascular Devices recalls

All 6,220