CLASS II ONGOING Device recall · Reported 17 Aug 2022 · FDA Enforcement Report
Service manuals for the CT, PET, or NM table as follows recalled
GE Healthcare is recalling Service manuals for the CT, PET, or NM table as follows: GT1700/GT1700V/VT1700, distributed nationwide in the US. The recall was initiated on 2 Jun 2022 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1511-2022
- FDA event ID
- 90368
- Recalling firm
- GE Healthcare, Waukesha, WI
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 2 Jun 2022
- FDA classified
- 10 Aug 2022
- Reported
- 17 Aug 2022
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 2 Jun 2022 | Recall initiated by company | |
| 10 Aug 2022 | Classified by FDA | +69 days |
| 17 Aug 2022 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place resulting in injury to service personnel.
Product as listed by the FDA
Service manuals for the CT, PET, or NM table as follows: GT1700/GT1700V/VT1700, GT2000/VT2000, GT2000x/VT2000x, High Capacity Table, Lite Table, Kunlun Table, Global PET-CT Table, NP Table, P2005AA Table, LCT Table, CTE TWIN TABLE Assy. All devices are patient tables for PET/CT systems.
Where it was distributed
US Nationwide distribution including PR. There was also government/military distribution. The countries of Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Plurinational State of Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Burkina Faso, Cambodia, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Gabon, Georgia, Germany, Ghana, Greece, Guadeloupe, Guatemala, Guyana, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Kyrgyzstan, Lao PDR, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Maldives, Mali, Martinique, Mauritania, Mexico, Moldova, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic of Serbia, Reunion, Romania, Russia, Rwanda, Saudi Arabia, Senegal, Serbia, Sierra Leone, Singapore, Slovakia, Slovenia, Somalia, South Africa, Spain, Sri Lanka, Sudan, Swaziland, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Togo, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Bolivarian Republic of Venezuela, Viet Nam, Yemen, and Yugoslavia.
Questions about this recall
Why was it recalled?
As a result of unclear guidance in the service manual, the table can drop during a specific service activity if the wrong screws are removed by the service personnel without the appropriate table supports in place resulting in injury to service personnel.
Which lots are affected?
Following are the service manual revision numbers and the GTINs for the associated systems: (1) Service manual identifier 5193754-800, Rev 23 and prior, GTIN: N/A; (2) Service manual identifier 5307450-2EN, Rev 30 and prior, GTIN: Revolution Discovery CT: 00840682102490; Revolution Discovery CT A: 00840682125574; Revolution HD: N/A; Discovery CT750 HD: N/A; Discovery CT: N/A. NOTE: Revolution Discovery CT has two different GTINs to identify different versions of the device. (3) Service manual identifier 5350500-8EN, Rev 28 and prior, GTIN: Optima CT540: 00840682102551; Optima CT540 N: 00840682138963. Note: Optima CT540 has two different GTINs to identify different versions of the device. (4) Service manual identifier 5836560-8EN, Rev 2 and prior, GTIN: 00195278563408. (5) Service manual identifier 5366080-8EN, Rev 39 and prior, GTIN: Optima CT660 Cj2.5 Upgrade: 00840682116510. (6) Service manual identifier 5487410-8EN, Rev 14 and prior, GTIN: N/A. (7) Service manual identifier 5809950-8EN, Rev 3 and prior, GTIN: Revolution Maxima: 00840682146180; Revolution Maxima US: 00840682147392; Revolution Ace: 00840682146159; Optima CT680 Expert: 008406821469372; Revolution Maxima M: N/A. (8) Service manual identifier 5848536-8EN, Rev 1 and prior, GTIN: 00195278420893. (9) Service manual identifier 5796213-2EN, Rev 8 and prior, GTIN: 00840682137553. (10) Service manual identifier 5366638-8EN, Rev 24 and prior, GTIN: Discovery RT: 00840682118699, Discovery CT590RT and CT580: N/A. (11) Service manual identifier 5196837-800, Rev 12 and prior, GTIN: N/A. (12) Service manual identifier 5193574-800, Rev 36 and prior, GTIN: 00840682105286. (13) Service manual identifier 5193574-800, Rev 36 and prior, GTIN: N/A. (14) Service manual identifier 5866660-8EN, Rev 26 and prior, GTIN: 00840682109796. (15) Service manual identifier 5987660-8EN, Rev3 and prior, GTIN: 00840682146173. (16) Service manual identifier 5307450-2EN, Rev 30 and prior, GTIN: Revolution Discovery CT: 00840682102490; Revolution Discovery CT A: 00840682125574; Revolution HD: N/A; Discovery CT750 HD: N/A; Discovery CT: N/A. NOTE: Revolution Discovery CT has two different GTINs to identify different versions of the device. (17) Service manual identifier 5366080-8EN, Rev 39 and prior, GTIN: Optima CT660 Cj2.5 Upgrade: 00840682116510. (18) Service manual identifier 5809950-8EN, Rev 3 and prior, GTIN: Revolution Maxima: 00840682146180; Revolution Maxima US: 00840682147392; Revolution Ace: 00840682146159; Optima CT680 Expert: 008406821469372; Revolution Maxima M: Not applicable. NOTE: Revolution Maxima has two different GTINs to identify different versions of the device. (19) Service manual identifier 5796213-2EN, Rev 8 and prior, GTIN: 00840682137553. (20) Service manual identifier 5196837-800, Rev 12 and prior, GTIN: N/A. (21) Service manual identifier 5661743-2EN, Rev 9 and prior, GTIN: Discovery MI weldment: 00840682143523; Discovery MI non-weldment: 00840682108218; Discovery MI Gen2: 00195278098603. NOTE: Discovery MI has three different GTINs to identify different versions of the device. (22) Service manual identifier 5146686-2EN, Rev 7 and prior, GTIN: Discovery MI-DR non-weldment: 00840682120975; Discovery MI-DR weldment: 00840682146166. NOTE: Discovery MI-DR has two different GTINs to identify different versions of the device. (23) Service manual identifier 5495002-2EN, Rev 11 and prior, GTIN: 00840682125703. (24) Service manual identifier 5428822-2EN, Rev 5 and prior, GTIN: Discovery 610: 00840682103053; Discovery 710: 00840682102995. (25) Service manual identifier 5440658-2EN, Rev 5 and prior, GTIN: Discovery 610: 00840682103053; Discovery 710: 00840682102995; Optima 560: N/A. (26) Service manual identifier 5322777-2EN, Rev 12 and prior, GTIN: Discovery 600 Upgrade: 00840682143981; Discovery 690 Elite: N/A; Optima 560: N/A. (27) Service manual identifier 5266353-2EN, Rev 14 and prior, GTIN: N/A. (28) Service manual identifier 5193574-800, Rev 36 and prior, GTIN: 00840682105286. (29) Service manual identifier 5193574-800, Rev 36 and prior, GTIN: N/A. (30) Service manual identifier 5866660-8EN, Rev 26 and prior, GTIN: 00840682109796. (31) Service manual identifier 5987660-8EN, Rev 3 and prior, GTIN: 00840682146173. (32) Service manual identifier 5307450-2EN, Rev 30 and prior, GTIN: Revolution Discovery CT: 00840682102490; Revolution Discovery CT A: 00840682125574; Revolution HD: N/A. (33) Service manual identifier 5796213-2EN, Rev 8 and prior, GTIN: 00840682137553. (34) Service manual identifier 5366080-8EN, Rev 39 and prior, GTIN: Optima CT660 Cj2.5 Upgrade: 00840682116510. (35) Service manual identifier 5866660-8EN, Rev 26 and prior, GTIN: 00840682109796. (36) Service manual identifier 5987660-8EN, Rev 3 and prior, GTIN: 00840682146173. (37) Service manual identifier 5116374, Rev 21 and prior, GTIN: N/A. (38) Service manual identifier 2326233, Rev 17 and prior, GTIN: N/A. (39) Service manual identifier 5661743-2EN, Rev 9 and prior, GTIN: Discovery MI weldment: 00840682143523; Discovery MI non-weldment: 00840682108218; Discovery MI Gen2: 00195278098603. NOTE: Discovery MI has three different GTINs to identify different versions of the device. (40) Service manual identifier 5146686-2EN, Rev 7 and prior, GTIN: Discovery MI-DR non-weldment: 00840682120975; Discovery MI-DR weldment: 00840682146166. NOTE: Discovery MI-DR has two different GTINs to identify different versions of the device. (41) Service manual identifier 5495002-2EN, Rev 11 and prior, GTIN: 00840682125703. (42) Service manual identifier 5428822-2EN, Rev 5 and prior, GTIN: Discovery 610: 00840682103053; Discovery 710: 00840682102995. (43) Service manual identifier 5440658-2EN, Rev 5 and prior, GTIN: Discovery 610: 00840682103053; Discovery 710: 00840682102995; Optima 560: N/A. (44) Service manual identifier 5322777-2EN, Rev 12 and prior, GTIN: Discovery 600 Upgrade: 00840682143981; Discovery 690 Elite: N/A; Optima 560: N/A. (45) Service manual identifier 5266353-2EN, Rev 14 and prior, GTIN: N/A. (46) Service manual identifier 5366638-8EN, Rev 24 and prior, GTIN: Discovery RT: 00840682118699, Discovery CT590RT; Optima CT580: N/A. (47) Service manual identifier 5196837-800, Rev 12 and prior, GTIN: N/A. (48) Service manual identifier 5478834-1EN, Rev 6 and prior, GTIN: N/A. (49) Service manual identifier 5458022-8EN, Rev 15 and prior, GTIN: N/A (50) Service manual identifier 5487206-8EN, Rev 8 and prior, GTIN: Revolution ACT: 00840682135313; Revolution ACT LMB: 00840682146197, Revolution ACT EL: 00195278568373, Revolution ACTs: 00195278534811; Revolution Star: N/A; Revolution Star ES: N/A; ACT: N/A; ACTs: N/A. (51) Service manual identifier 2244773, Rev 36 and prior, GTIN: N/A. (52) Service manual identifier 5181178-800, Rev 24 and prior, GTIN: N/A. (53) Service manual identifier 5411810-8EN, Rev 10 and prior, GTIN: Optima CT520: 00840682102568. (54) Service manual identifier 5457941-8EN, Rev 16 and prior, GTIN: Optima CT520: 00840682102568; Optima Advance: 00840682147415. (55) Service manual identifier 5836560-8EN, Rev 2 and prior, GTIN: 00195278563408. (56) Service manual identifier 5366080-8EN, Rev 39 and prior, GTIN: Optima CT660 Cj2.5 Upgrade: 00840682116510. (57) Service manual identifier 5487410-8EN, Rev 14 and prior, GTIN: N/A. (58) Service manual identifier 5809950-8EN, Rev 3 and prior, GTIN: Revolution Maxima: 00840682146180; Revolution Maxima US: 00840682147392; Revolution Ace: 00840682146159; Optima CT680 Expert: 008406821469372; Revolution Maxima M: N/A. NOTE: Revolution Maxima has two different GTINs to identify different versions of the device. (59) Service manual identifier 5850006-8EN, Rev 1 and prior, GTIN: Revolution Advance: 00195278385215; Revolution Eagle: 00195278403087. (60) Service manual identifier 2188543, Rev 44 and prior, GTIN: N/A. (61) Service manual identifier 2244773, Rev 36 and prior, GTIN: N/A.
Where was it sold?
US Nationwide distribution including PR. There was also government/military distribution. The countries of Afghanistan, Albania, Algeria, Angola, Argentina, Armenia, Australia, Austria, Azerbaijan, Bahrain, Bangladesh, Belarus, Belgium, Bolivia, Plurinational State of Bolivia, Bosnia and Herzegovina, Botswana, Brazil, Brunei Darussalam, Bulgaria, Burkina Faso, Cambodia, Cameroon, Canada, Chile, China, Colombia, Costa Rica, Cote D'Ivoire, Croatia, Cyprus, Czech Republic, Denmark, Dominican Republic, Ecuador, Egypt, El Salvador, Estonia, Ethiopia, Finland, France, Gabon, Georgia, Germany, Ghana, Greece, Guadeloupe, Guatemala, Guyana, Honduras, Hong Kong, Hungary, Iceland, India, Indonesia, Iran, Iraq, Ireland, Israel, Italy, Jamaica, Japan, Jordan, Kazakhstan, Kenya, Republic of Korea, Kuwait, Kyrgyzstan, Lao PDR, Latvia, Lebanon, Libya, Lithuania, Luxembourg, Macedonia, Malaysia, Maldives, Mali, Martinique, Mauritania, Mexico, Moldova, Morocco, Mozambique, Myanmar, Nepal, Netherlands, New Zealand, Nicaragua, Nigeria, Norway, Oman, Pakistan, Palestine, Panama, Paraguay, Peru, Philippines, Poland, Portugal, Qatar, Republic of Serbia, Reunion, Romania, Russia, Rwanda, Saudi Arabia, Senegal, Serbia, Sierra Leone, Singapore, Slovakia, Slovenia, Somalia, South Africa, Spain, Sri Lanka, Sudan, Swaziland, Sweden, Switzerland, Syria, Taiwan, Tanzania, Thailand, Togo, Tunisia, Turkey, Uganda, Ukraine, United Arab Emirates, United Kingdom, Uruguay, Uzbekistan, Venezuela, Bolivarian Republic of Venezuela, Viet Nam, Yemen, and Yugoslavia.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from GE Healthcare
All 512| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Signa Premier systems recalled by GE Healthcare | GE Healthcare | Other | Nationwide |
| Device | CLASS II | Centricity Universal Viewer Zero Footprint Client recalled | GE Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | GE HealthCare ViewPoint 6, Catalog Number H47601AA; Radiological recalled | GE Healthcare | Device Malfunction | |
| Device | CLASS II | Centricity Pacs-iw (Pacs-iw) recalled by GE Healthcare | GE Healthcare | Other | IL |
| Device | CLASS II | Centricity Pacs-iw with Universal Viewer recalled by GE Healthcare | GE Healthcare | Other | IL |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | AMS Convenience Kit: Dsaek Pack - RX recalled | Windstone Medical Packaging | Device Malfunction | Nationwide |