CLASS II Device recall · Reported 28 Oct 2020 · FDA Enforcement Report

Siemens Dimension Vista Akaline Phosphatase recalled

Siemens Healthcare Diagnostics is recalling Siemens Dimension Vista Akaline Phosphatase (ALPI) Flex reagent cartridge - In vitro, distributed nationwide in the US. The recall was initiated on 25 Aug 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot Number/UDI/Exp. Date: 1) 19247AB UDI: 0084276803903419247AB20090310642444840 Exp.2020-09-03, (2) 19282BB UDI: 0084276803903419282BB20100810642444840 Exp. 2020-10-08, (3) 19330BD UDI:0084276803903419330BD20112510642444840 Exp. 2020-11-25. Expanded Recall: 3/10/21 20337AB, 20311BA, 20275AA, 20218AA, 20203BA, 20170AA, 20125BA, 20078AB.
Quantity recalledUS:5970 units ; OUS: 1446 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0134-2021
FDA event ID
86430
Recalling firm
Siemens Healthcare Diagnostics, Newark, DE
Classification
Class II
Status
Terminated
Recall initiated
25 Aug 2020
FDA classified
20 Oct 2020
Reported
28 Oct 2020
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

25 Aug 2020Recall initiated by company
20 Oct 2020Classified by FDA+56 days
28 Oct 2020Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Low outlier resultsgenerated with Alkaline Phosphatase (ALPI) Flex reagent cartridge.The maximum difference observed with a patient sample was approximately -27% at a concentration of 129 U/L (2.15 ukat/L). The low outlier results are generated from the last five (5) tests of the well

Product as listed by the FDA

Siemens Dimension Vista Akaline Phosphatase (ALPI) Flex reagent cartridge - In vitro diagnostic test for the quantitative measurement of alkaline phosphatase in human serum and plasma Catalog Number: K2115, SMN# 10642444

Where it was distributed

Worldwide distribution - US Nationwide and the countries of Canada, Australia, Austria, Belgium, Bahamas, Czech Republic, Denmark, France, Germany, Italy, Japan, Korea, Netherlands, Norway, New Zealand, Portugal, Slovakia, Slovenia, Spain, Switzerland.

Nationwide

Questions about this recall

Is Siemens Dimension Vista Akaline Phosphatase (ALPI) Flex reagent cartridge - In vitro recalled?

Yes. Siemens Healthcare Diagnostics recalled Siemens Dimension Vista Akaline Phosphatase (ALPI) Flex reagent cartridge - In vitro on 25 Aug 2020. The recall is Class II and its status is Terminated. Recall number Z-0134-2021.

Why was it recalled?

Low outlier resultsgenerated with Alkaline Phosphatase (ALPI) Flex reagent cartridge.The maximum difference observed with a patient sample was approximately -27% at a concentration of 129 U/L (2.15 ukat/L). The low outlier results are generated from the last five (5) tests of the well

Which lots are affected?

Lot Number/UDI/Exp. Date: 1) 19247AB UDI: 0084276803903419247AB20090310642444840 Exp.2020-09-03, (2) 19282BB UDI: 0084276803903419282BB20100810642444840 Exp. 2020-10-08, (3) 19330BD UDI:0084276803903419330BD20112510642444840 Exp. 2020-11-25. Expanded Recall: 3/10/21 20337AB, 20311BA, 20275AA, 20218AA, 20203BA, 20170AA, 20125BA, 20078AB.

Where was it sold?

Worldwide distribution - US Nationwide and the countries of Canada, Australia, Austria, Belgium, Bahamas, Czech Republic, Denmark, France, Germany, Italy, Japan, Korea, Netherlands, Norway, New Zealand, Portugal, Slovakia, Slovenia, Spain, Switzerland.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Siemens Healthcare Diagnostics

All 694

Other Test Kits & Diagnostics recalls

All 2,656