CLASS II ONGOING Device recall · Reported 19 Nov 2025 · FDA Enforcement Report
SimplyGo Mini, Standard Battery Kit recalled by Philips Respironics
Philips Respironics is recalling SimplyGo Mini, Standard Battery Kit, distributed nationwide in the US. The recall was initiated on 7 Oct 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0481-2026
- FDA event ID
- 97736
- Recalling firm
- Philips Respironics, Murrysville, PA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 7 Oct 2025
- FDA classified
- 7 Nov 2025
- Reported
- 19 Nov 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Temperature Abuse
- Device type
- CPAP & Ventilators
Timeline
| 7 Oct 2025 | Recall initiated by company | |
| 7 Nov 2025 | Classified by FDA | +31 days |
| 19 Nov 2025 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Certain serial numbers of the SimplyGo Mini standard battery kit contain a defective chip which can lead the battery to either fail to charge or lose voltage output when in use in the device. If the fault occurs while the device is running on battery power, the SimplyGo Mini will shut down and stop providing oxygen to the user. This may happen at a temperature within the specified operating temperature range of the SimplyGo Mini.
Foods that need refrigeration and were stored or shipped too warm can grow harmful bacteria without any change in look or smell.
Product as listed by the FDA
SimplyGo Mini, Standard Battery Kit; Product code: 1116816; The affected battery kits are sold as accessories for the SimplyGo Mini: SimplyGo Mini Portable Oxygen Concentrator Product code: 1113600
Where it was distributed
US Nationwide distribution in the states of GA, ID, KY, NJ, NY and the country of Canada.
Questions about this recall
Why was it recalled?
Certain serial numbers of the SimplyGo Mini standard battery kit contain a defective chip which can lead the battery to either fail to charge or lose voltage output when in use in the device. If the fault occurs while the device is running on battery power, the SimplyGo Mini will shut down and stop providing oxygen to the user. This may happen at a temperature within the specified operating temperature range of the SimplyGo Mini.
Which lots are affected?
Product Code: 1116816; UDI-DI: 00606959032385; Serial Numbers: 2438V115830, 2438V115960, 2438V116033, 2438V115899, 2438V115832, 2438V115964, 2438V116034, 2438V115837, 2438V115965, 2438V116040, 2438V115840, 2438V115967, 2438V116044, 2438V115842, 2438V115968, 2438V116045, 2438V115844, 2438V115970, 2438V116046, 2438V115878, 2438V115972, 2438V116047, 2438V115890, 2438V115973, 2438V116048, 2438V115894, 2438V115974, 2438V116049, 2438V115895, 2438V115980, 2438V116050, 2438V115896, 2438V115981, 2438V116051, 2438V115897, 2438V115983, 2438V116052, 2438V115919, 2438V115984, 2438V116226, 2438V115937, 2438V115991, 2438V116229, 2438V115939, 2438V115992, 2438V116233, 2438V115942, 2438V115993, 2438V116249, 2438V115948, 2438V115994, 2438V116253, 2438V115957, 2438V115995, 2438V116258, 2438V117762;
Where was it sold?
US Nationwide distribution in the states of GA, ID, KY, NJ, NY and the country of Canada.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Philips Respironics
All 76| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Trilogy Evo recalled by Philips Respironics | Philips Respironics | Device Malfunction | Nationwide |
| Device | CLASS I | Trilogy EV300 recalled by Philips Respironics | Philips Respironics | Device Malfunction | Nationwide |
| Device | CLASS I | Trilogy Evo Universal recalled by Philips Respironics | Philips Respironics | Device Malfunction | Nationwide |
| Device | CLASS I | Trilogy Evo O2 recalled by Philips Respironics | Philips Respironics | Device Malfunction | Nationwide |
| Device | CLASS I | Trilogy Evo recalled by Philips Respironics | Philips Respironics | Device Malfunction | Nationwide |
Other CPAP & Ventilators recalls
All 554| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Inca (Infant Nasal Cannulae Assy) Infant Nasal CPAP, Complete Set recalled | CooperSurgical | Other | Nationwide |
| Device | CLASS I | Inca (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP recalled | CooperSurgical | Other | Nationwide |
| Device | CLASS II | Medline Hudson Rci, Dbd-circuit,Vent,Inspiratory,Heated,Neo recalled | Medline Industries | Device Malfunction | Nationwide |
| Device | CLASS II | Getinge Y Sensor Module, Servo-u, Servo recalled | Maquet Critical Care AB | Device Malfunction | Nationwide |
| Device | CLASS II | Biofire Respiratory Panel 2.1 (RP2.1) recalled by BioFire Diagnostics | BioFire Diagnostics | Other | Nationwide |