CLASS II ONGOING Device recall · Reported 29 Jul 2026 · FDA Enforcement Report
SnapshotGLO Wound Imaging Device recalled by Kent Imaging
Kent Imaging is recalling SnapshotGLO Wound Imaging Device, distributed nationwide in the US. The recall was initiated on 30 Jun 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2775-2026
- FDA event ID
- 99406
- Recalling firm
- Kent Imaging, Calgary, N/A
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 30 Jun 2026
- FDA classified
- 22 Jul 2026
- Reported
- 29 Jul 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- N/A
- Reason category
- Device Malfunction
- Device type
- Imaging Equipment
Timeline
| 30 Jun 2026 | Recall initiated by company | |
| 22 Jul 2026 | Classified by FDA | +22 days |
| 29 Jul 2026 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully discharged into a low-power "shipping mode." When this occurs, an affected device will not resume charging when reconnected to power and becomes temporarily unusable/ unavailable for wound imaging until corrected.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
SnapshotGLO Wound Imaging Device, Model KB100, KB100AA.
Where it was distributed
US Nationwide distribution in the states of PA, IA, FL, TX, GA, NV, CA, UT, IN, DC, KS, IL, TN, AR, AZ.
Nationwide ArkansasArizonaCaliforniaDistrict of ColumbiaFloridaGeorgiaIowaIllinoisIndianaKansasNevadaPennsylvaniaTennesseeTexasUtah
Questions about this recall
Why was it recalled?
The software controlling the wound imaging device's battery charging circuit was not designed to recognize and recover a battery that has fully discharged into a low-power "shipping mode." When this occurs, an affected device will not resume charging when reconnected to power and becomes temporarily unusable/ unavailable for wound imaging until corrected.
Which lots are affected?
UDI-DI: B376KB1000, Software Revisions v4.2.0 and v4.1.3
Where was it sold?
US Nationwide distribution in the states of PA, IA, FL, TX, GA, NV, CA, UT, IN, DC, KS, IL, TN, AR, AZ.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Kent Imaging
All 2| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Snapshot NIR recalled by Kent Imaging | Kent Imaging | Other | Nationwide |
Other Imaging Equipment recalls
All 2,349| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Xact mole marker for mammography recalled by Solstice | Solstice | Other | Nationwide |
| Device | CLASS I | Philips Lumify Diagnostic Ultrasound System recalled | Philips Ultrasound | Other | Nationwide |
| Device | CLASS II | Computed Tomography X-ray System recalled by Shanghai United Imaging | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Computed Tomography X-ray System recalled by Shanghai United Imaging | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Computed Tomography X-ray System recalled by Shanghai United Imaging | Shanghai United Imaging Healthcare | Device Malfunction | Nationwide |