CLASS II Device recall · Reported 8 Mar 2017 · FDA Enforcement Report
Sonnet Mini Battery Pack CableProduct Usage recalled
Med-el Elektromedizinische Gereate is recalling Sonnet Mini Battery Pack CableProduct Usage: The Sonnet Mini Battery Pack Cable is part, distributed nationwide in the US. The recall was initiated on 25 Oct 2016 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1395-2017
- FDA event ID
- 76503
- Recalling firm
- Med-el Elektromedizinische Gereate, Innsbruck, N/A
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 25 Oct 2016
- FDA classified
- 2 Mar 2017
- Reported
- 8 Mar 2017
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Processing Deviation
- Device type
- Hip & Knee Implants
Timeline
| 25 Oct 2016 | Recall initiated by company | |
| 2 Mar 2017 | Classified by FDA | +128 days |
| 8 Mar 2017 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Unit Defect: Firm inspection revealed units did not have the compartment lock firmly welded to the housing.
The USDA issues these when meat or poultry was produced without proper inspection, or when a step such as cooking or pasteurisation did not meet requirements.
Product as listed by the FDA
SONNET Mini Battery Pack CableProduct Usage: The SONNET Mini Battery Pack Cable is part of the SONNET Mini Battery Pack, which is an external battery pack used to power one MED-EL SONNET control unit. It can be worn on the upper body or the hip using different SONNET Mini Battery Pack Cable lengths to connect the SONNET Mini Battery Pack to one processor unit.
Where it was distributed
Worldwide Distribution - US Nationwide in the states of CA, NC and TX; and the countries of Belgium, Belarus, Canada, France, Germany, Hong Kong, Israel, Netherlands, Philippines, Poland, Portugal, Turkey, United Kingdom, and Uruguay.
Questions about this recall
Why was it recalled?
Unit Defect: Firm inspection revealed units did not have the compartment lock firmly welded to the housing.
Which lots are affected?
Part No. 32291, Kit 33372, Kit UDI - Primary ID: 09008737333720 (anthracite 50); Part No. 32292, Kit 33374, Kit UDI - Primary ID: 09008737333744 (ebony 50); Part No. 32293, Kit 33375, Kit UDI - Primary ID: 09008737333751 (black 50); Part No. 32294, Kit 33376, Kit UDI - Primary ID: 09008737333768 (white 50); Part No. 32295, Kit 33377, Kit UDI - Primary ID: 09008737333775 (beige 50) and Part No. 32296, Kit 33378, Kit UDI - Primary ID: 09008737333782 (grey 50).
Where was it sold?
Worldwide Distribution - US Nationwide in the states of CA, NC and TX; and the countries of Belgium, Belarus, Canada, France, Germany, Hong Kong, Israel, Netherlands, Philippines, Poland, Portugal, Turkey, United Kingdom, and Uruguay.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Med-el Elektromedizinische Gereate
All 5| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Mi1250 Synchrony recalled by MED-EL Elektromedizinische | Med-el Elektromedizinische Gereate | Device Malfunction | NC |
| Device | CLASS II | MED EL Cochlear Implant System, PULSARci100 Standard recalled | Med-el Elektromedizinische Gereate | Packaging Defects | Nationwide |
| Device | CLASS II | MED EL Cochlear Implant System, SONATATIi100 recalled | Med-el Elektromedizinische Gereate | Packaging Defects | Nationwide |
| Device | CLASS II | MED EL Cochlear Implant System, Mi1000 MED EL Concerto recalled | Med-el Elektromedizinische Gereate | Packaging Defects | Nationwide |
Other Hip & Knee Implants recalls
All 3,097| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline convenience kits labeled as: Femoral Pack, Medline Kit recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Arvis Surgical Navigation System Gen recalled by Kico Knee Innovation | Kico Knee Innovation | Device Malfunction | 19 states |
| Device | CLASS II | Veritas Standard Reverse Total Shoulder Veritas Standard recalled | restor3d | Device Malfunction | Nationwide |
| Device | CLASS II | Endo-Model, Replacement Set, Tibial Plateau, right/left, L, Uhmwpe recalled | Waldemar Link GmbH & Co. KG (Mfg Site) | Other | IN |
| Device | CLASS II | U2 Total Knee System Femoral offset bushing recalled | United Orthopedic | Device Malfunction | Nationwide |