CLASS II Device recall · Reported 14 Aug 2013 · FDA Enforcement Report
Sonopet Ultrasonic Aspirator Console 5450-851-000(Console 100V) recalled
Stryker Instruments Div. of Stryker is recalling Sonopet Ultrasonic Aspirator Console 5450-851-000(Console 100V), distributed nationwide in the US. The recall was initiated on 24 Jun 2013 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1879-2013
- FDA event ID
- 65486
- Recalling firm
- Stryker Instruments Div. of Stryker, Portage, MI
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 24 Jun 2013
- FDA classified
- 5 Aug 2013
- Reported
- 14 Aug 2013
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Hip & Knee Implants
Timeline
| 24 Jun 2013 | Recall initiated by company | |
| 5 Aug 2013 | Classified by FDA | +42 days |
| 14 Aug 2013 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
A supplied component exhibiting corrosion was built into the power supply boards of the Sonopet Consoles listed above. The issue is limited to 2 lots of the supplied component. The presence of the corrosion on the relay of the power supply board can potentially lead to a conduction defect which could cause the irrigation pump to stop working. The purpose of the irrigation pump is to supply sali
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Sonopet Ultrasonic Aspirator Console 5450-851-000(Console 100V) Product Usage: The Sonopet Console is indicated for use in surgical procedures where fragmentation, emulsification, and aspiration of soft and hard tissue is desirable, including neurosurgery, gastrointestinal and affiliated organ surgery, urological surgery, plastic and reconstructive surgery, general surgery, orthopedic surgery, gynecological surgery, thoracic surgery, laparoscopic surgery, and thoracoscopic surgery.
Where it was distributed
Worldwide - USA Nationwide including the states of CA, CO, FL, HI, IA, IL, KS, KY, LA, MD, MI, MN, MS, NC, NH, NJ, NV, NY, OH, PA, TX, VA, ITALY, and the countries of AUSTRALIA, CANADA, UNITED KINGDOM, NETHERLANDS, INDIA, JAPAN, NEW ZEALAND, SWITZERLAND, HONG KONG, and TAIWAN.
Nationwide CaliforniaColoradoFloridaHawaiiIowaIllinoisKansasKentuckyLouisianaMarylandMichiganMinnesotaMississippiNorth CarolinaNew HampshireNew JerseyNevadaNew YorkOhioPennsylvania
Questions about this recall
Why was it recalled?
A supplied component exhibiting corrosion was built into the power supply boards of the Sonopet Consoles listed above. The issue is limited to 2 lots of the supplied component. The presence of the corrosion on the relay of the power supply board can potentially lead to a conduction defect which could cause the irrigation pump to stop working. The purpose of the irrigation pump is to supply sali
Which lots are affected?
5450-851-000(Console 100V) S/N 1127900134, 1133400114, 1202400224, 1205800094, 1226800284, 1127900144, 1133400124, 1202400234, 1205800104, 1226800294, 1127900154, 1133400134, 1202400244, 1206200574, 1226800304, 1127900164, 1133400144, 1203700594, 1206200584, 1228500194, 1127900174, 1133400154, 1203700604, 1206200594, 1228500204, 1127900184, 1133900014, 1205100514, 1222800034, 1228500214, 1127900194, 1133900024, 1205100524, 1222800044, 01J0018, 1127900204, 1133900034, 1205100534, 1224800314, 02B0052, 1130500054, 1133900044, 1205100544, 1225500084, 02G0091, 1130500064, 1133900054, 1205100554, 1225500094, 05H0386, 1130500074, 1201300384, 1205100564, 1225500104, 1130500084, 1202400154, 1205100574, 1225500114, 1130800374, 1202400164, 1205100584, 1225500124, 1130800384, 1202400174, 1205100594, 1226800234, 1130800394, 1202400184, 1205100604, 1226800244, 1130800404, 1202400194, 1205800064, 1226800254, 1130800414, 1202400204, 1205800074, 1226800264, 1130800424, 1202400214, 1205800084, 1226800274.
Where was it sold?
Worldwide - USA Nationwide including the states of CA, CO, FL, HI, IA, IL, KS, KY, LA, MD, MI, MN, MS, NC, NH, NJ, NV, NY, OH, PA, TX, VA, ITALY, and the countries of AUSTRALIA, CANADA, UNITED KINGDOM, NETHERLANDS, INDIA, JAPAN, NEW ZEALAND, SWITZERLAND, HONG KONG, and TAIWAN.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
2 other products · FDA event 65486| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS II | Sonopet Ultrasonic Aspirator Console 5450-852-000 (Console 230V) recalled | Stryker Instruments Div. of Stryker | Device Malfunction | Nationwide | |
| CLASS II | Sonopet Ultrasonic Aspirator Console (Console 110V recalled | Stryker Instruments Div. of Stryker | Device Malfunction | Nationwide |
More recalls from Stryker Instruments Div. of Stryker
All 161| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS III | Triton Sponge System recalled by Stryker Instruments Div. of | Stryker Instruments Div. of Stryker | Device Malfunction | Nationwide |
| Device | CLASS II | Triton Canister System (finished part numbers FG recalled | Stryker Instruments Div. of Stryker | Device Malfunction | Nationwide |
| Device | CLASS II | SurgiCount+ System Application recalled by Stryker Instruments Div. of | Stryker Instruments Div. of Stryker | Device Malfunction | KS, ME |
| Device | CLASS II | Stryker Zyphr Disposable Cranial Perforator recalled | Stryker Instruments Div. of Stryker | Foreign Material | Nationwide |
| Device | CLASS II | Neptune E-sep Blade Electrode recalled by Stryker Instruments Div. of | Stryker Instruments Div. of Stryker | Other | MI |
Other Hip & Knee Implants recalls
All 3,097| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline convenience kits labeled as: Femoral Pack, Medline Kit recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Arvis Surgical Navigation System Gen recalled by Kico Knee Innovation | Kico Knee Innovation | Device Malfunction | 19 states |
| Device | CLASS II | Veritas Standard Reverse Total Shoulder Veritas Standard recalled | restor3d | Device Malfunction | Nationwide |
| Device | CLASS II | Endo-Model, Replacement Set, Tibial Plateau, right/left, L, Uhmwpe recalled | Waldemar Link GmbH & Co. KG (Mfg Site) | Other | IN |
| Device | CLASS II | U2 Total Knee System Femoral offset bushing recalled | United Orthopedic | Device Malfunction | Nationwide |