CLASS II Device recall · Reported 5 Mar 2014 · FDA Enforcement Report

Specialty Straight Static Inserter Assembly recalled by Stryker Spine

Stryker Spine is recalling Specialty Straight Static Inserter Assembly, distributed nationwide in the US. The recall was initiated on 18 Dec 2013 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesD109140 Catalog # IS3009AVS Lot Serial # 3664, 3759, 3843, 3913, 4079, 4195 and 4242
Quantity recalled21 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1086-2014
FDA event ID
67388
Recalling firm
Stryker Spine, Allendale, NJ
Classification
Class II
Status
Terminated
Recall initiated
18 Dec 2013
FDA classified
24 Feb 2014
Reported
5 Mar 2014
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons
Device type
Spinal Implants

Timeline

18 Dec 2013Recall initiated by company
24 Feb 2014Classified by FDA+68 days
5 Mar 2014Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Stryker has received complaints from customers relating to fracture along the weld that binds the handle to the shaft of the static inserter.

Product as listed by the FDA

Specialty Straight Static Inserter Assembly Product Usage: This instrument is a part of the Navigator product brand . It is used for implant insertion in to the disc space. The AVS navigator Peek Spacer System offers implants that are interbody fusion devices intended for use as an aid in spinal fixation. These hollow implants are offered in a variety of lengths, heights and lordotic angles designed to adapt to different patient anatomies. They have serrations on the superior and inferior surfaces designed for multidirectional fixation and ergonomically shaped anterior edges and flat posterior edges.

Where it was distributed

US Nationwide Distribution

Nationwide

Questions about this recall

Why was it recalled?

Stryker has received complaints from customers relating to fracture along the weld that binds the handle to the shaft of the static inserter.

Which lots are affected?

D109140 Catalog # IS3009AVS Lot Serial # 3664, 3759, 3843, 3913, 4079, 4195 and 4242

Where was it sold?

US Nationwide Distribution

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Stryker Spine

All 29

Other Spinal Implants recalls

All 1,159