CLASS II ONGOING Device recall · Reported 3 Sep 2025 · FDA Enforcement Report

Spectrum IQ Infusion Pump recalled by Baxter Healthcare

Baxter Healthcare is recalling Spectrum IQ Infusion Pump, distributed in Arizona, Florida. The recall was initiated on 14 Jul 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 00085412610900; Serial Numbers: 3815255 3815261 3815265 3815266 3815267 3815271
Quantity recalled6 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2415-2025
FDA event ID
97335
Recalling firm
Baxter Healthcare, Deerfield, IL
Classification
Class II
Status
Ongoing
Recall initiated
14 Jul 2025
FDA classified
22 Aug 2025
Reported
3 Sep 2025
Type
Voluntary: Firm initiated
Customers notified by
Telephone
Reason category
Device Malfunction
Device type
Infusion Pumps

Timeline

14 Jul 2025Recall initiated by company
22 Aug 2025Classified by FDA+39 days
3 Sep 2025Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Spectrum IQ Infusion Pump, Product Code 3570009

Where it was distributed

US distribution to states of: AZ, FL

ArizonaFlorida

Questions about this recall

Why was it recalled?

Baxter has identified that certain pumps have potentially been released with the grease applied to the cam and motor gears that will break down quickly. The breakdown of the grease will lead to the device having insufficient or ineffective grease applied to the cam, which could lead to premature wear of the mechanism assembly, resulting in excessive therapy.

Which lots are affected?

UDI-DI: 00085412610900; Serial Numbers: 3815255 3815261 3815265 3815266 3815267 3815271

Where was it sold?

US distribution to states of: AZ, FL

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.