CLASS II ONGOING Device recall · Reported 10 Jul 2024 · FDA Enforcement Report
Sterile convenience kits recalled by MEDLINE INDUSTRIES, LP
Medline Industries LP - Northfield is recalling Sterile convenience kits: a. Medline Retrograde PK-22025203-LF, Pack #DYNJ47873K, distributed nationwide in the US. The recall was initiated on 14 Mar 2024 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2228-2024
- FDA event ID
- 94239
- Recalling firm
- Medline Industries LP - Northfield, Northfield, IL
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 14 Mar 2024
- FDA classified
- 28 Jun 2024
- Reported
- 10 Jul 2024
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Sterility
- Device type
- Surgical Instruments
Timeline
| 14 Mar 2024 | Recall initiated by company | |
| 28 Jun 2024 | Classified by FDA | +106 days |
| 10 Jul 2024 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Surgical convenience kits containing Pure Pouch component that has weak seals which may result in a breach in sterility of the contents if the seal fails.
A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.
Product as listed by the FDA
Sterile convenience kits: a. Medline Retrograde PK-22025203-LF, Pack #DYNJ47873K, containing a Pure Pouch containing component Sterile Curved Kelly Hemostat, Reorder #66175. b. Medline Cath Lab Pack, Pack #DYNJ56141A, containing a Pure Pouch containing component Medline DYND0400 O.R. Scissors, S/B, 5.5 Sterile. c. Medline Sterile Loop Insertion Tray, Pack DYNJ62843A, containing a Pure Pouch containing Medline Reorder #DYNJ04048, Kelly Forceps/Std Str Sterile. d. Medline Radiology Minor MVH, Pack #DYNJ63555C, containing a Pure Pouch containing component Sterile Curved Kelly Hemostat, Reorder #66175. e. Medline Basic Pack, Pack #DYNJ80649B, containing a Pure Pouch containing component Medline DYND0400 O.R. Scissors, S/B, 5.5 Sterile.
Where it was distributed
US Nationwide distribution in the states of AL, CA, CO, CT, FL, IL, IN, MD, MN, MT, NC, NJ, NV, NY, OH, PA, SC, TN, TX, VA, WA, WI, and WY.
Nationwide AlabamaCaliforniaColoradoConnecticutFloridaIllinoisIndianaMarylandMinnesotaMontanaNorth CarolinaNew JerseyNevadaNew YorkOhioPennsylvaniaSouth CarolinaTennesseeTexasVirginia
Questions about this recall
Why was it recalled?
Surgical convenience kits containing Pure Pouch component that has weak seals which may result in a breach in sterility of the contents if the seal fails.
Which lots are affected?
a. Pack #DYNJ47873K - lot numbers 23HBK868, exp. 8/31/2027, 23HBC082exp. 2/29/2028, 23GBC462, exp. 1/31/2028, and 23DBK792, exp. 11/30/2027, UDI-DI each 10193489466812 and UDI-DI case-40193489466813. b. Pack #DYNJ56141A - lot numbers 23IBK595, exp. 1/31/2025, UDI-DI each-10193489555585 and UDI-DI case-40193489555586. c. Pack DYNJ62843A - lot numbers 24ABP043, exp. 12/31/2024, and 23IBM893, exp. 3/31/2025, UDI-DI each-10195327001766 and UDI-DI case-40195327001224. d. Pack #DYNJ63555C - lot numbers 24ALA613, exp. 3/31/2025; 23KLA748, 1/31/2025; 23KLA017, exp. 1/31/2025; 23JLA853, exp. 1/31/2025; 23JLA232. exp. 1/31/2025; 23ILB006, exp. 1/31/2025; 23ILA201, exp. 12/31/2024; and 23HLA429, exp. 12/31/2024; UDI-DI each-10193489816525 and UDI-DI case-40193489816526. e. Pack #DYNJ80649B - lot number 23JBG046, exp. 5/31/2026, UDI-DI each-10195327269944 and UDI-DI case-40195327269945.
Where was it sold?
US Nationwide distribution in the states of AL, CA, CO, CT, FL, IL, IN, MD, MN, MT, NC, NJ, NV, NY, OH, PA, SC, TN, TX, VA, WA, WI, and WY.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
8 other products · FDA event 94239More recalls from Medline Industries LP - Northfield
All 792| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Medline medical convenience kits labeled as recalled | Medline Industries LP - Northfield | Device Malfunction | 6 states |
| Device | CLASS I | Medline medical convenience kits labeled as recalled | Medline Industries LP - Northfield | Device Malfunction | 6 states |
| Device | CLASS I | Medline ReNewal Reprocessed St. Jude Medical Response recalled | Medline Industries LP - Northfield | Particulate Matter | Nationwide |
| Device | CLASS II | The Medline General Surgery Tray is customized to meet recalled | Medline Industries LP - Northfield | Sterility | Nationwide |
| Device | CLASS II | Medline Kits containing BD SureStep Foley Tray Systems recalled | Medline Industries LP - Northfield | Labeling Errors | FL, IL |
Other Surgical Instruments recalls
All 3,809| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits: IV Start KIT recalled by Medline Industries | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Packaging Defects | Nationwide |
| Device | CLASS II | Medline convenience kits labeled as: Open Heart Pack, Medline Kit recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline convenience kits labeled as: GYN Robotic CDS, Medline Kit recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Linvatec Genesys CrossFT Suture Anchor with Disposable Driver recalled | ConMed | Other | Nationwide |