CLASS III Device recall · Reported 29 Jul 2020 · FDA Enforcement Report

Sterile Disposable Connector recalled by LivaNova USA

LivaNova USA is recalling Sterile Disposable Connector,1, distributed in Kansas, Minnesota, Missouri, Utah, Washington. The recall was initiated on 18 Nov 2019 and the FDA has classified it as Class III.

Check your package

Lot, batch or date codesItem# EC2135S, Lot No. 1923900091
Quantity recalled140 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2642-2020
FDA event ID
85858
Recalling firm
LivaNova USA, Arvada, CO
Classification
Class III
Status
Terminated
Recall initiated
18 Nov 2019
FDA classified
17 Jul 2020
Reported
29 Jul 2020
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Sterility

Timeline

18 Nov 2019Recall initiated by company
17 Jul 2020Classified by FDA+242 days
29 Jul 2020Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Some packages of sterile disposable connectors contained a 3/8x1/4 reducer instead of the specified 1/2x1/2 connector. No significant risk to health is expected as the issue is easily detectable and the different shape would prevent use of the connector.

A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.

Product as listed by the FDA

LivaNova USA Sterile Disposable Connector,1/2x1/2, 20 units per case

Where it was distributed

Distributed to medical facilities in 5 states including: Minnesota, Kansas, Missouri, Utah and Washington.

KansasMinnesotaMissouriUtahWashington

Questions about this recall

Why was it recalled?

Some packages of sterile disposable connectors contained a 3/8x1/4 reducer instead of the specified 1/2x1/2 connector. No significant risk to health is expected as the issue is easily detectable and the different shape would prevent use of the connector.

Which lots are affected?

Item# EC2135S, Lot No. 1923900091

Where was it sold?

Distributed to medical facilities in 5 states including: Minnesota, Kansas, Missouri, Utah and Washington.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class III mean?

Class III recalls are for products unlikely to cause any adverse health consequence, usually labelling or quality issues.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from LivaNova USA

All 47

Other Other Medical Devices recalls

All 13,908