CLASS II Device recall · Reported 24 Aug 2016 · FDA Enforcement Report

Stryker AutoPlex System recalled by Stryker Instruments Div. of

Stryker Instruments Div. of Stryker is recalling Stryker AutoPlex System, distributed nationwide in the US. The recall was initiated on 14 Jun 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel Number(s): 0607-687-000 0605-887-000 0605-687-000 0605-683-000 0607-687-000  Lot numbers: 16021012, 16027022, 16049012, 16061012, 16092012, 16107012, 16022012, 16028012, 16049022, 16063012, 16098012, 16109012, 16025012, 16028022, 16049032, 16063022, 16100012, 16109022, 16025022, 16033012, 16053012, 16070012, 16104012, 16113022, 16026012, 16041012, 16056012, 16070022, 16104022, 16113032, 16026022, 16048012, 16056022, 16074012, 16105012, 16118012, 16027012, 16048022, 16060012, 16088012, 16106012, 16118022 0605-887-000  Lot numbers: 16041012, 16040012, 16050012, 16057012, 16069012, 16078012, 16104012, 16112012, 16124012 0605-687-000  Lot numbers: 16041012, 16047012, 16111012 0605-683-000  Lot numbers: 16071012
Quantity recalled4,841 boxes (9,682 units)

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2536-2016
FDA event ID
74582
Recalling firm
Stryker Instruments Div. of Stryker, Portage, MI
Classification
Class II
Status
Terminated
Recall initiated
14 Jun 2016
FDA classified
15 Aug 2016
Reported
24 Aug 2016
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

14 Jun 2016Recall initiated by company
15 Aug 2016Classified by FDA+62 days
24 Aug 2016Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The Piston Head can become separated from the Delivery Piston, blocking the injection assembly Valve resulting in a cement backflow towards the injector handle. Potential for delay in surgery if additional cement needs to be prepared for the injection procedure.

Product as listed by the FDA

Stryker AutoPlex System. Model Numbers 0607-687-000, 0605-887-000, 0605-687-000, and 0605-683-000. For bone cement. Stryker AutoPlex System kits contain a cement mixer, cement injector, funnel, and an extension tube assembly whereby bone cement powder and liquid monomer are mixed inside a chamber and transferred into a cement injector automatically. The bone cement is then dispensed directly into a fracture site using the cement injector and introducer needle combination.

Where it was distributed

Worldwide distribution. US nationwide, France, Netherlands, Spain, Italy, Canada, Israel, Romania, Switzerland, and United Kingdom.

Nationwide

Questions about this recall

Is Stryker AutoPlex System recalled?

Yes. Stryker Instruments Div. of Stryker recalled Stryker AutoPlex System on 14 Jun 2016. The recall is Class II and its status is Terminated. Recall number Z-2536-2016.

Why was it recalled?

The Piston Head can become separated from the Delivery Piston, blocking the injection assembly Valve resulting in a cement backflow towards the injector handle. Potential for delay in surgery if additional cement needs to be prepared for the injection procedure.

Which lots are affected?

Model Number(s): 0607-687-000 0605-887-000 0605-687-000 0605-683-000 0607-687-000  Lot numbers: 16021012, 16027022, 16049012, 16061012, 16092012, 16107012, 16022012, 16028012, 16049022, 16063012, 16098012, 16109012, 16025012, 16028022, 16049032, 16063022, 16100012, 16109022, 16025022, 16033012, 16053012, 16070012, 16104012, 16113022, 16026012, 16041012, 16056012, 16070022, 16104022, 16113032, 16026022, 16048012, 16056022, 16074012, 16105012, 16118012, 16027012, 16048022, 16060012, 16088012, 16106012, 16118022 0605-887-000  Lot numbers: 16041012, 16040012, 16050012, 16057012, 16069012, 16078012, 16104012, 16112012, 16124012 0605-687-000  Lot numbers: 16041012, 16047012, 16111012 0605-683-000  Lot numbers: 16071012

Where was it sold?

Worldwide distribution. US nationwide, France, Netherlands, Spain, Italy, Canada, Israel, Romania, Switzerland, and United Kingdom.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Stryker Instruments Div. of Stryker

All 161

Other Heart & Vascular Devices recalls

All 6,190