CLASS II ONGOING Device recall · Reported 5 Jun 2024 · FDA Enforcement Report
Synchro .014"guide wire tip recalled by Stryker Neurovascular
Stryker Neurovascular is recalling Synchro .014"guide wire tip, distributed nationwide in the US. The recall was initiated on 23 Apr 2024 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1909-2024
- FDA event ID
- 94437
- Recalling firm
- Stryker Neurovascular, Fremont, CA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 23 Apr 2024
- FDA classified
- 28 May 2024
- Reported
- 5 Jun 2024
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Sterility
- Device type
- Other Medical Devices
Timeline
| 23 Apr 2024 | Recall initiated by company | |
| 28 May 2024 | Classified by FDA | +35 days |
| 5 Jun 2024 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.
A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.
Product as listed by the FDA
Synchro .014"guide wire, 35 cm tip, Catalog: M00313010 Synchro .014 guide wire, 300 cm length, Catalog: M00313310 SYNCHRO .010 WIRE SOFT STRAIGHT 200cm, Catalog: M00316310 SYNCHRO .010 WIRE SOFT STRAIGHT 300cm, Catalog: M00316330
Where it was distributed
Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA, SC, GU, WA, MO, OR, DC, OK, LA, MD, ID, NH, SD, NV, RI, ND, HI, MS, AR, IA, NM, VT, KS and the countries of Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, SAR China, Hungary, India, Ireland, Israel, Italy, Japan, Korea (South), Lithuania, Luxembourg, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Republic of China, Thailand, Turkey, United Arab Emirates, United Kingdom.
Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusetts
Questions about this recall
Why was it recalled?
Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.
Which lots are affected?
Catalog/UDI-DI/Lot: M00313010/07613252186922/123279, 123364, 123404, 123985, 124214, 124216; M00313310/07613252186946/123494; M00316310/07613252187158/123851; M00316330/07613252187165/124156
Where was it sold?
Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA, SC, GU, WA, MO, OR, DC, OK, LA, MD, ID, NH, SD, NV, RI, ND, HI, MS, AR, IA, NM, VT, KS and the countries of Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, SAR China, Hungary, India, Ireland, Israel, Italy, Japan, Korea (South), Lithuania, Luxembourg, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Republic of China, Thailand, Turkey, United Arab Emirates, United Kingdom.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
7 other products · FDA event 94437This product is part of a recall event; the main page for the event is Trevo ProVue recalled by Stryker Neurovascular.
| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS II | Trevo ProVue recalled by Stryker Neurovascular | Stryker Neurovascular | Sterility | Nationwide | |
| CLASS II | Trevo Trak 21 Microcatheter recalled by Stryker Neurovascular | Stryker Neurovascular | Sterility | Nationwide | |
| CLASS II | DAC 044 115cm, Catalog recalled by Stryker Neurovascular | Stryker Neurovascular | Sterility | Nationwide | |
| CLASS II | Trevo NXT ProVue , US recalled by Stryker Neurovascular | Stryker Neurovascular | Sterility | Nationwide | |
| CLASS II | Trevo XP ProVue Retriever recalled by Stryker Neurovascular | Stryker Neurovascular | Sterility | Nationwide | |
| CLASS II | FlowGate2 8F recalled by Stryker Neurovascular | Stryker Neurovascular | Sterility | Nationwide | |
| CLASS II | Trevo Pro 14 Microcatheter recalled by Stryker Neurovascular | Stryker Neurovascular | Sterility | Nationwide |
More recalls from Stryker Neurovascular
All 109| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Inzone Detachment System recalled by Stryker Neurovascular | Stryker Neurovascular | Device Malfunction | Nationwide |
| Device | CLASS II | The Trevo Trak 21 Microcatheter is a single-lumen recalled | Stryker Neurovascular | Packaging Defects | Nationwide |
| Device | CLASS II | The Synchro Neuro Guidewires are a steerable guidewire recalled | Stryker Neurovascular | Other | Nationwide |
| Device | CLASS II | The Synchro Select Guidewires are a steerable guidewire recalled | Stryker Neurovascular | Other | Nationwide |
| Device | CLASS II | The Synchro2 Guidewires are a steerable guidewire family recalled | Stryker Neurovascular | Other | Nationwide |
Other Other Medical Devices recalls
All 13,908| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Plastics Tray kits recalled by Avid Medical | Avid Medical | Foreign Material | Nationwide |
| Device | CLASS II | Heart Prep Tray kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Pack Open Heart Cfh kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Sperm Ab Elisa (96 well plate test) recalled by DRG International | DRG International | Other | Nationwide |
| Device | CLASS III | BeneView Series recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |