CLASS II ONGOING Device recall · Reported 5 Jun 2024 · FDA Enforcement Report

Trevo XP ProVue Retriever recalled by Stryker Neurovascular

Stryker Neurovascular is recalling Trevo XP ProVue Retriever, distributed nationwide in the US. The recall was initiated on 23 Apr 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCatalog/UDI-DI/Lot: 90182/00815742001822/116706; 90183/00815742001839/117514; 90185/00815742001853/118900; 90186/00815742001860/114054; 80051/00815742000511/47307561, 47328877, 47351811, 47398058, 47398074, 48521577, 48521585, 48540280, 48540281, 48540282, 48540283, 48540284, 48556786, 48588490, 48607589, 48676218, 48696629, 48715720, 48736116, 48736122, 48754915, 48754920, 48772012, 48772014, 48772025, 50065499, 50065511, 50065512, 50065513, 50065514, 50065515, 50065518, 50113384, 50113385, 50133176, 50133191, 50193582, 50216273, 50216274, 50216277, 50239676, 50260575, 50260583, 50280979, 50281011, 50304688, 50304712, 50304713, 50304714, 50397877, 50422572, 50422582, 50422589, 50448279, 50519274, 50519275, 50519276, 50519277, 50519285, 50519295, 50577572, 50607374, 50607376, 50607377, 50607378, 50650278, 50650279, 50650288, 50673881, 50673884, 50673908, 50673909, 50673910, 50673911, 50673930, 50694279, 50738912, 50786710, 50850697, 50869887, 50869892, 50869905, 50886487, 50912493, 50912514, 50912515, 50912516, 50912517, 50931490, 50931518, 50931519, 50950402, 50967093, 51320497, 51350503, 51350505, 51489228, 51489235, 51513311, 51513516, 51535421, 51729133; 80052/00815742000528/49021114, 49021115, 49021116, 49021117, 49021118, 49033114, 49033115, 49033116, 49033117, 49052517, 49052525, 49192125, 49192140, 49192141, 49216723, 49216724, 49216737, 53175402; 93067/00815742000672/53320201; 93068/00815742000689/48963313, 49056315, 49155966, 49216725, 49318831, 49318836, 49318837, 49418740, 49418743, 49441346, 50000932, 50000933, 50000934, 50029333, 50029340, 50029361, 50047054, 50065482, 50065485, 50065486, 50065487; M0033PK32021002/07613327126334/48500990, 48624226, 48641523, 48641533, 50156372, 50519272, 50650290, 50912513, 50967091; M0033PK42022002/07613327126358/48995917, 49033123, 49140624, 49236926; M0033PK62523002/07613327126365/49033118, 49236935, 50047053
Quantity recalled461

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1902-2024
FDA event ID
94437
Recalling firm
Stryker Neurovascular, Fremont, CA
Classification
Class II
Status
Ongoing
Recall initiated
23 Apr 2024
FDA classified
28 May 2024
Reported
5 Jun 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Sterility

Timeline

23 Apr 2024Recall initiated by company
28 May 2024Classified by FDA+35 days
5 Jun 2024Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.

A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.

Product as listed by the FDA

Trevo XP ProVue Retriever 4 x 20, Catalog: 90182 Trevo XP ProVue Retriever 3 x 20, Catalog: 90183 Trevo XP ProVue Retriever 4 x 30, Catalog: 90185 Trevo XP ProVue Retriever 6 x 25, Catalog: 90186 Kits: TREVO XP 3X20+PRO14 2-PACK, Catalog: 80051 TREVO XP 4X20+PRO18 2-PACK, Catalog: 80052 TREVO XP 4X30+PRO18 2-PACK, Catalog: 93067 TREVO XP 6X25+XT27 2-PACK, Catalog: 93068 TREVO XP 3X20+PRO14+CAT6 3-PACK - CE, Catalog: M0033PK32021002 TREVO XP 4X20+PRO18+CAT6 3-PACK - CE, Catalog: M0033PK42022002 TREVO XP 6X25+XT27+CAT6 3-PACK - CE, Catalog: M0033PK62523002

Where it was distributed

Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA, SC, GU, WA, MO, OR, DC, OK, LA, MD, ID, NH, SD, NV, RI, ND, HI, MS, AR, IA, NM, VT, KS and the countries of Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, SAR China, Hungary, India, Ireland, Israel, Italy, Japan, Korea (South), Lithuania, Luxembourg, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Republic of China, Thailand, Turkey, United Arab Emirates, United Kingdom.

Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusetts

Questions about this recall

Why was it recalled?

Neurovascular devices did not meet the minimum sample size required for endotoxin testing during manufacturing processes and are out of compliance with the AAMI ST72 standard, so these devices may exceed the acceptable limits of endotoxin levels; high levels of endotoxins may increase the risk of fever, shock and aseptic meningitis.

Which lots are affected?

Catalog/UDI-DI/Lot: 90182/00815742001822/116706; 90183/00815742001839/117514; 90185/00815742001853/118900; 90186/00815742001860/114054; 80051/00815742000511/47307561, 47328877, 47351811, 47398058, 47398074, 48521577, 48521585, 48540280, 48540281, 48540282, 48540283, 48540284, 48556786, 48588490, 48607589, 48676218, 48696629, 48715720, 48736116, 48736122, 48754915, 48754920, 48772012, 48772014, 48772025, 50065499, 50065511, 50065512, 50065513, 50065514, 50065515, 50065518, 50113384, 50113385, 50133176, 50133191, 50193582, 50216273, 50216274, 50216277, 50239676, 50260575, 50260583, 50280979, 50281011, 50304688, 50304712, 50304713, 50304714, 50397877, 50422572, 50422582, 50422589, 50448279, 50519274, 50519275, 50519276, 50519277, 50519285, 50519295, 50577572, 50607374, 50607376, 50607377, 50607378, 50650278, 50650279, 50650288, 50673881, 50673884, 50673908, 50673909, 50673910, 50673911, 50673930, 50694279, 50738912, 50786710, 50850697, 50869887, 50869892, 50869905, 50886487, 50912493, 50912514, 50912515, 50912516, 50912517, 50931490, 50931518, 50931519, 50950402, 50967093, 51320497, 51350503, 51350505, 51489228, 51489235, 51513311, 51513516, 51535421, 51729133; 80052/00815742000528/49021114, 49021115, 49021116, 49021117, 49021118, 49033114, 49033115, 49033116, 49033117, 49052517, 49052525, 49192125, 49192140, 49192141, 49216723, 49216724, 49216737, 53175402; 93067/00815742000672/53320201; 93068/00815742000689/48963313, 49056315, 49155966, 49216725, 49318831, 49318836, 49318837, 49418740, 49418743, 49441346, 50000932, 50000933, 50000934, 50029333, 50029340, 50029361, 50047054, 50065482, 50065485, 50065486, 50065487; M0033PK32021002/07613327126334/48500990, 48624226, 48641523, 48641533, 50156372, 50519272, 50650290, 50912513, 50967091; M0033PK42022002/07613327126358/48995917, 49033123, 49140624, 49236926; M0033PK62523002/07613327126365/49033118, 49236935, 50047053

Where was it sold?

Worldwide - US Nationwide distribution including in the states of MN, PA, IL, AK, NY, CA, MI, WI, TX, AZ, FL, KY, MA, IN, NE, MT, CT, NJ, TN, VA, AL, CO, DE, OH, UT, NC, ME, GA, SC, GU, WA, MO, OR, DC, OK, LA, MD, ID, NH, SD, NV, RI, ND, HI, MS, AR, IA, NM, VT, KS and the countries of Australia, Austria, Belgium, Brazil, Canada, Canary Islands, Chile, China, Czech Republic, Denmark, Finland, France, Germany, Hong Kong, SAR China, Hungary, India, Ireland, Israel, Italy, Japan, Korea (South), Lithuania, Luxembourg, Netherlands, Northern Ireland, Norway, Poland, Portugal, Romania, Russian Federation, Saudi Arabia, Singapore, Slovakia, Spain, Sweden, Switzerland, Taiwan, Republic of China, Thailand, Turkey, United Arab Emirates, United Kingdom.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

7 other products · FDA event 94437

This product is part of a recall event; the main page for the event is Trevo ProVue recalled by Stryker Neurovascular.

More recalls from Stryker Neurovascular

All 109

Other Other Medical Devices recalls

All 13,985