CLASS II Device recall · Reported 4 Sep 2013 · FDA Enforcement Report

Synthes Va-lcp Two Column Volar Distal Radius Plate recalled

Synthes (USA) Products is recalling Synthes Va-lcp Two Column Volar Distal Radius Plate and Torque limiting attachment, 2Nm, distributed nationwide in the US. The recall was initiated on 31 May 2013 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesPart number J8301-F
Quantity recalled119, 891

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2057-2013
FDA event ID
65556
Recalling firm
Synthes (USA) Products, West Chester, PA
Classification
Class II
Status
Terminated
Recall initiated
31 May 2013
FDA classified
26 Aug 2013
Reported
4 Sep 2013
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

31 May 2013Recall initiated by company
26 Aug 2013Classified by FDA+87 days
4 Sep 2013Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The firm is initiating a Field Safety Correction for the labeling of the 2.4mm VA-LCP Two Column Volar Distal Radius Plate and 1.2 Nm Torque Limiting Attachment. The US Technique guide lists 1.2Nm TLA as an option for use with the 2.4mm Variable Angle LCP (VA-LCP) Two-Column (Standard & Narrow) Volar Distal Radius Plate. The potential exists for the VA locking screw to be driven through the 2.4mm

Product as listed by the FDA

Synthes 2.4mm VA-LCP Two Column Volar Distal Radius Plate and Torque limiting attachment, 1.2Nm Synthes VA-LCP Distal Radius System is intended for fixation of complex intra- and extra-articular fractures and osteotomies of the distal radius and other small bones.

Where it was distributed

Nationwide Distribution including the states of AK. AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MD, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI SC, SD, TN, TX, UT, VA, VT, WA WI and WV.

Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusetts

Questions about this recall

Is Synthes Va-lcp Two Column Volar Distal Radius Plate and Torque limiting attachment, 2Nm recalled?

Yes. Synthes (USA) Products recalled Synthes Va-lcp Two Column Volar Distal Radius Plate and Torque limiting attachment, 2Nm on 31 May 2013. The recall is Class II and its status is Terminated. Recall number Z-2057-2013.

Why was it recalled?

The firm is initiating a Field Safety Correction for the labeling of the 2.4mm VA-LCP Two Column Volar Distal Radius Plate and 1.2 Nm Torque Limiting Attachment. The US Technique guide lists 1.2Nm TLA as an option for use with the 2.4mm Variable Angle LCP (VA-LCP) Two-Column (Standard & Narrow) Volar Distal Radius Plate. The potential exists for the VA locking screw to be driven through the 2.4mm

Which lots are affected?

Part number J8301-F

Where was it sold?

Nationwide Distribution including the states of AK. AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MD, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, RI SC, SD, TN, TX, UT, VA, VT, WA WI and WV.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Synthes (USA) Products

All 164

Other Other Medical Devices recalls