CLASS II Device recall · Reported 10 May 2017 · FDA Enforcement Report
Sysmex, Automated Hematology Analyzer XN Series recalled
Sysmex America is recalling Sysmex, Automated Hematology Analyzer XN Series, Sysmex Corporation, 1-5-1, distributed nationwide in the US. The recall was initiated on 23 Mar 2017 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2034-2017
- FDA event ID
- 76828
- Recalling firm
- Sysmex America, Lincolnshire, IL
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 23 Mar 2017
- FDA classified
- 4 May 2017
- Reported
- 10 May 2017
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Test Kits & Diagnostics
Timeline
| 23 Mar 2017 | Recall initiated by company | |
| 4 May 2017 | Classified by FDA | +42 days |
| 10 May 2017 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Reported customer complaints of automated hematology analyzers catching fire due to misalignment of an internal protective sheet that shields electronic components from salty reagents. Build-up of salt deposits on unprotected electronic components may combust due to heat generated by electronics when instrument cycles. Completed CAPA identified the root cause as supplier providing defective component (sensor ready valve) and the misalignment of the electronic protective shield during manufacturing of the device.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Sysmex, Automated Hematology Analyzer XN Series, Sysmex Corporation, 1-5-1 Wakinohama-Kaigandori, Chuo-ku, Kobe 651-0073 Japan
Where it was distributed
Nationwide, Canada
Questions about this recall
Why was it recalled?
Reported customer complaints of automated hematology analyzers catching fire due to misalignment of an internal protective sheet that shields electronic components from salty reagents. Build-up of salt deposits on unprotected electronic components may combust due to heat generated by electronics when instrument cycles. Completed CAPA identified the root cause as supplier providing defective component (sensor ready valve) and the misalignment of the electronic protective shield during manufacturing of the device.
Which lots are affected?
**Model: XN-10, Product Number: AP795756, Serial Numbers: 11001 29656, 29668-29786, 29788-29798, 29800-29801, 29803, 29804, 29807, 29812, 29816-29867 **Model: XN-11, Product Number: CS297931, Serial Numbers: A1001-A1124, A1127, A1130, A1135-A1140, A1142-A1144, A1146-A1149, A1151, A1153-A1159, A1161, A1163, A1171
Where was it sold?
Nationwide, Canada
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Sysmex America
All 14| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | TS-10/TS-10H Tube Sorter, automatic sorting system recalled | Sysmex America | Device Malfunction | Nationwide |
| Device | CLASS II | Sysmex PS-10 Sample Preparation System recalled by Sysmex America | Sysmex America | Device Malfunction | Nationwide |
| Device | CLASS II | Sysmex PS-10 Sample Preparation System recalled by Sysmex America | Sysmex America | Device Malfunction | Nationwide |
| Device | CLASS II | The CF-70 instrument (product recalled by Sysmex America | Sysmex America | Device Malfunction | Nationwide |
| Device | CLASS II | Sysmex CV-11 Sample Unit, an integrated modular sampler recalled | Sysmex America | Other | FL, IL, VT |
Other Test Kits & Diagnostics recalls
All 2,663| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | DxC 500 AU Clinical Chemistry Analyzers: AU w/out ISE recalled | Beckman Coulter | Device Malfunction | Nationwide |
| Device | CLASS II | DxC 500i Clinical Analyzer Modules: DxC 500 AU Module w/ISE recalled | Beckman Coulter | Device Malfunction | Nationwide |
| Device | CLASS II | McKesson Consult hCG Urine Cassette recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |
| Device | CLASS II | Systems/units containing affected recalled | Beckman Coulter Mishima K.K | Device Malfunction | Nationwide |
| Device | CLASS II | Fisher Scientific Sure-Vue Urine hCG recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |