CLASS II Device recall · Reported 15 Aug 2012 · FDA Enforcement Report
System 1 base 100/120V The Terumo Advanced Perfusion System recalled
Terumo Cardiovascular Systems is recalling System 1 base 100/120V The Terumo Advanced Perfusion System 1 is indicated for use in, distributed nationwide in the US. The recall was initiated on 26 Jul 2012 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2177-2012
- FDA event ID
- 62444
- Recalling firm
- Terumo Cardiovascular Systems, Ann Arbor, MI
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 26 Jul 2012
- FDA classified
- 9 Aug 2012
- Reported
- 15 Aug 2012
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Other Medical Devices
Timeline
| 26 Jul 2012 | Recall initiated by company | |
| 9 Aug 2012 | Classified by FDA | +14 days |
| 15 Aug 2012 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Terumo Cardiovascular Systems sent a Safety Advisory (AA-2010-007-C) on December 2, 2010 (Recall# Z-0882/0883-2011 RES 57416) regarding malfunctions of the Occluder and Air Bubble Detection (ABD) Modules to a confined number of Terumo Advanced Perfusion System 1 users. The investigation of the malfunction found a faulty component on the modules' application board caused the failures. Terumo CVS re
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
System 1 base 100/120V The Terumo Advanced Perfusion System 1 is indicated for use in extracorporeal circulation of blood for arterial perfusion, regional perfusion, and cardiopulmonary bypass procedures only when used by a qualified perfusionist who is experienced in the operation of Sarns or similar equipment.
Where it was distributed
Worldwide distribution: USA (nationwide) and countries of: ARGENTINA, AUSTRALIA, BELGIUM, CANADA, CHILE, COLOMBIA, Costa Rica, Dominican Republic, Egypt, Germany, Guatemala, Honduras, Hong Kong, India, Indonesia, Japan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), and Vietnam.
Stores named: CVS
Questions about this recall
Is System 1 base 100/120V The Terumo Advanced Perfusion System 1 is indicated for use in recalled?
Yes. Terumo Cardiovascular Systems recalled System 1 base 100/120V The Terumo Advanced Perfusion System 1 is indicated for use in on 26 Jul 2012. The recall is Class II and its status is Terminated. Recall number Z-2177-2012.
Why was it recalled?
Terumo Cardiovascular Systems sent a Safety Advisory (AA-2010-007-C) on December 2, 2010 (Recall# Z-0882/0883-2011 RES 57416) regarding malfunctions of the Occluder and Air Bubble Detection (ABD) Modules to a confined number of Terumo Advanced Perfusion System 1 users. The investigation of the malfunction found a faulty component on the modules' application board caused the failures. Terumo CVS re
Which lots are affected?
catalog number: 801763 and serial numbers: 0011-0335, 0337-0610,1001-1034, 1037-1049,1100,1103-1106,1109-1450, and 1452.
Where was it sold?
Worldwide distribution: USA (nationwide) and countries of: ARGENTINA, AUSTRALIA, BELGIUM, CANADA, CHILE, COLOMBIA, Costa Rica, Dominican Republic, Egypt, Germany, Guatemala, Honduras, Hong Kong, India, Indonesia, Japan, Kuwait, Malaysia, Mexico, Pakistan, Philippines, Philippines, Russia, Saudi Arabia, Singapore, South Africa, South Korea, Taiwan, Thailand, Turkey, UNITED ARAB EMIRATES (UAE), and Vietnam.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Terumo Cardiovascular Systems
All 222| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | CDI OneView System BPM Probe recalled by Terumo Cardiovascular Systems | Terumo Cardiovascular Systems | Labeling Errors | Nationwide |
| Device | CLASS II | Terumo CDI OneView Monitoring System Hematocrit / Oxygen recalled | Terumo Cardiovascular Systems | Device Malfunction | 6 states |
| Device | CLASS II | Cardiovascular Procedure Kits that include High Crack Check recalled | Terumo Cardiovascular Systems | Device Malfunction | Nationwide |
| Device | CLASS II | Capiox RX15 Hollow Fiber Oxygenator recalled | Terumo Cardiovascular Systems | Packaging Defects | Nationwide |
| Device | CLASS II | Capiox FX25 Hollow Fiber Oxygenator recalled | Terumo Cardiovascular Systems | Packaging Defects | Nationwide |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | AMS Convenience Kit: Dsaek Pack - RX recalled | Windstone Medical Packaging | Device Malfunction | Nationwide |