CLASS II ONGOING Device recall · Reported 7 Oct 2026 · FDA Enforcement Report

Tandem t:slim mobile application t:slim Mobile Application, iOS recalled

Tandem Diabetes Care is recalling Tandem t:slim mobile application t:slim Mobile Application, iOS, distributed nationwide in the US. The recall was initiated on 16 Jul 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesLot Code: UDI-DI: 00389152470641 Software Version: 2.13 (iOS only)
Quantity recalled177

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-3330-2026
FDA event ID
99589
Recalling firm
Tandem Diabetes Care, San Diego, CA
Classification
Class II
Status
Ongoing
Recall initiated
16 Jul 2026
FDA classified
29 Sep 2026
Reported
7 Oct 2026
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction
Device type
Medical Software

Timeline

16 Jul 2026Recall initiated by company
29 Sep 2026Classified by FDA+75 days
7 Oct 2026Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P) continuous glucose monitoring (CGM) sensors, resulting in the insulin pump being unable to establish a near-field communication (NFC) connection with the CGM and requiring FSL3P users to use their t:slim insulin pumps in manual mode. FSL3P users are unable to access their CGM's glucose values from within the t:slim app and cannot use their pumps in automated control (closed-loop) mode, which could lead to users experiencing hypoglycemia or hyperglycemia.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Brand Name: Tandem t:slim mobile application Product Name: t:slim Mobile Application, iOS Model/Catalog Number: iOS software version 2.13 Software Version: iOS software version 2.13 UDI-DI: 00389152470740 Component: No

Where it was distributed

US Nationwide distribution.

Nationwide

Questions about this recall

Why was it recalled?

A beta (early-release) version of the iOS application, version 2.13 (TestFlight) had a software issue preventing successful pairing of the t:slim application with Abbott FreeStyle Libre 3 Plus (FSL3P) continuous glucose monitoring (CGM) sensors, resulting in the insulin pump being unable to establish a near-field communication (NFC) connection with the CGM and requiring FSL3P users to use their t:slim insulin pumps in manual mode. FSL3P users are unable to access their CGM's glucose values from within the t:slim app and cannot use their pumps in automated control (closed-loop) mode, which could lead to users experiencing hypoglycemia or hyperglycemia.

Which lots are affected?

Lot Code: UDI-DI: 00389152470641 Software Version: 2.13 (iOS only)

Where was it sold?

US Nationwide distribution.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 13 Oct 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Tandem Diabetes Care

All 13

Other Medical Software recalls

All 930