CLASS I ONGOING Device recall · Reported 12 Nov 2025 · FDA Enforcement Report
Tandem Mobi Insulin Pump: Description recalled by Tandem Diabetes Care
Tandem Diabetes Care is recalling Tandem Mobi Insulin Pump: Description, distributed in 52 states. The recall was initiated on 6 Oct 2025 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0427-2026
- FDA event ID
- 97662
- Recalling firm
- Tandem Diabetes Care, San Diego, CA
- Classification
- Class I
- Status
- Ongoing
- Recall initiated
- 6 Oct 2025
- FDA classified
- 5 Nov 2025
- Reported
- 12 Nov 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Insulin Pumps & Glucose Monitors
Timeline
| 6 Oct 2025 | Recall initiated by company | |
| 5 Nov 2025 | Classified by FDA | +30 days |
| 12 Nov 2025 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Insulin pump includes a vibration motor that gives tactile feedback for any alerts, alarms, or malfunctions. Pump may exhibit false vibration motor failure due to a software issue causing Malfunction 12: "Pump cannot operate, the mobile app can no longer receive data from the pump. Insulin delivery and any active CGM Sessions have been stopped", which could result in hyperglycemia.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Tandem Mobi Insulin Pump: Description/Catalog: Tandem insulin pumps are battery-operated infusion pumps capable of both basal and bolus delivery of insulin. The pumps utilize a motor-driven mechanism to deliver insulin from within a disposable cartridge, through an infusion set, into a patient s subcutaneous tissue. As with current insulin infusion pumps on the U.S. market, the desired timing and quantity of the insulin delivery is programmed by the user (i.e., the patient). Pump, Tandem Mobi, Control-IQ/1010750, Pump, Tandem Mobi, Control-IQ, Replacement/1012719, Pump, Tandem Mobi, Control-IQ 7.6.0.3, Medicare/1013501, Pump, Tandem Mobi, Control-IQ 7.7/1013655, Pump, Tandem Mobi, Control-IQ, 7.7, Replacement/1013656, Pump, Tandem Mobi, Control-IQ 7.7, Medicare/1013700, Pharmacy Kit, Starter Pack, Tandem Mobi 7.7/1014081
Where it was distributed
US: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, PT, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY
AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusetts
Questions about this recall
Why was it recalled?
Insulin pump includes a vibration motor that gives tactile feedback for any alerts, alarms, or malfunctions. Pump may exhibit false vibration motor failure due to a software issue causing Malfunction 12: "Pump cannot operate, the mobile app can no longer receive data from the pump. Insulin delivery and any active CGM Sessions have been stopped", which could result in hyperglycemia.
Which lots are affected?
Software versions: 7.6.0.1, 7.6.0.3, and 7.7.0.1 Catalog/UDI-DI: 1010750/00389152075013, 1012719/00389152271910, 1013501/00389152350110, 1013655/00389152365510, 1013656/00389152365619, 1013700/00389152370019, 1014081/00389152408170
Where was it sold?
US: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, PT, RI, SC, SD, TN, TX, UT, VA, VI, VT, WA, WI, WV, WY
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Tandem Diabetes Care
All 13| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Tandem t:slim mobile application t:slim Mobile Application, iOS recalled | Tandem Diabetes Care | Device Malfunction | Nationwide |
| Device | CLASS II | Tandem pumps are battery-operated infusion pumps capable recalled | Tandem Diabetes Care | Device Malfunction | CA |
| Device | CLASS II | t:slim X2 insulin pump with Interoperable Technology recalled | Tandem Diabetes Care | Device Malfunction | Nationwide |
| Device | CLASS II | t:slim X2 Insulin Pump with Interoperable Technology recalled | Tandem Diabetes Care | Device Malfunction | CA |
| Device | CLASS I | Tandem Mobi Insulin Pump with Interoperable Technology recalled | Tandem Diabetes Care | Labeling Errors | 29 states |
Other Insulin Pumps & Glucose Monitors recalls
All 466| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Fisher Scientific Sure-Vue Urine hCG Strip recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |
| Device | ANNOUNCED | StatStrip Glucose Hospital Meter System recalled by Nova BiomedicalNova Biomedical StatStrip | Nova Biomedical | Device Malfunction | |
| Device | CLASS I | StatStrip Glucose Hospital Meter System recalled by Nova Biomedical | Nova Biomedical | Device Malfunction | Nationwide |
| Device | CLASS I | StatStrip Glucose and B- Ketone meter recalled by Nova Biomedical | Nova Biomedical | Device Malfunction | Nationwide |