CLASS II ONGOING Device recall · Reported 18 Nov 2026 · FDA Enforcement Report
TCD Electrodes (Description recalled by Boston Scientific
Boston Scientific Neuromodulation is recalling TCD Electrodes (Description, distributed nationwide in the US. The recall was initiated on 5 Oct 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0368-2027
- FDA event ID
- 100005
- Recalling firm
- Boston Scientific Neuromodulation, Valencia, CA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 5 Oct 2026
- FDA classified
- 6 Nov 2026
- Reported
- 18 Nov 2026
- Reason category
- Foreign Material
- Device type
- Other Medical Devices
Timeline
| 5 Oct 2026 | Recall initiated by company | |
| 6 Nov 2026 | Classified by FDA | +32 days |
| 18 Nov 2026 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Radiofrequency probes may have small cut marks on the Tyvek side of the sterile pouch, and if the sterile barrier is compromised and devices are used, this could cause microbial or other foreign material contamination to be introduced into patient tissue, which may lead to infection, pain, inflammation, allergic reaction, toxicity, the need for additional medical and/or surgical intervention.
Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.
Product as listed by the FDA
TCD Electrodes (Description/REF(Inner/Outer)): TCD Electrode, Multi-Pack, 3M Cable, 10cm x 28ga (0.36mm)/(TCD-10-3M/TCD-10-3M-P); TCD Electrode, Multi-Pack, 10cm x 28ga (0.36mm)/(TCD-10/TCD-10-P); TCD Electrode, Multi-Pack, 3M Cable, 15cm x 28ga (0.36mm)/(TCD-15-3M/TCD-15-3M-P); TCD Electrode, Multi-Pack, 15cm x 28ga (0.36mm)/(TCD-15/TCD-15-P); TCD Electrode, Multi-Pack, 5cm x 28ga (0.36mm)/(TCD-5/TCD-5-P);
Where it was distributed
US Nationwide distribution including in the states of OH, ID, KY, NJ, GA, NY, MO, SC, MN, IA, MI, MA, WI, TX, ME, IN, AR, CO, IL, VT, VA, MS, WA, OR, NV, AZ, UT, FL, WY, CA, OK, NC, TN, MD, DE, LA, HI, CT, WV, AL, NH, ND.
NationwideAlabamaArkansasArizonaCaliforniaColoradoConnecticutDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKentuckyLouisianaMassachusettsMarylandMaineMichigan
Questions about this recall
Why was it recalled?
Radiofrequency probes may have small cut marks on the Tyvek side of the sterile pouch, and if the sterile barrier is compromised and devices are used, this could cause microbial or other foreign material contamination to be introduced into patient tissue, which may lead to infection, pain, inflammation, allergic reaction, toxicity, the need for additional medical and/or surgical intervention.
Which lots are affected?
Inner REF/UDI-DI/Lot(Expiration): TCD-10-3M/00813250014396/39749333(7/6/2030), 85009822(7/12/2030); TCD-10/00813250010978/39741131(7/5/2030), 39750861(7/6/2030), 39757187(7/7/2030), 39766608(7/8/2030), 39768774(7/8/2030), 39772208(7/9/2030), 39778427(7/9/2030), 39780344(7/9/2030), 39781367(7/10/2030), 85006486(7/12/2030), 85011925(7/13/2030); TCD-15-3M/00813250014402/39770714(7/8/2030); TCD-15/00813250010985/39741129(7/5/2030), 39746187(7/6/2030), 39780349(7/10/2030), 39783782(7/10/2030); TCD-5/00813250010961/39751579(7/7/2030), 39783599(7/10/2030)
Where was it sold?
US Nationwide distribution including in the states of OH, ID, KY, NJ, GA, NY, MO, SC, MN, IA, MI, MA, WI, TX, ME, IN, AR, CO, IL, VT, VA, MS, WA, OR, NV, AZ, UT, FL, WY, CA, OK, NC, TN, MD, DE, LA, HI, CT, WV, AL, NH, ND.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 25 Nov 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Boston Scientific Neuromodulation
All 26| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Nalu TD2 Therapy Disc with firmware recalled by Boston Scientific | Boston Scientific Neuromodulation | Other | Nationwide |
| Device | CLASS I | Infinion Cx Lead Kit recalled by Boston Scientific | Boston Scientific Neuromodulation | Other | NY |
| Device | CLASS II | Intracept Access Instruments, Description recalled by Boston Scientific | Boston Scientific Neuromodulation | Sterility | Nationwide |
| Device | CLASS II | Boston Scientific TCN Electrodes recalled | Boston Scientific Neuromodulation | Device Malfunction | Nationwide |
| Device | CLASS II | Boston Scientific recalled | Boston Scientific Neuromodulation | Sterility | Nationwide |
Other Other Medical Devices recalls
All 14,022| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | ANNOUNCED | Luna Bilevel Positive Airway Pressure (BPAP) System recalledLuna | React Health | Other | VT |
| Device | CLASS II | 1.5T Signa Artist, whole body magnetic resonance scanner recalled | GE Medical Systems | Other | Nationwide |
| Device | CLASS II | Artis icono biplane recalled by Siemens Medical Solutions USA | Siemens Medical Solutions USA | Other | Nationwide |
| Device | CLASS II | Physitemp Eso-1 For Esophageal And General Purpose Use recalled | Physitemp Instruments | Temperature Abuse | Nationwide |
| Device | CLASS II | Artis icono ceiling recalled by Siemens Medical Solutions USA | Siemens Medical Solutions USA | Other | Nationwide |