CLASS II ONGOING Device recall · Reported 18 Nov 2026 · FDA Enforcement Report

TCD Electrodes (Description recalled by Boston Scientific

Boston Scientific Neuromodulation is recalling TCD Electrodes (Description, distributed nationwide in the US. The recall was initiated on 5 Oct 2026 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesInner REF/UDI-DI/Lot(Expiration): TCD-10-3M/00813250014396/39749333(7/6/2030), 85009822(7/12/2030); TCD-10/00813250010978/39741131(7/5/2030), 39750861(7/6/2030), 39757187(7/7/2030), 39766608(7/8/2030), 39768774(7/8/2030), 39772208(7/9/2030), 39778427(7/9/2030), 39780344(7/9/2030), 39781367(7/10/2030), 85006486(7/12/2030), 85011925(7/13/2030); TCD-15-3M/00813250014402/39770714(7/8/2030); TCD-15/00813250010985/39741129(7/5/2030), 39746187(7/6/2030), 39780349(7/10/2030), 39783782(7/10/2030); TCD-5/00813250010961/39751579(7/7/2030), 39783599(7/10/2030)
Quantity recalled13320

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0368-2027
FDA event ID
100005
Recalling firm
Boston Scientific Neuromodulation, Valencia, CA
Classification
Class II
Status
Ongoing
Recall initiated
5 Oct 2026
FDA classified
6 Nov 2026
Reported
18 Nov 2026
Reason category
Foreign Material

Timeline

5 Oct 2026Recall initiated by company
6 Nov 2026Classified by FDA+32 days
18 Nov 2026Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Radiofrequency probes may have small cut marks on the Tyvek side of the sterile pouch, and if the sterile barrier is compromised and devices are used, this could cause microbial or other foreign material contamination to be introduced into patient tissue, which may lead to infection, pain, inflammation, allergic reaction, toxicity, the need for additional medical and/or surgical intervention.

Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.

Product as listed by the FDA

TCD Electrodes (Description/REF(Inner/Outer)): TCD Electrode, Multi-Pack, 3M Cable, 10cm x 28ga (0.36mm)/(TCD-10-3M/TCD-10-3M-P); TCD Electrode, Multi-Pack, 10cm x 28ga (0.36mm)/(TCD-10/TCD-10-P); TCD Electrode, Multi-Pack, 3M Cable, 15cm x 28ga (0.36mm)/(TCD-15-3M/TCD-15-3M-P); TCD Electrode, Multi-Pack, 15cm x 28ga (0.36mm)/(TCD-15/TCD-15-P); TCD Electrode, Multi-Pack, 5cm x 28ga (0.36mm)/(TCD-5/TCD-5-P);

Where it was distributed

US Nationwide distribution including in the states of OH, ID, KY, NJ, GA, NY, MO, SC, MN, IA, MI, MA, WI, TX, ME, IN, AR, CO, IL, VT, VA, MS, WA, OR, NV, AZ, UT, FL, WY, CA, OK, NC, TN, MD, DE, LA, HI, CT, WV, AL, NH, ND.

NationwideAlabamaArkansasArizonaCaliforniaColoradoConnecticutDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKentuckyLouisianaMassachusettsMarylandMaineMichigan

Questions about this recall

Why was it recalled?

Radiofrequency probes may have small cut marks on the Tyvek side of the sterile pouch, and if the sterile barrier is compromised and devices are used, this could cause microbial or other foreign material contamination to be introduced into patient tissue, which may lead to infection, pain, inflammation, allergic reaction, toxicity, the need for additional medical and/or surgical intervention.

Which lots are affected?

Inner REF/UDI-DI/Lot(Expiration): TCD-10-3M/00813250014396/39749333(7/6/2030), 85009822(7/12/2030); TCD-10/00813250010978/39741131(7/5/2030), 39750861(7/6/2030), 39757187(7/7/2030), 39766608(7/8/2030), 39768774(7/8/2030), 39772208(7/9/2030), 39778427(7/9/2030), 39780344(7/9/2030), 39781367(7/10/2030), 85006486(7/12/2030), 85011925(7/13/2030); TCD-15-3M/00813250014402/39770714(7/8/2030); TCD-15/00813250010985/39741129(7/5/2030), 39746187(7/6/2030), 39780349(7/10/2030), 39783782(7/10/2030); TCD-5/00813250010961/39751579(7/7/2030), 39783599(7/10/2030)

Where was it sold?

US Nationwide distribution including in the states of OH, ID, KY, NJ, GA, NY, MO, SC, MN, IA, MI, MA, WI, TX, ME, IN, AR, CO, IL, VT, VA, MS, WA, OR, NV, AZ, UT, FL, WY, CA, OK, NC, TN, MD, DE, LA, HI, CT, WV, AL, NH, ND.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.