CLASS II ONGOING Device recall · Reported 26 Aug 2026 · FDA Enforcement Report
TD Navio Midi Speech Generating Device recalled by Tobii Dynavox
Tobii Dynavox is recalling TD Navio Midi Speech Generating Device, distributed nationwide in the US. The recall was initiated on 8 Jul 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2959-2026
- FDA event ID
- 99410
- Recalling firm
- Tobii Dynavox, Pittsburgh, PA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 8 Jul 2026
- FDA classified
- 14 Aug 2026
- Reported
- 26 Aug 2026
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Other Medical Devices
Timeline
| 8 Jul 2026 | Recall initiated by company | |
| 14 Aug 2026 | Classified by FDA | +37 days |
| 26 Aug 2026 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
TD Navio Midi Speech Generating Device Model/Catalog Number: 1000798
Where it was distributed
US Nationwide distribution.
Questions about this recall
Why was it recalled?
Under certain conditions, the battery mounting tabs may become damaged, allowing the battery to move within the device enclosure, potentially leading to an internal short circuit. This may cause localized heating and gas venting within the device enclosure.
Which lots are affected?
UDI-DI 073400747602056 Serial Numbers TDNM1-240628000004 through TDNM1-260617070001. Corrected/future production is differentiated by new serialization from TDNM1-260617070001 onward (production batch 20260615, improved design).
Where was it sold?
US Nationwide distribution.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
1 other product · FDA event 99410| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS II | TD Navio Maxi Speech Generating Device recalled by Tobii Dynavox | Tobii Dynavox | Device Malfunction | Nationwide |
More recalls from Tobii Dynavox
All 3| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | TD Navio Maxi Speech Generating Device recalled by Tobii Dynavox | Tobii Dynavox | Device Malfunction | Nationwide |
| Device | CLASS II | TD I-110 Speech Device-tablet style speech generating recalled | Tobii Dynavox | Device Malfunction | Nationwide |
Other Other Medical Devices recalls
All 13,986| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | ANNOUNCED | Luna Bilevel Positive Airway Pressure (BPAP) System recalledLuna | React Health | Other | VT |
| Device | ANNOUNCED | Beds, Rails, Mattresses, and accessories recalled by MedlineMedline | Medline | Other | |
| Device | CLASS II | Hemochron Jr., Whole Blood Microcoagulation System recalled | Accriva Diagnostics | Other | Nationwide |
| Device | CLASS II | Senographe Pristina 3D recalled by GE Medical Systems | GE Medical Systems | Other | WI |
| Device | CLASS II | Mobile column TruSystem recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |