CLASS II ONGOING Device recall · Reported 16 Oct 2024 · FDA Enforcement Report

Triathlon Universal TS Baseplate Size recalled by Howmedica Osteonics

Howmedica Osteonics is recalling Triathlon Universal TS Baseplate Size 2 - A single use sterile implantable device, distributed nationwide in the US. The recall was initiated on 4 Sep 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 07613327026306 Lot numbers: LZX4LB, OLH7VA
Quantity recalled16 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0023-2025
FDA event ID
95315
Recalling firm
Howmedica Osteonics, Mahwah, NJ
Classification
Class II
Status
Ongoing
Recall initiated
4 Sep 2024
FDA classified
7 Oct 2024
Reported
16 Oct 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay

Timeline

4 Sep 2024Recall initiated by company
7 Oct 2024Classified by FDA+33 days
16 Oct 2024Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Anterior locking tabs on the Triathlon Universal Baseplate are out of specification which may prevent engagement of either of the two locking tabs with the locking wire of the tibial insert.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

Triathlon Universal TS Baseplate Size 2 - A single use sterile implantable device intended for cemented applications during primary and revision Total Knee Arthroplasty (TKA) when the Triathlon Knee System is selected for use. Catalog Number: 5521-B-200

Where it was distributed

US Nationwide distribution in the states of AZ, FL, NC, OH, PA, TN, WI. Stryker issued Urgent Medical Device Recall (UMDR) PFA 3554908 by third party Sedgwick on September 04, 2024, via UPS 2-day air. Letter states reason for recall, health risk and action to take: 1. Please inform users of this Urgent Medical Device Recall and forward this notice to all individuals who need to be made aware or organizations who have consigned product. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. 3. Quarantine and discontinue use of the recalled devices identified in the affected product list (Table 1). 4. Complete and sign the enclosed Urgent Medical Device Recall Business Reply Form: " Medical Facility: Please sign the Business Reply Form and email to: [email protected]/ Fax: (800) 871-7417 " Stryker Branch/Agency: please sign the AdobeSign form arriving to you via email. 5. Please contact your Local Sales Office or your Stryker Sales Representative directly for product replacement and inventory questions. 6. Please return the devices to Stryker using the following address: Stryker Orthopaedics/PFA Product Returns Attn: Distribution Inventory Team 325 Corporate Drive Dock M-East Mahwah, NJ 07430 Ref. PFA 3554908 Please add the tracking number Please assist us in meeting our regulatory obligation by completing and emailing the attached Urgent Medical Device Recall Business Reply Form within 5 days. A response is required, even though you may not have any physical inventory on site. If you have any questions or concerns after reviewing this letter, please contact Customer Service at (201) 831-5000. For questions pertaining to the recall, email [email protected]

Nationwide ArizonaFloridaNorth CarolinaNew JerseyOhioPennsylvaniaTennesseeWisconsin

Questions about this recall

Why was it recalled?

Anterior locking tabs on the Triathlon Universal Baseplate are out of specification which may prevent engagement of either of the two locking tabs with the locking wire of the tibial insert.

Which lots are affected?

UDI-DI: 07613327026306 Lot numbers: LZX4LB, OLH7VA

Where was it sold?

US Nationwide distribution in the states of AZ, FL, NC, OH, PA, TN, WI. Stryker issued Urgent Medical Device Recall (UMDR) PFA 3554908 by third party Sedgwick on September 04, 2024, via UPS 2-day air. Letter states reason for recall, health risk and action to take: 1. Please inform users of this Urgent Medical Device Recall and forward this notice to all individuals who need to be made aware or organizations who have consigned product. 2. Immediately check all stock areas and/or operating room storage to determine if any devices from the affected product list are at your facility. 3. Quarantine and discontinue use of the recalled devices identified in the affected product list (Table 1). 4. Complete and sign the enclosed Urgent Medical Device Recall Business Reply Form: " Medical Facility: Please sign the Business Reply Form and email to: [email protected]/ Fax: (800) 871-7417 " Stryker Branch/Agency: please sign the AdobeSign form arriving to you via email. 5. Please contact your Local Sales Office or your Stryker Sales Representative directly for product replacement and inventory questions. 6. Please return the devices to Stryker using the following address: Stryker Orthopaedics/PFA Product Returns Attn: Distribution Inventory Team 325 Corporate Drive Dock M-East Mahwah, NJ 07430 Ref. PFA 3554908 Please add the tracking number Please assist us in meeting our regulatory obligation by completing and emailing the attached Urgent Medical Device Recall Business Reply Form within 5 days. A response is required, even though you may not have any physical inventory on site. If you have any questions or concerns after reviewing this letter, please contact Customer Service at (201) 831-5000. For questions pertaining to the recall, email [email protected]

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Howmedica Osteonics

All 244

Other Hip & Knee Implants recalls

All 3,110