CLASS II Device recall · Reported 19 Feb 2014 · FDA Enforcement Report
The UF-1000i is a fully automated urine particle analyzer recalled
Sysmex America is recalling The UF-1000i is a fully automated urine particle analyzer intended for vitro diagnostic, distributed nationwide in the US. The recall was initiated on 31 Mar 2013 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0949-2014
- FDA event ID
- 65214
- Recalling firm
- Sysmex America, Mundelein, IL
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 31 Mar 2013
- FDA classified
- 10 Feb 2014
- Reported
- 19 Feb 2014
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- Test Kits & Diagnostics
Timeline
| 31 Mar 2013 | Recall initiated by company | |
| 10 Feb 2014 | Classified by FDA | +316 days |
| 19 Feb 2014 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The current Century Break patch did not contain the same setting for the Century Break year for both the Red Hat Linux and GUI files. Due to the mismatch in file settings, the software could not properly calculate a purge date for the database files based on the customer data retention setting in the application. This resulted in the database files meeting its maximum capacity and causing system p
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
The UF-1000i is a fully automated urine particle analyzer intended for vitro diagnostic use in urinalysis. The instrument can perform screening for abnormal samples with a high degree of accuracy, so it increases automation and efficiency in laboratories.
Where it was distributed
Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico and the states of AL, AR, CA, CO, CT, FL, GA, HI, IA, IS, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX, VA, WA and WI and the country of Canada.
Nationwide AlabamaArkansasCaliforniaColoradoConnecticutFloridaGeorgiaHawaiiIowaIllinoisIndianaKentuckyLouisianaMassachusettsMarylandMichiganMinnesotaMissouriMississippiMontana
Questions about this recall
Why was it recalled?
The current Century Break patch did not contain the same setting for the Century Break year for both the Red Hat Linux and GUI files. Due to the mismatch in file settings, the software could not properly calculate a purge date for the database files based on the customer data retention setting in the application. This resulted in the database files meeting its maximum capacity and causing system p
Which lots are affected?
Software versions 1.0, 1.01, 1.1.1, 1.1.2
Where was it sold?
Worldwide Distribution - USA (nationwide) Distribution including Puerto Rico and the states of AL, AR, CA, CO, CT, FL, GA, HI, IA, IS, IL, IN, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, PA, SC, SD, TN, TX, VA, WA and WI and the country of Canada.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Sysmex America
All 14| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | TS-10/TS-10H Tube Sorter, automatic sorting system recalled | Sysmex America | Device Malfunction | Nationwide |
| Device | CLASS II | Sysmex PS-10 Sample Preparation System recalled by Sysmex America | Sysmex America | Device Malfunction | Nationwide |
| Device | CLASS II | Sysmex PS-10 Sample Preparation System recalled by Sysmex America | Sysmex America | Device Malfunction | Nationwide |
| Device | CLASS II | The CF-70 instrument (product recalled by Sysmex America | Sysmex America | Device Malfunction | Nationwide |
| Device | CLASS II | Sysmex CV-11 Sample Unit, an integrated modular sampler recalled | Sysmex America | Other | FL, IL, VT |
Other Test Kits & Diagnostics recalls
All 2,663| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | DxC 500 AU Clinical Chemistry Analyzers: AU w/out ISE recalled | Beckman Coulter | Device Malfunction | Nationwide |
| Device | CLASS II | DxC 500i Clinical Analyzer Modules: DxC 500 AU Module w/ISE recalled | Beckman Coulter | Device Malfunction | Nationwide |
| Device | CLASS II | McKesson Consult hCG Urine Cassette recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |
| Device | CLASS II | Systems/units containing affected recalled | Beckman Coulter Mishima K.K | Device Malfunction | Nationwide |
| Device | CLASS II | Fisher Scientific Sure-Vue Urine hCG recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |