CLASS II ONGOING Device recall · Reported 13 Aug 2025 · FDA Enforcement Report
UIBC (Unsaturated Iron Binding Capacity) recalled by Beckman Coulter
Beckman Coulter is recalling UIBC (Unsaturated Iron Binding Capacity), distributed in 11 states. The recall was initiated on 26 Jun 2025 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-2264-2025
- FDA event ID
- 97178
- Recalling firm
- Beckman Coulter, Brea, CA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 26 Jun 2025
- FDA classified
- 4 Aug 2025
- Reported
- 13 Aug 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Potency & Assay
- Device type
- Other Medical Devices
Timeline
| 26 Jun 2025 | Recall initiated by company | |
| 4 Aug 2025 | Classified by FDA | +39 days |
| 13 Aug 2025 | Published in enforcement report | +9 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Beckman Coulter identified that the Unsaturated Iron Binding Capacity (UIBC) assay is not meeting their labeled hemoglobin interference claims of no significant interference (i.e., less than 10%) up to 200 mg/dL hemoglobin as indicated in their instructions for use. Initial internal testing confirmed that samples with low UIBC levels failed the claimed hemolysis interference specifications with reported biases up to -43.6 % UIBC when hemolyzed serum samples contained 200 mg/dL hemoglobin.
Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.
Product as listed by the FDA
UIBC (Unsaturated Iron Binding Capacity), REF: OSR61205,
Where it was distributed
US: AK, AL, AR, CO, CT, DC, DE, ID, UT OUS: Algeria, Argentina, Australia, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, China, Colombia, Costa Rica, Croatia, Czechia, Egypt, El Salvador, Ethiopia, Georgia, Germany, Ghana, Gibraltar, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Jordan, Kazakhstan, Korea, Republic of, Kuwait, Lebanon, Libya, Lithuania, Malaysia, Maldives, Mexico, Morocco, Mozambique, Myanmar, Nepal, Netherlands, Nigeria, North Macedonia, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Somalia, South Africa, Spain, Sri Lanka, Sweden, Taiwan, Province of China, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Uzbekistan, Viet Nam, Yemen.
AlaskaAlabamaArkansasColoradoConnecticutDistrict of ColumbiaDelawareGeorgiaIdahoPuerto RicoUtah
Questions about this recall
Why was it recalled?
Beckman Coulter identified that the Unsaturated Iron Binding Capacity (UIBC) assay is not meeting their labeled hemoglobin interference claims of no significant interference (i.e., less than 10%) up to 200 mg/dL hemoglobin as indicated in their instructions for use. Initial internal testing confirmed that samples with low UIBC levels failed the claimed hemolysis interference specifications with reported biases up to -43.6 % UIBC when hemolyzed serum samples contained 200 mg/dL hemoglobin.
Which lots are affected?
All Lots/(01)15099590011925
Where was it sold?
US: AK, AL, AR, CO, CT, DC, DE, ID, UT OUS: Algeria, Argentina, Australia, Bahrain, Bangladesh, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, China, Colombia, Costa Rica, Croatia, Czechia, Egypt, El Salvador, Ethiopia, Georgia, Germany, Ghana, Gibraltar, Greece, Hong Kong, Hungary, India, Indonesia, Ireland, Israel, Italy, Jordan, Kazakhstan, Korea, Republic of, Kuwait, Lebanon, Libya, Lithuania, Malaysia, Maldives, Mexico, Morocco, Mozambique, Myanmar, Nepal, Netherlands, Nigeria, North Macedonia, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Puerto Rico, Qatar, Romania, Russian Federation, Saudi Arabia, Serbia, Singapore, Slovakia, Somalia, South Africa, Spain, Sri Lanka, Sweden, Taiwan, Province of China, Thailand, Tunisia, Turkey, United Arab Emirates, United Kingdom of Great Britain and Northern Ireland, Uzbekistan, Viet Nam, Yemen.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Beckman Coulter
All 336| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | DxC 500 AU Clinical Chemistry Analyzers: AU w/out ISE recalled | Beckman Coulter | Device Malfunction | Nationwide |
| Device | CLASS II | DxC 500i Clinical Analyzer Modules: DxC 500 AU Module w/ISE recalled | Beckman Coulter | Device Malfunction | Nationwide |
| Device | CLASS II | DxC 500i Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | DxC 500 AU Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | REF # D23352, MicroScan LabPro V5.03 Panel Update 11 (US) recalled | Beckman Coulter | Other | Nationwide |
Other Other Medical Devices recalls
All 13,908| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Plastics Tray kits recalled by Avid Medical | Avid Medical | Foreign Material | Nationwide |
| Device | CLASS II | Heart Prep Tray kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Pack Open Heart Cfh kits recalled by Avid Medical | Avid Medical | Other | Nationwide |
| Device | CLASS II | Sperm Ab Elisa (96 well plate test) recalled by DRG International | DRG International | Other | Nationwide |
| Device | CLASS III | BeneView Series recalled by Mindray DS USA | Mindray DS USA | Device Malfunction | Nationwide |