CLASS I Device recall · Reported 7 Mar 2018 · FDA Enforcement Report
ultraView SISH Detection Kit, Ventana recalled
Ventana Medical Systems is recalling ultraView SISH Detection Kit, Ventana, distributed nationwide in the US. The recall was initiated on 19 Dec 2017 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0732-2018
- FDA event ID
- 78942
- Recalling firm
- Ventana Medical Systems, Oro Valley, AZ
- Classification
- Class I
- Status
- Terminated
- Recall initiated
- 19 Dec 2017
- FDA classified
- 14 Sep 2018
- Reported
- 7 Mar 2018
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Packaging Defects
- Device type
- Test Kits & Diagnostics
Timeline
| 19 Dec 2017 | Recall initiated by company | |
| 7 Mar 2018 | Published in enforcement report | +78 days |
| 14 Sep 2018 | Classified by FDA | +191 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.
Product as listed by the FDA
ultraView SISH Detection Kit, Ventana Part Number 780-001, Roche GMMI 05271967001 in vitro diagnostic
Where it was distributed
Worldwide distribution, including US nationwide, Argentina, Australia, Hong Kong, Indonesia, Philippines, Brazil, Chile, Thailand, Korea, Malaysia, New Zealand, Ecuador, Mexico, Pakistan, Vietnam, Germany, Japan, India, Colombia, Singapore, Uruguay, Peru, Canada, Taiwan, China,
Questions about this recall
Why was it recalled?
Increased reports of leaking and sticking reagent dispensers for horseradish peroxidase (HRP) dispensers (part of the iView, ultraView and OptiView detection kits, as well as CINtec PLUS Cytology Kit) and with Hematoxylin II.
Which lots are affected?
UDI 04015630972296, Lot Number: Y15133
Where was it sold?
Worldwide distribution, including US nationwide, Argentina, Australia, Hong Kong, Indonesia, Philippines, Brazil, Chile, Thailand, Korea, Malaysia, New Zealand, Ecuador, Mexico, Pakistan, Vietnam, Germany, Japan, India, Colombia, Singapore, Uruguay, Peru, Canada, Taiwan, China,
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
11 other products · FDA event 78942This product is part of a recall event; the main page for the event is ultraView Universal DAB Detection Kit, Ventana recalled.
More recalls from Ventana Medical Systems
All 42| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | navify Digital Pathology (on-prem) v2.5 recalled | Ventana Medical Systems | Device Malfunction | Nationwide |
| Device | CLASS II | Ventana anti-CD10 (SP67) Rabbit Monoclonal Primary Antibody recalled | Ventana Medical Systems | Other | Nationwide |
| Device | CLASS II | Ventana Pd-L1 (SP142) Ventana Pd-L1 (SP142) Assay recalled | Ventana Medical Systems | Potency & Assay | Nationwide |
| Device | CLASS II | Ventana HE 600 System, automated slide preparer, for use recalled | Ventana Medical Systems | Packaging Defects | 35 states |
| Device | CLASS II | BenchMark Ultra and Discovery Ultra Instruments recalled | Ventana Medical Systems | Packaging Defects | 52 states |
Other Test Kits & Diagnostics recalls
All 2,656| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | STA Staclot Protein recalled by Diagnostica Stago | Diagnostica Stago | Potency & Assay | Nationwide |
| Device | CLASS II | DxC 500i Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | DxC 500 AU Clinical Analyzer Modules using recalled by Beckman Coulter | Beckman Coulter | Packaging Defects | Nationwide |
| Device | CLASS II | Syva EMIT 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |
| Device | CLASS II | Beckman Coulter, Emit 2000 Quinidine Assay recalled | Siemens Healthcare Diagnostics | Potency & Assay | Nationwide |