CLASS II Device recall · Reported 12 Feb 2020 · FDA Enforcement Report
Urethral Dilator Set, Global Product No. G14185 recalled by Cook
Cook is recalling Urethral Dilator Set, Global Product No. G14185, distributed nationwide in the US. The recall was initiated on 4 Jan 2019 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0971-2020
- FDA event ID
- 82026
- Recalling firm
- Cook, Bloomington, IN
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 4 Jan 2019
- FDA classified
- 4 Feb 2020
- Reported
- 12 Feb 2020
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 4 Jan 2019 | Recall initiated by company | |
| 4 Feb 2020 | Classified by FDA | +396 days |
| 12 Feb 2020 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
Product as listed by the FDA
Urethral Dilator Set, Global Product No. G14185
Where it was distributed
Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.
Questions about this recall
Is Urethral Dilator Set, Global Product No. G14185 recalled?
Yes. Cook recalled Urethral Dilator Set, Global Product No. G14185 on 4 Jan 2019. The recall is Class II and its status is Terminated. Recall number Z-0971-2020.
Why was it recalled?
The wire guide may be incorrectly loaded into the wire guide holder, leading to the stiff tip of the wire guide being introduced into the patient instead of the flexible tip.
Which lots are affected?
8401849 8404945 8404946 8417702 8404946X
Where was it sold?
Domestic distribution to nationwide US. Foreign distribution to Australia, Belgium, Bolivia, Canada, Canary Islands, Chile, China, Colombia, Croatia, Czech Republic, Denmark, France, French Polynesia, Germany, Guatemala, Honduras, Hungary, India, Ireland, Italy, Japan, Korea, Luxembourg, Malaysia, Malta, Mexico, Mongolia, Netherlands, New Caledonia, New Zealand, Oman, Panama, Paraguay, Poland, Puerto Rico, Romania, Russian Federation, South Africa, Spain, Sweden, Switzerland, Taiwan, Trinidad and Tobago, United Arab Emirates, and the United Kingdom.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Cook
All 264| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Skinny Needle with Chiba Tip recalled by Cook | Cook | Other | Nationwide |
| Device | CLASS II | Flexor Check-Flo Introducer Ansel Modification recalled by Cook | Cook | Labeling Errors | 14 states |
| Device | CLASS II | Flexor Check-Flo Introducer Ansel Modification recalled by Cook | Cook | Labeling Errors | 14 states |
| Device | CLASS II | Transseptal Needle, Trocar recalled by Cook | Cook | Device Malfunction | Nationwide |
| Device | CLASS II | Transseptal Needle with Catheter recalled by Cook | Cook | Foreign Material | Nationwide |
Other Other Medical Devices recalls
All 13,778| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Citric Complete Dry Citric Acid Concentrate recalled | Nipro Renal Soultions USA | Other | Nationwide |
| Device | CLASS I | Kit: Epidural/Pain TQC 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | Halyard IV Start Kit Non-Sterile recalled by Avid Medical | Avid Medical | Sterility | Nationwide |
| Device | CLASS I | Kit: Single Shot Epidural 20/Cs recalled by Medical Action Industries | Medical Action Industries Inc. 306 | Sterility | 6 states |
| Device | CLASS II | AMS Convenience Kit: Dsaek Pack - RX recalled | Windstone Medical Packaging | Device Malfunction | Nationwide |