CLASS II Device recall · Reported 4 Feb 2015 · FDA Enforcement Report

VC150 Vital Signs Monitor (VC150 monitor equipped recalled

Innokas Medical OY is recalling VC150 Vital Signs Monitor (VC150 monitor equipped with Masimo SpO2 technique only). The recall was initiated on 3 Dec 2014 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSK514470022YA, SK514470023YA, SK514470024YA, SK514470020YA, SK514470021YA, SK514100003YA, SK514470019YA, SK514470001YA, SK514470017YA, SK514470018YA, SK514450021YA, SK514450022YA, SK514450019YA, SK514470004YA, SK514470005YA, SK514390001YA, SK514150002YA, SK514150004YA, SK514150005YA, SK514300002YA
Quantity recalled20 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1022-2015
FDA event ID
70068
Recalling firm
Innokas Medical OY, KEMPELE, N/A
Classification
Class II
Status
Terminated
Recall initiated
3 Dec 2014
FDA classified
28 Jan 2015
Reported
4 Feb 2015
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction
Device type
Patient Monitors

Timeline

3 Dec 2014Recall initiated by company
28 Jan 2015Classified by FDA+56 days
4 Feb 2015Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

If the SpO2 or RRa value is violating the respective alarm limit at the time of switching to monitoring mode, the monitor does not issue visual and audible alarms.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

VC150 Vital Signs Monitor (VC150 monitor equipped with Masimo SpO2 technique only); Innokas Medical CARESCAPE VC150 Rx Only Innokas Medical Oy. Catalog numbers 2067980-002, 2067980-006, 2067980-010, and 2067980-014.

Where it was distributed

Distributed in the US and Europe.

Questions about this recall

Why was it recalled?

If the SpO2 or RRa value is violating the respective alarm limit at the time of switching to monitoring mode, the monitor does not issue visual and audible alarms.

Which lots are affected?

SK514470022YA, SK514470023YA, SK514470024YA, SK514470020YA, SK514470021YA, SK514100003YA, SK514470019YA, SK514470001YA, SK514470017YA, SK514470018YA, SK514450021YA, SK514450022YA, SK514450019YA, SK514470004YA, SK514470005YA, SK514390001YA, SK514150002YA, SK514150004YA, SK514150005YA, SK514300002YA

Where was it sold?

Distributed in the US and Europe.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.