CLASS II Device recall · Reported 14 Nov 2012 · FDA Enforcement Report

Vector TAS Modular Driver, a manual driver recalled by Ormco/Sybronendo

Ormco/Sybronendo is recalling Vector TAS Modular Driver, a manual driver for Vector TAS dental screws, distributed nationwide in the US. The recall was initiated on 12 Dec 2011 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesPart Numbers 601-0007 and 601-0010, all lots
Quantity recalled4,543

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0205-2013
FDA event ID
63412
Recalling firm
Ormco/Sybronendo, Glendora, CA
Classification
Class II
Status
Terminated
Recall initiated
12 Dec 2011
FDA classified
5 Nov 2012
Reported
14 Nov 2012
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Sterility
Device type
Dental Devices

Timeline

12 Dec 2011Recall initiated by company
5 Nov 2012Classified by FDA+329 days
14 Nov 2012Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The Instructions For Use (IFUs) for Vector TAS Modular Driver for some foreign language sections incorrectly lists the sterilization temperature as 130¿C. The correct sterilization temperature should be listed as 132¿C and/or 270¿F.

A sterility recall means a product that must be sterile, such as an injection or eye drop, may be contaminated. Using a non-sterile product can cause serious infection.

Product as listed by the FDA

Vector TAS Modular Driver, a manual driver for Vector TAS dental screws.

Where it was distributed

Worldwide distribution, including Nationwide (USA) and the countries of Canada, Austria, Belgium, Bulgaria, Switzerland, Germany, Denmark, France, Estonia, Spain, Finland, Iceland, United Kingdom, Greece, Croatia, Hungary, Ireland, Italy, Luxembourg, Netherlands, Monaco, Poland, Romania, Serbia, Slovenia, Turkey, Jersey, and Martinique.

Nationwide

Questions about this recall

Why was it recalled?

The Instructions For Use (IFUs) for Vector TAS Modular Driver for some foreign language sections incorrectly lists the sterilization temperature as 130¿C. The correct sterilization temperature should be listed as 132¿C and/or 270¿F.

Which lots are affected?

Part Numbers 601-0007 and 601-0010, all lots

Where was it sold?

Worldwide distribution, including Nationwide (USA) and the countries of Canada, Austria, Belgium, Bulgaria, Switzerland, Germany, Denmark, France, Estonia, Spain, Finland, Iceland, United Kingdom, Greece, Croatia, Hungary, Ireland, Italy, Luxembourg, Netherlands, Monaco, Poland, Romania, Serbia, Slovenia, Turkey, Jersey, and Martinique.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.