CLASS II ONGOING Device recall · Reported 26 Mar 2025 · FDA Enforcement Report

Vericis, Merge Cardio recalled by Merge Healthcare

Merge Healthcare is recalling Vericis, Merge Cardio. The recall was initiated on 19 Feb 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSoftware Versions and Part Numbers: 10.0: Part Numbers: 88-00527-00, 88-00528-00, 88-00529-00, 88-00530-00, 88-00531-00, 88-00532-00, 88-00533-00, 70-00008-00; 10.1.1.1: Part Numbers: 88-01094-00; 10.1.2: Part Numbers: 88-01074-00, 88-01075-00, 88-01076-00, 88-01077-00, 88-01078-00, 88-01079-00, 88-00781-00, 88-00803-00, 88-01080-00, 70-00473-00; 10.2.0: Part Numbers: 88-01115-00, 88-01116-00, 88-01117-00, 88-01118-00, 88-01119-00, 88-01120-00, 88-01121-00, 88-01122-00, 88-01123-00, 70-00491-00, 10.2 P2: Part Numbers: 88-01203-00 10.3 P1 UDI - (01)00842000100041(10)10.3.0.2009(11)170309
Quantity recalled11

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1399-2025
FDA event ID
96344
Recalling firm
Merge Healthcare, Hartland, WI
Classification
Class II
Status
Ongoing
Recall initiated
19 Feb 2025
FDA classified
20 Mar 2025
Reported
26 Mar 2025
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Device Malfunction

Timeline

19 Feb 2025Recall initiated by company
20 Mar 2025Classified by FDA+29 days
26 Mar 2025Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

VERICIS, Merge Cardio, Model/Catalog Number: Version 10.x, Software Version: 10.0, 10.1.1.1, 10.1.2, 10.2.0, 10.2 P2 , 10.3; The VERICIS is an integrated cardiovascular information system classified as a picture archiving and communications system.

Where it was distributed

US

Questions about this recall

Why was it recalled?

When following a specific less typical workflow in which the user remeasures using the ultrasound device or manually edits primitive or derived measurement values directly in Clinical Reporting, derived measurements may be inconsistent with the associated primitive measurements in the final patient report.

Which lots are affected?

Software Versions and Part Numbers: 10.0: Part Numbers: 88-00527-00, 88-00528-00, 88-00529-00, 88-00530-00, 88-00531-00, 88-00532-00, 88-00533-00, 70-00008-00; 10.1.1.1: Part Numbers: 88-01094-00; 10.1.2: Part Numbers: 88-01074-00, 88-01075-00, 88-01076-00, 88-01077-00, 88-01078-00, 88-01079-00, 88-00781-00, 88-00803-00, 88-01080-00, 70-00473-00; 10.2.0: Part Numbers: 88-01115-00, 88-01116-00, 88-01117-00, 88-01118-00, 88-01119-00, 88-01120-00, 88-01121-00, 88-01122-00, 88-01123-00, 70-00491-00, 10.2 P2: Part Numbers: 88-01203-00 10.3 P1 UDI - (01)00842000100041(10)10.3.0.2009(11)170309

Where was it sold?

US

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

4 other products · FDA event 96344

This product is part of a recall event; the main page for the event is Vericis, Merge Cardio recalled by Merge Healthcare.

More recalls from Merge Healthcare

All 96

Other Heart & Vascular Devices recalls

All 6,217