CLASS II ONGOING Device recall · Reported 18 Nov 2026 · FDA Enforcement Report
Vesco Q Small Bore Bag Feed Set recalled by Vesco Medical
Vesco Medical is recalling Vesco Q Small Bore Bag Feed Set, distributed nationwide in the US. The recall was initiated on 8 Oct 2026 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0379-2027
- FDA event ID
- 100054
- Recalling firm
- Vesco Medical, Westerville, OH
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 8 Oct 2026
- FDA classified
- 6 Nov 2026
- Reported
- 18 Nov 2026
- Reason category
- Other Reasons
- Device type
- Other Medical Devices
Timeline
| 8 Oct 2026 | Recall initiated by company | |
| 6 Nov 2026 | Classified by FDA | +29 days |
| 18 Nov 2026 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the prescribed dose, if proper controls are not in place. Over- or under-delivery of enteral nutrition through the pump and associated feed sets could cause temporary or medically reversible adverse health consequences, with a remote probability of serious adverse health consequences.
Product as listed by the FDA
Vesco Q 100mL Small Bore Bag Feed Set, Model Number VED-059
Where it was distributed
US Nationwide distribution.
Questions about this recall
Why was it recalled?
The feeding pump may deliver enteral fluids outside its labeled accuracy specification of +/-5%. This inaccuracy can result in over-delivery or under-delivery of enteral fluids relative to the prescribed dose, if proper controls are not in place. Over- or under-delivery of enteral nutrition through the pump and associated feed sets could cause temporary or medically reversible adverse health consequences, with a remote probability of serious adverse health consequences.
Which lots are affected?
UDI-DI 00810096151377; Lot Numbers: all lots manufactured through 03/26/2026
Where was it sold?
US Nationwide distribution.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 25 Nov 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
6 other products · FDA event 100054This product is part of a recall event; the main page for the event is Vesco Q Screw Cap Feed Set recalled by Vesco Medical.
| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS II | Vesco Q Screw Cap Feed Set recalled by Vesco Medical | Vesco Medical | Other | Nationwide | |
| CLASS II | Vesco Q Bag Feed Set (LL) recalled by Vesco Medical | Vesco Medical | Other | Nationwide | |
| CLASS II | Vesco Q Bag Feed Set (PVC) recalled by Vesco Medical | Vesco Medical | Other | Nationwide | |
| CLASS II | Vesco Q ENPlus Spike Feed Set recalled by Vesco Medical | Vesco Medical | Other | Nationwide | |
| CLASS II | Vesco Q Bag Feed Set (PVC) recalled by Vesco Medical | Vesco Medical | Other | Nationwide | |
| CLASS II | Vesco Q Enteral Feeding Pump recalled by Vesco Medical | Vesco Medical | Device Malfunction | Nationwide |
More recalls from Vesco Medical
All 7| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Vesco Q Enteral Feeding Pump recalled by Vesco Medical | Vesco Medical | Device Malfunction | Nationwide |
| Device | CLASS II | Vesco Q Bag Feed Set (PVC) recalled by Vesco Medical | Vesco Medical | Other | Nationwide |
| Device | CLASS II | Vesco Q ENPlus Spike Feed Set recalled by Vesco Medical | Vesco Medical | Other | Nationwide |
| Device | CLASS II | Vesco Q Bag Feed Set (PVC) recalled by Vesco Medical | Vesco Medical | Other | Nationwide |
| Device | CLASS II | Vesco Q Bag Feed Set (LL) recalled by Vesco Medical | Vesco Medical | Other | Nationwide |
Other Other Medical Devices recalls
All 14,022| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | ANNOUNCED | Luna Bilevel Positive Airway Pressure (BPAP) System recalledLuna | React Health | Other | VT |
| Device | CLASS II | 1.5T Signa Artist, whole body magnetic resonance scanner recalled | GE Medical Systems | Other | Nationwide |
| Device | CLASS II | Artis icono biplane recalled by Siemens Medical Solutions USA | Siemens Medical Solutions USA | Other | Nationwide |
| Device | CLASS II | Physitemp Eso-1 For Esophageal And General Purpose Use recalled | Physitemp Instruments | Temperature Abuse | Nationwide |
| Device | CLASS II | Artis icono ceiling recalled by Siemens Medical Solutions USA | Siemens Medical Solutions USA | Other | Nationwide |