CLASS II Device recall · Reported 24 Apr 2013 · FDA Enforcement Report

Vitros 5,1 FS Chemistry System, Product recalled

Ortho-Clinical Diagnostics is recalling Vitros 5,1 FS Chemistry System, distributed nationwide in the US. The recall was initiated on 7 Feb 2013 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesVITROS 5,1 FS Chemistry System Serial Numbers: 34000119 through 34002273; VITROS 5,1 FS Chemistry System Refurbished Serial Numbers: 34000133, 34001240, 34000193, 34001243, 34000212, 34001248, 34000316, 34001251, 34000318, 34001274, 34000337, 34001297, 34000360, 34001311, 34000420, 34001317, 34000451, 34001345, 34000478, 34001378, 34000495, 34001389, 34000497, 34001390, 34000529, 34001396, 34000538, 34001422, 34000552, 34001472, 34000557, 34001504, 34000596, 34001569, 34000601, 34001572, 34000631, 34001589, 34000642, 34001610, 34000799, 34001633, 34000805, 34001668, 34000833, 34001809, 34000851, 34001840, 34000893, 34001850, 34000947, 34001897, 34001172, 34001218
Quantity recalledVITROS 5,1 FS: USA: 670 units, Foreign: 1249 units; VITROS 5,1 FS Refurbished: USA: 52 units, Foreign: 2 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1106-2013
FDA event ID
64577
Recalling firm
Ortho-Clinical Diagnostics, Rochester, NY
Classification
Class II
Status
Terminated
Recall initiated
7 Feb 2013
FDA classified
13 Apr 2013
Reported
24 Apr 2013
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

7 Feb 2013Recall initiated by company
13 Apr 2013Classified by FDA+65 days
24 Apr 2013Published in enforcement report+11 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Ortho-Clinical Diagnostics, Inc. (OCD) determined that extension springs on some VITROS¿ Systems may be out of its manufacturing specifications. Extension springs are part of the slide alignment guide assembly. During an internal evaluation in its manufacturing facility, it was discovered that an extension spring had insufficient force. If the extension spring does not provide sufficient force, MicroSlides may not be properly positioned within the MicroSlide incubator which has the potential to cause imprecise results. Extension springs are just one component in the slide alignment assembly. OCD's evaluation determined that the probability of an out of specification extension spring alone adversely affecting results is unlikely.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

VITROS 5,1 FS Chemistry System, Product Code 6801375 and VITROS 5,1 FS Chemistry System Refurbished, Product Code 6801890. Intended for use in the measurement of a variety of chemistry analytes.

Where it was distributed

Worldwide Distribution- USA (nationwide) including Puerto Rico and countries of: Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Republic of Panama, Singapore, Spain, and Venezuela.

Nationwide Puerto Rico

Questions about this recall

Why was it recalled?

Ortho-Clinical Diagnostics, Inc. (OCD) determined that extension springs on some VITROS¿ Systems may be out of its manufacturing specifications. Extension springs are part of the slide alignment guide assembly. During an internal evaluation in its manufacturing facility, it was discovered that an extension spring had insufficient force. If the extension spring does not provide sufficient force, MicroSlides may not be properly positioned within the MicroSlide incubator which has the potential to cause imprecise results. Extension springs are just one component in the slide alignment assembly. OCD's evaluation determined that the probability of an out of specification extension spring alone adversely affecting results is unlikely.

Which lots are affected?

VITROS 5,1 FS Chemistry System Serial Numbers: 34000119 through 34002273; VITROS 5,1 FS Chemistry System Refurbished Serial Numbers: 34000133, 34001240, 34000193, 34001243, 34000212, 34001248, 34000316, 34001251, 34000318, 34001274, 34000337, 34001297, 34000360, 34001311, 34000420, 34001317, 34000451, 34001345, 34000478, 34001378, 34000495, 34001389, 34000497, 34001390, 34000529, 34001396, 34000538, 34001422, 34000552, 34001472, 34000557, 34001504, 34000596, 34001569, 34000601, 34001572, 34000631, 34001589, 34000642, 34001610, 34000799, 34001633, 34000805, 34001668, 34000833, 34001809, 34000851, 34001840, 34000893, 34001850, 34000947, 34001897, 34001172, 34001218

Where was it sold?

Worldwide Distribution- USA (nationwide) including Puerto Rico and countries of: Australia, Brazil, Canada, Chile, China, Colombia, England, France, Germany, India, Italy, Japan, Republic of Panama, Singapore, Spain, and Venezuela.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

2 other products · FDA event 64577

This product is part of a recall event; the main page for the event is Vitros 4600 Chemistry System recalled by Ortho-Clinical Diagnostics.