CLASS II Device recall · Reported 21 Nov 2018 · FDA Enforcement Report

Xper Flex Cardio Physiomonitoring System recalled by Invivo

Invivo is recalling Xper Flex Cardio Physiomonitoring System. The recall was initiated on 24 Sep 2018 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel Number 453564669081; System Serial Number (UDI): US1814000001 ((01)00884838091474(21)US1814000001), US1814000002 ((01)00884838091474(21)US1814000002), US1814000003 ((01)00884838091474(21)US1814000003), US1814000004 ((01)00884838091474(21)US1814000004), US1814000006 ((01)00884838091474(21)US1814000006), US1814000007 ((01)00884838091474(21)US1814000007), US1814000008 ((01)00884838091474(21)US1814000008), US1814000011 ((01)00884838091474(21)US1814000011), US1814000012 ((01)00884838091474(21)US1814000012), US1814000013 ((01)00884838091474(21)US1814000013), US1814000014 ((01)00884838091474(21)US1814000014), US1814000016 ((01)00884838091474(21)US1814000016), US1814000017 ((01)00884838091474(21)US1814000017), US1814000018 ((01)00884838091474(21)US1814000018), US1814000019 ((01)00884838091474(21)US1814000019), US1814000020 ((01)00884838091474(21)US1814000020), US1814000022 ((01)00884838091474(21)US1814000022), US1814000023 ((01)00884838091474(21)US1814000023), US1814000024 ((01)00884838091474(21)US1814000024), US1814000027 ((01)00884838091474(21)US1814000027), US1814000028 ((01)00884838091474(21)US1814000028), US1814000029 ((01)00884838091474(21)US1814000029), US1814000030 ((01)00884838091474(21)US1814000030), US1814000031 ((01)00884838091474(21)US1814000031), US1814000032 ((01)00884838091474(21)US1814000032), US1814000033 ((01)00884838091474(21)US1814000033), US1814000034 ((01)00884838091474(21)US1814000034), US1814000035 ((01)00884838091474(21)US1814000035), US1814000036 ((01)00884838091474(21)US1814000036), US1814000037 ((01)00884838091474(21)US1814000037), US1814000038 ((01)00884838091474(21)US1814000038), US1814000039 ((01)00884838091474(21)US1814000039), US1814000040 ((01)00884838091474(21)US1814000040), US1814000041 ((01)00884838091474(21)US1814000041), US1814000043 ((01)00884838091474(21)US1814000043), US1814000044 ((01)00884838091474(21)US1814000044), US1814000045 ((01)00884838091474(21)US1814000045), US1814000046 ((01)00884838091474(21)US1814000046), US1814000047 ((01)00884838091474(21)US1814000047), US1814000048 ((01)00884838091474(21)US1814000048), US1814000049 ((01)00884838091474(21)US1814000049), US1814000050 ((01)00884838091474(21)US1814000050), US1814000051 ((01)00884838091474(21)US1814000051), US1814000052 ((01)00884838091474(21)US1814000052), US1814000053 ((01)00884838091474(21)US1814000053), US1814000054 ((01)00884838091474(21)US1814000054), US1814000055 ((01)00884838091474(21)US1814000055), US1814000056 ((01)00884838091474(21)US1814000056), US1814000058 ((01)00884838091474(21)US1814000058), US1814000059 ((01)00884838091474(21)US1814000059), US1814000060 ((01)00884838091474(21)US1814000060), US1814000061 ((01)00884838091474(21)US1814000061), US1814000062 ((01)00884838091474(21)US1814000062), US1814000063 ((01)00884838091474(21)US1814000063), US1814000064 ((01)00884838091474(21)US1814000064), US1814000065 ((01)00884838091474(21)US1814000065), US1814000066 ((01)00884838091474(21)US1814000066), US1814000067 ((01)00884838091474(21)US1814000067), US1814000068 ((01)00884838091474(21)US1814000068), US1814000069 ((01)00884838091474(21)US1814000069), US1814000070 ((01)00884838091474(21)US1814000070), US1814000072 ((01)00884838091474(21)US1814000072), US1814000074 ((01)00884838091474(21)US1814000074), US1814000078 ((01)00884838091474(21)US1814000078), US1814000081 ((01)00884838091474(21)US1814000081), US1814000082 ((01)00884838091474(21)US1814000082), US1814000083 ((01)00884838091474(21)US1814000083), US1814000084 ((01)00884838091474(21)US1814000084), US1814000085 ((01)00884838091474(21)US1814000085), US1814000086 ((01)00884838091474(21)US1814000086), US1814000087 ((01)00884838091474(21)US1814000087), US1814000088 ((01)00884838091474(21)US1814000088), US1814000089 ((01)00884838091474(21)US1814000089), US1814000092 ((01)00884838091474(21)US1814000092), US1814000094 ((01)00884838091474(21)US1814000094), US1814000096 ((01)00884838091474(21)US1814000096),
Quantity recalled76 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0462-2019
FDA event ID
81092
Recalling firm
Invivo, Orlando, FL
Classification
Class II
Status
Completed
Recall initiated
24 Sep 2018
FDA classified
14 Nov 2018
Reported
21 Nov 2018
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction
Device type
Patient Monitors

Timeline

24 Sep 2018Recall initiated by company
14 Nov 2018Classified by FDA+51 days
21 Nov 2018Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The Invasive Blood Pressure (IBP) numeric values will freeze on the display when any of the following functions are performed: change an IBP Site Label using the Site Label icon or the keyboard, activate or deactivate the mean pressure waveform display option (On/Off) using the Invasive Pressure Options Dialog or the keyboard, activate or deactivate the waveform hide option (On/Off) using the Invasive Pressure Options Dialog, activate or deactivate the pressure waveform filter option (On/Off) using the keyboard, and/or perform a Pullback using the Pullback icon or the keyboard

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Xper Flex Cardio Physiomonitoring System, Software version 1.6.0.0388

Where it was distributed

Distributed to the Netherlands

Questions about this recall

Is Xper Flex Cardio Physiomonitoring System recalled?

Yes. Invivo recalled Xper Flex Cardio Physiomonitoring System on 24 Sep 2018. The recall is Class II and its status is Completed. Recall number Z-0462-2019.

Why was it recalled?

The Invasive Blood Pressure (IBP) numeric values will freeze on the display when any of the following functions are performed: change an IBP Site Label using the Site Label icon or the keyboard, activate or deactivate the mean pressure waveform display option (On/Off) using the Invasive Pressure Options Dialog or the keyboard, activate or deactivate the waveform hide option (On/Off) using the Invasive Pressure Options Dialog, activate or deactivate the pressure waveform filter option (On/Off) using the keyboard, and/or perform a Pullback using the Pullback icon or the keyboard

Which lots are affected?

Model Number 453564669081; System Serial Number (UDI): US1814000001 ((01)00884838091474(21)US1814000001), US1814000002 ((01)00884838091474(21)US1814000002), US1814000003 ((01)00884838091474(21)US1814000003), US1814000004 ((01)00884838091474(21)US1814000004), US1814000006 ((01)00884838091474(21)US1814000006), US1814000007 ((01)00884838091474(21)US1814000007), US1814000008 ((01)00884838091474(21)US1814000008), US1814000011 ((01)00884838091474(21)US1814000011), US1814000012 ((01)00884838091474(21)US1814000012), US1814000013 ((01)00884838091474(21)US1814000013), US1814000014 ((01)00884838091474(21)US1814000014), US1814000016 ((01)00884838091474(21)US1814000016), US1814000017 ((01)00884838091474(21)US1814000017), US1814000018 ((01)00884838091474(21)US1814000018), US1814000019 ((01)00884838091474(21)US1814000019), US1814000020 ((01)00884838091474(21)US1814000020), US1814000022 ((01)00884838091474(21)US1814000022), US1814000023 ((01)00884838091474(21)US1814000023), US1814000024 ((01)00884838091474(21)US1814000024), US1814000027 ((01)00884838091474(21)US1814000027), US1814000028 ((01)00884838091474(21)US1814000028), US1814000029 ((01)00884838091474(21)US1814000029), US1814000030 ((01)00884838091474(21)US1814000030), US1814000031 ((01)00884838091474(21)US1814000031), US1814000032 ((01)00884838091474(21)US1814000032), US1814000033 ((01)00884838091474(21)US1814000033), US1814000034 ((01)00884838091474(21)US1814000034), US1814000035 ((01)00884838091474(21)US1814000035), US1814000036 ((01)00884838091474(21)US1814000036), US1814000037 ((01)00884838091474(21)US1814000037), US1814000038 ((01)00884838091474(21)US1814000038), US1814000039 ((01)00884838091474(21)US1814000039), US1814000040 ((01)00884838091474(21)US1814000040), US1814000041 ((01)00884838091474(21)US1814000041), US1814000043 ((01)00884838091474(21)US1814000043), US1814000044 ((01)00884838091474(21)US1814000044), US1814000045 ((01)00884838091474(21)US1814000045), US1814000046 ((01)00884838091474(21)US1814000046), US1814000047 ((01)00884838091474(21)US1814000047), US1814000048 ((01)00884838091474(21)US1814000048), US1814000049 ((01)00884838091474(21)US1814000049), US1814000050 ((01)00884838091474(21)US1814000050), US1814000051 ((01)00884838091474(21)US1814000051), US1814000052 ((01)00884838091474(21)US1814000052), US1814000053 ((01)00884838091474(21)US1814000053), US1814000054 ((01)00884838091474(21)US1814000054), US1814000055 ((01)00884838091474(21)US1814000055), US1814000056 ((01)00884838091474(21)US1814000056), US1814000058 ((01)00884838091474(21)US1814000058), US1814000059 ((01)00884838091474(21)US1814000059), US1814000060 ((01)00884838091474(21)US1814000060), US1814000061 ((01)00884838091474(21)US1814000061), US1814000062 ((01)00884838091474(21)US1814000062), US1814000063 ((01)00884838091474(21)US1814000063), US1814000064 ((01)00884838091474(21)US1814000064), US1814000065 ((01)00884838091474(21)US1814000065), US1814000066 ((01)00884838091474(21)US1814000066), US1814000067 ((01)00884838091474(21)US1814000067), US1814000068 ((01)00884838091474(21)US1814000068), US1814000069 ((01)00884838091474(21)US1814000069), US1814000070 ((01)00884838091474(21)US1814000070), US1814000072 ((01)00884838091474(21)US1814000072), US1814000074 ((01)00884838091474(21)US1814000074), US1814000078 ((01)00884838091474(21)US1814000078), US1814000081 ((01)00884838091474(21)US1814000081), US1814000082 ((01)00884838091474(21)US1814000082), US1814000083 ((01)00884838091474(21)US1814000083), US1814000084 ((01)00884838091474(21)US1814000084), US1814000085 ((01)00884838091474(21)US1814000085), US1814000086 ((01)00884838091474(21)US1814000086), US1814000087 ((01)00884838091474(21)US1814000087), US1814000088 ((01)00884838091474(21)US1814000088), US1814000089 ((01)00884838091474(21)US1814000089), US1814000092 ((01)00884838091474(21)US1814000092), US1814000094 ((01)00884838091474(21)US1814000094), US1814000096 ((01)00884838091474(21)US1814000096),

Where was it sold?

Distributed to the Netherlands

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.