CLASS III Device recall · Reported 26 Aug 2015 · FDA Enforcement Report

Xpert CT /NG Vaginal/Endocervical/ Specimen Collection Kit recalled

Cepheid is recalling Xpert CT /NG Vaginal/Endocervical/ Specimen Collection Kit, distributed nationwide in the US. The recall was initiated on 27 Jul 2015 and the FDA has classified it as Class III.

Check your package

Lot, batch or date codesPart number: CT/NGSWAB-50: Lot numbers: 46A202C, 46A202D, 46A202E, 46A202F, 46A202G, Flock swab expires end of July 2015. Kit expires October 2015. Lot numbers 26N121H, and 28N134D.- Cleaning swabs expire end of June 2016. Kits expires November 2016.
Quantity recalled729 kits US, 47 kits ROW.

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2431-2015
FDA event ID
71841
Recalling firm
Cepheid, Sunnyvale, CA
Classification
Class III
Status
Terminated
Recall initiated
27 Jul 2015
FDA classified
20 Aug 2015
Reported
26 Aug 2015
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay

Timeline

27 Jul 2015Recall initiated by company
20 Aug 2015Classified by FDA+24 days
26 Aug 2015Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Firm determined that some lots of CT/NG Swab 50 and CT/NG Urine collection kits contain a sub component with an expiry date that is earlier than the kit expiration date.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

Xpert CT /NG Vaginal/Endocervical/ Specimen Collection Kit: Part number CT/NGSWAB-50; Microbiology: The Cepheid X pert CT /NG Vagina Endocervical/ Specimen Collection Kit is designed to collect, preserve and transport patient Chlamydia trachomatis and Neisseria gonorrhoeae DNA in endocervical and vaginal specimens from symptomatic and asymptomatic individuals prior to analysis with the Cepheid Xpert CT/NG Assay.

Where it was distributed

Worldwide distribution. US nationwide, including North Mariana Islands, Aruba, Canada, Chile, Colombia, Costa Rica, Cyprus, Guam, Guatemala, Italy, Malawi, Micronesia Federated States, Panama, San Marino, South Africa, Spain, Switzerland, and Zimbabwe.

Nationwide

Questions about this recall

Why was it recalled?

Firm determined that some lots of CT/NG Swab 50 and CT/NG Urine collection kits contain a sub component with an expiry date that is earlier than the kit expiration date.

Which lots are affected?

Part number: CT/NGSWAB-50: Lot numbers: 46A202C, 46A202D, 46A202E, 46A202F, 46A202G, Flock swab expires end of July 2015. Kit expires October 2015. Lot numbers 26N121H, and 28N134D.- Cleaning swabs expire end of June 2016. Kits expires November 2016.

Where was it sold?

Worldwide distribution. US nationwide, including North Mariana Islands, Aruba, Canada, Chile, Colombia, Costa Rica, Cyprus, Guam, Guatemala, Italy, Malawi, Micronesia Federated States, Panama, San Marino, South Africa, Spain, Switzerland, and Zimbabwe.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class III mean?

Class III recalls are for products unlikely to cause any adverse health consequence, usually labelling or quality issues.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

1 other product · FDA event 71841
ReportedClassRecallCompanyReasonWhere
CLASS IIIXpert CT/NG Urine Specimen Collection Kit recalled by CepheidCepheidPotency & AssayNationwide

More recalls from Cepheid

All 12

Other Test Kits & Diagnostics recalls

All 2,663