Zap-x Radiosurgery System recalled by Zap Surgical Systems
Zap Surgical Systems is recalling Zap-x Radiosurgery System, distributed nationwide in the US. The recall was initiated on 5 Jul 2022 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1742-2022
- FDA event ID
- 90716
- Recalling firm
- Zap Surgical Systems, San Carlos, CA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 5 Jul 2022
- FDA classified
- 14 Sep 2022
- Reported
- 21 Sep 2022
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Processing Deviation
- Device type
- Other Medical Devices
Timeline
| 5 Jul 2022 | Recall initiated by company | |
| 14 Sep 2022 | Classified by FDA | +71 days |
| 21 Sep 2022 | Published in enforcement report | +7 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Software issue identified in cases of initial patient setup with large (greater-than-or-equal-to 1.5 degrees) rotational deviations, between digitally reconstructed radiographs, and pre-delivery X-ray images. In these cases the transitional alignment algorithm may incorrectly calculate the new treatment table offset values, leading to an incorrect position for the subsequent treatment isocenter.
The USDA issues these when meat or poultry was produced without proper inspection, or when a step such as cooking or pasteurisation did not meet requirements.
Product as listed by the FDA
ZAP-X Radiosurgery System
Where it was distributed
Worldwide distribution - US Nationwide distribution in the states of AZ, MD, FL, CO and the countries of DEU, CHE, JPN.
Questions about this recall
Why was it recalled?
Software issue identified in cases of initial patient setup with large (greater-than-or-equal-to 1.5 degrees) rotational deviations, between digitally reconstructed radiographs, and pre-delivery X-ray images. In these cases the transitional alignment algorithm may incorrectly calculate the new treatment table offset values, leading to an incorrect position for the subsequent treatment isocenter.
Which lots are affected?
UDI-DI: 00860183001504, Serial Numbers, ZUC17152, ZUC17153, ZUC18155, ZUC20157, ZUC20158, ZUC20156, ZUC21161, ZUC21162, using Treatment Delivery Software Version: 1.8.53
Where was it sold?
Worldwide distribution - US Nationwide distribution in the states of AZ, MD, FL, CO and the countries of DEU, CHE, JPN.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Zap Surgical Systems
All 2| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Zap-x Radiosurgery System recalled by Zap Surgical Systems | Zap Surgical Systems | Device Malfunction | Nationwide |
Other Other Medical Devices recalls
All 13,986| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | ANNOUNCED | Luna Bilevel Positive Airway Pressure (BPAP) System recalledLuna | React Health | Other | VT |
| Device | ANNOUNCED | Beds, Rails, Mattresses, and accessories recalled by MedlineMedline | Medline | Other | |
| Device | CLASS II | Hemochron Jr., Whole Blood Microcoagulation System recalled | Accriva Diagnostics | Other | Nationwide |
| Device | CLASS II | Senographe Pristina 3D recalled by GE Medical Systems | GE Medical Systems | Other | WI |
| Device | CLASS II | Mobile column TruSystem recalled by Baxter Healthcare | Baxter Healthcare | Device Malfunction | Nationwide |