CLASS II ONGOING Device recall · Reported 21 Sep 2022 · FDA Enforcement Report

Zap-x Radiosurgery System recalled by Zap Surgical Systems

Zap Surgical Systems is recalling Zap-x Radiosurgery System, distributed nationwide in the US. The recall was initiated on 5 Jul 2022 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesUDI-DI: 00860183001504, Serial Numbers, ZUC17152, ZUC17153, ZUC18155, ZUC20157, ZUC20158, ZUC20156, ZUC21161, ZUC21162, using Treatment Delivery Software Version: 1.8.53
Quantity recalled8

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1742-2022
FDA event ID
90716
Recalling firm
Zap Surgical Systems, San Carlos, CA
Classification
Class II
Status
Ongoing
Recall initiated
5 Jul 2022
FDA classified
14 Sep 2022
Reported
21 Sep 2022
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Processing Deviation

Timeline

5 Jul 2022Recall initiated by company
14 Sep 2022Classified by FDA+71 days
21 Sep 2022Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Software issue identified in cases of initial patient setup with large (greater-than-or-equal-to 1.5 degrees) rotational deviations, between digitally reconstructed radiographs, and pre-delivery X-ray images. In these cases the transitional alignment algorithm may incorrectly calculate the new treatment table offset values, leading to an incorrect position for the subsequent treatment isocenter.

The USDA issues these when meat or poultry was produced without proper inspection, or when a step such as cooking or pasteurisation did not meet requirements.

Product as listed by the FDA

ZAP-X Radiosurgery System

Where it was distributed

Worldwide distribution - US Nationwide distribution in the states of AZ, MD, FL, CO and the countries of DEU, CHE, JPN.

Nationwide ArizonaColoradoFloridaMaryland

Questions about this recall

Why was it recalled?

Software issue identified in cases of initial patient setup with large (greater-than-or-equal-to 1.5 degrees) rotational deviations, between digitally reconstructed radiographs, and pre-delivery X-ray images. In these cases the transitional alignment algorithm may incorrectly calculate the new treatment table offset values, leading to an incorrect position for the subsequent treatment isocenter.

Which lots are affected?

UDI-DI: 00860183001504, Serial Numbers, ZUC17152, ZUC17153, ZUC18155, ZUC20157, ZUC20158, ZUC20156, ZUC21161, ZUC21162, using Treatment Delivery Software Version: 1.8.53

Where was it sold?

Worldwide distribution - US Nationwide distribution in the states of AZ, MD, FL, CO and the countries of DEU, CHE, JPN.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Zap Surgical Systems

All 2
ReportedClassRecallCompanyReasonWhere
DeviceCLASS IIZap-x Radiosurgery System recalled by Zap Surgical SystemsZap Surgical SystemsDevice MalfunctionNationwide

Other Other Medical Devices recalls

All 13,986