CLASS II Device recall · Reported 22 May 2019 · FDA Enforcement Report

Abbott Alinity ci-series System Control Module recalled

Abbott Gmbh & Co. KG is recalling Abbott Alinity ci-series System Control Module, distributed in 19 states. The recall was initiated on 7 Mar 2019 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesAlinity ci series System Control Module; LN 3R70 01 Level Sensor, Bulk Solution; LN 04S68 01 Accessory kit, Alinity i; LN 03R66 01, LN 03R66 02 Accessory kit, Alinity c; LN 03R69 01, LN03R69 02
Quantity recalled1557

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1316-2019
FDA event ID
82475
Recalling firm
Abbott Gmbh & Co. KG, Wiesbaden
Classification
Class II
Status
Terminated
Recall initiated
7 Mar 2019
FDA classified
13 May 2019
Reported
22 May 2019
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

7 Mar 2019Recall initiated by company
13 May 2019Classified by FDA+67 days
22 May 2019Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Potential performance issues in the Alinity-ci software version 2.5.1

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Abbott Alinity ci-series System Control Module software version 2.5.1; LN 3R70-01 with the following hardware: Level Sensor, Bulk Solution; 04S68-01, Accessory Kit, Alinity I; 03R66-01and 03R66-02 , and Accessory Kit, Alinity c; 03R69-01 and 03R69-02. Product Usage: The Alinity i (LN 03R65 01) processing module is a fully automated immunoassay analyzer allowing random and continuous access, as well as priority and automated retest processing using chemiluminescent microparticle immunoassay (CMIA) technology. CMIA technology is used to determine the presence of antigens, antibodies, and analytes in samples. The Alinity c (LN 03R67 01) processing module is a fully automated chemistry analyzer allowing random and continuous access, as well as priority and automated retest processing using photometric and potentiometric detection technology. The Alinity c processing module uses photometric detection technology to measure sample absorbance for the quantification of analyte concentration and uses potentiometric detection technology to measure the electrical potential in a sample. In addition, the Alinity c processing module uses an integrated chip technology (ICT) module to measure potentiometric assays (electrolytes).

Where it was distributed

AR, AZ, CA, FL, GA, IL, MA, ME, MN, MO, NY, OK, SC, TN, TX, UT, VA, WI, and Puerto Rico ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, GUATEMALA, HONG KONG, INDONESIA, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, KENYA, KUWAIT, LATVIA, LEBANON, MADAGASCAR, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, PAKISTAN, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD & TOBAGO, TURKEY, UAE, UK, and VIETNAM.

ArkansasArizonaCaliforniaFloridaGeorgiaIllinoisMassachusettsMaineMinnesotaMissouriNew YorkOklahomaPuerto RicoSouth CarolinaTennesseeTexasUtahVirginiaWisconsin

Questions about this recall

Why was it recalled?

Potential performance issues in the Alinity-ci software version 2.5.1

Which lots are affected?

Alinity ci series System Control Module; LN 3R70 01 Level Sensor, Bulk Solution; LN 04S68 01 Accessory kit, Alinity i; LN 03R66 01, LN 03R66 02 Accessory kit, Alinity c; LN 03R69 01, LN03R69 02

Where was it sold?

AR, AZ, CA, FL, GA, IL, MA, ME, MN, MO, NY, OK, SC, TN, TX, UT, VA, WI, and Puerto Rico ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BAHRAIN, BANGLADESH, BELGIUM, BOSNIA AND HERZEGOVINA, BRAZIL, BULGARIA, CANADA, CHILE, CHINA, COLOMBIA, CROATIA, CZECH REPUBLIC, DENMARK, ESTONIA, FINLAND, FRANCE, GERMANY, GREECE, GUATEMALA, HONG KONG, INDONESIA, ISRAEL, ITALY, JAPAN, JORDAN, KAZAKHSTAN, KENYA, KUWAIT, LATVIA, LEBANON, MADAGASCAR, MALAYSIA, MEXICO, NETHERLANDS, NEW ZEALAND, NORWAY, PAKISTAN, PHILIPPINES, POLAND, PORTUGAL, ROMANIA, SAUDI ARABIA, SERBIA, SINGAPORE, SLOVAKIA, SOUTH AFRICA, SOUTH KOREA, SPAIN, SWEDEN, SWITZERLAND, TAIWAN, THAILAND, TRINIDAD & TOBAGO, TURKEY, UAE, UK, and VIETNAM.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.