CLASS II Device recall · Reported 27 Jan 2021 · FDA Enforcement Report

Abbott i-STAT CG8+ cartridges - IVD Intended recalled

Abbott Point Of Care is recalling Abbott i-STAT CG8+ cartridges - IVD Intended for quantitative measurement of sodium, distributed nationwide in the US. The recall was initiated on 22 Dec 2020 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesStored at room temperature (18 to 30¿C) for periods of time in excess of: 3 days for lot numbers *20100 to *20339 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y Primary GTIN (01) 00054749000163)
Quantity recalled2,820,000 cartridges US and 2,273,453 catridges OUS

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0880-2021
FDA event ID
87073
Recalling firm
Abbott Point Of Care, Princeton, NJ
Classification
Class II
Status
Terminated
Recall initiated
22 Dec 2020
FDA classified
19 Jan 2021
Reported
27 Jan 2021
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Temperature Abuse

Timeline

22 Dec 2020Recall initiated by company
19 Jan 2021Classified by FDA+28 days
27 Jan 2021Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y.

Foods that need refrigeration and were stored or shipped too warm can grow harmful bacteria without any change in look or smell.

Product as listed by the FDA

Abbott i-STAT CG8+ cartridges - IVD Intended for quantitative measurement of sodium, potassium, ionized calcium, glucose, pH, pCO2, pO2 and hematocrit in venous, arterial or capillary whole blood List Number: 03P88-25

Where it was distributed

Nationwide

Nationwide

Questions about this recall

Why was it recalled?

Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y.

Which lots are affected?

Stored at room temperature (18 to 30¿C) for periods of time in excess of: 3 days for lot numbers *20100 to *20339 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y Primary GTIN (01) 00054749000163)

Where was it sold?

Nationwide

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.