CLASS II Device recall · Reported 27 Jan 2021 · FDA Enforcement Report
Abbott i-STAT CG8+ cartridges - IVD Intended recalled
Abbott Point Of Care is recalling Abbott i-STAT CG8+ cartridges - IVD Intended for quantitative measurement of sodium, distributed nationwide in the US. The recall was initiated on 22 Dec 2020 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0880-2021
- FDA event ID
- 87073
- Recalling firm
- Abbott Point Of Care, Princeton, NJ
- Classification
- Class II
- Status
- Terminated
- Recall initiated
- 22 Dec 2020
- FDA classified
- 19 Jan 2021
- Reported
- 27 Jan 2021
- Type
- Voluntary: Firm initiated
- Customers notified by
- Reason category
- Temperature Abuse
- Device type
- Insulin Pumps & Glucose Monitors
Timeline
| 22 Dec 2020 | Recall initiated by company | |
| 19 Jan 2021 | Classified by FDA | +28 days |
| 27 Jan 2021 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y.
Foods that need refrigeration and were stored or shipped too warm can grow harmful bacteria without any change in look or smell.
Product as listed by the FDA
Abbott i-STAT CG8+ cartridges - IVD Intended for quantitative measurement of sodium, potassium, ionized calcium, glucose, pH, pCO2, pO2 and hematocrit in venous, arterial or capillary whole blood List Number: 03P88-25
Where it was distributed
Nationwide
Questions about this recall
Why was it recalled?
Ionized calcium test on some i-STAT CG8+ and EG7+ cartridges may exhibit higher than expected reported results when stored at room temperature (18 to 30C) for periods of time in excess of 3 days for lot numbers *20100 to *20339 and 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y.
Which lots are affected?
Stored at room temperature (18 to 30¿C) for periods of time in excess of: 3 days for lot numbers *20100 to *20339 7 days for lot numbers *20340 and above Where * = lot letter A, K, L, M, N, W or Y Primary GTIN (01) 00054749000163)
Where was it sold?
Nationwide
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
1 other product · FDA event 87073| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| CLASS II | Abbott i-STAT EG7+ cartridge-IVD recalled by Abbott Point Of Care | Abbott Point Of Care | Temperature Abuse | Nationwide |
More recalls from Abbott Point Of Care
All 31| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | i-STAT EG6+ cartridge recalled by Abbott Point Of Care | Abbott Point Of Care | Other | Nationwide |
| Device | CLASS II | i-STAT EG7+ cartridge recalled by Abbott Point Of Care | Abbott Point Of Care | Other | Nationwide |
| Device | CLASS II | i-STAT G3+ cartridge recalled by Abbott Point Of Care | Abbott Point Of Care | Other | Nationwide |
| Device | CLASS II | i-STAT EG7+ cartridge recalled by Abbott Point Of Care | Abbott Point Of Care | Other | Nationwide |
| Device | CLASS II | i-STAT CG8+ cartridge recalled by Abbott Point Of Care | Abbott Point Of Care | Other | Nationwide |
Other Insulin Pumps & Glucose Monitors recalls
All 467| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | MiniMed Flex Insulin Pump recalled by Medtronic MiniMed | Medtronic MiniMed | Labeling Errors | Nationwide |
| Device | CLASS II | Fisher Scientific Sure-Vue Urine hCG Strip recalled by Alere San Diego | Alere San Diego | Potency & Assay | Nationwide |
| Device | ANNOUNCED | StatStrip Glucose Hospital Meter System recalled by Nova BiomedicalNova Biomedical StatStrip | Nova Biomedical | Device Malfunction | |
| Device | CLASS I | StatStrip Glucose Hospital Meter System recalled by Nova Biomedical | Nova Biomedical | Device Malfunction | Nationwide |
| Device | CLASS I | StatStrip Glucose and B- Ketone meter recalled by Nova Biomedical | Nova Biomedical | Device Malfunction | Nationwide |