CLASS II ONGOING Device recall · Reported 12 Aug 2015 · FDA Enforcement Report
Analytical X-ray System recalled by Olympus Scientific Solutions
Olympus Scientific Solutions Americas is recalling Analytical X-ray System, distributed nationwide in the US. The recall was initiated on 9 Jun 2015 and the FDA has classified it as Class II.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-1991-2015
- FDA event ID
- 71649
- Recalling firm
- Olympus Scientific Solutions Americas, Waltham, MA
- Classification
- Class II
- Status
- Ongoing
- Recall initiated
- 9 Jun 2015
- FDA classified
- 31 Jul 2015
- Reported
- 12 Aug 2015
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Foreign Material
- Device type
- Imaging Equipment
Timeline
| 9 Jun 2015 | Recall initiated by company | |
| 31 Jul 2015 | Classified by FDA | +52 days |
| 12 Aug 2015 | Published in enforcement report | +12 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Higher than published scatter around the probe head when testing some samples. At the point the customer would interact directly with the instrument (Trigger position), most readings were lower. However, measurements around the probe head were higher particularly on samples know to be highly scattering (plastics, soils) with minimal increases on high density samples (steel, aluminum).
Foreign material recalls cover pieces of metal, plastic, glass, wood or other objects that got into the product during processing and could cause injury if eaten.
Product as listed by the FDA
Olympus Scientific Solutions Americas Analytical X-ray System. Delta, Delta Element, Delta Plus, DPO-4000-XX-U-EN-EN-PM-SP12, DPO-6000-XX-U-EN-EN-AP-SP10. Mobile X-ray System.
Where it was distributed
Nationwide Distribution.
Questions about this recall
Why was it recalled?
Higher than published scatter around the probe head when testing some samples. At the point the customer would interact directly with the instrument (Trigger position), most readings were lower. However, measurements around the probe head were higher particularly on samples know to be highly scattering (plastics, soils) with minimal increases on high density samples (steel, aluminum).
Which lots are affected?
Delta, Delta Element, Delta Plus, DPO-4000-XX-U-EN-EN-PM-SP12, DPO-6000-XX-U-EN-EN-AP-SP10.
Where was it sold?
Nationwide Distribution.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class II mean?
Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Olympus Scientific Solutions Americas
All 6| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Olympus Scientific Solutions Americas Corporation (Ossa) Model recalled | Olympus Scientific Solutions Americas | Device Malfunction | Nationwide |
| Device | CLASS II | (Ossa) Vanta¿ Xrf Analyzer recalled by Olympus Scientific Solutions | Olympus Scientific Solutions Americas | Other | 22 states |
| Device | CLASS II | Olympus Scientific Solutions Americas Corporation (Ossa) Model recalled | Olympus Scientific Solutions Americas | Other | 16 states |
| Device | CLASS II | Olympus Scientific Solutions Americas Corporation (Ossa) Model recalled | Olympus Scientific Solutions Americas | Device Malfunction | Nationwide |
| Device | CLASS II | ( OSSA ) Delta XRF Analyzer . This is a Analytical X-ray system recalled | Olympus Scientific Solutions Americas | Device Malfunction | Nationwide |
Other Imaging Equipment recalls
All 2,336| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Packaging Defects | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Ultrasound Breast Pack recalled | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Alphenix Models Infx-8000C Infx-8000F Infx-8000V This device recalled | Canon Medical System USA | Other | Nationwide |
| Device | CLASS II | Merge PACS and Merge OrthoPACS recalled by Merge Healthcare | Merge Healthcare | Labeling Errors | Nationwide |
| Device | CLASS II | Philips Azurion recalled by Philips Medical Systems | Philips Medical Systems Nederl | Potency & Assay |