CLASS I ONGOING Device recall · Reported 19 Nov 2025 · FDA Enforcement Report
Anesthesia Circuit Kits and other various kits that contain recalled
Medline Industries is recalling Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits, distributed nationwide in the US. The recall was initiated on 10 Oct 2025 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0444-2026
- FDA event ID
- 97882
- Recalling firm
- Medline Industries, Northfield, IL
- Classification
- Class I
- Status
- Ongoing
- Recall initiated
- 10 Oct 2025
- FDA classified
- 13 Nov 2025
- Reported
- 19 Nov 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Reason category
- Packaging Defects
- Device type
- Heart & Vascular Devices
Timeline
| 10 Oct 2025 | Recall initiated by company | |
| 13 Nov 2025 | Classified by FDA | +34 days |
| 19 Nov 2025 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.
Product as listed by the FDA
Anesthesia Circuit Kits and other various kits that contain Anesthesia Circuits; OPEN HEART SUPPLEMENT DYNJ909116D; VASCULAR DYNJ909124F
Where it was distributed
International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;
Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutFloridaGeorgiaIowaIllinoisKansasKentuckyMassachusettsMichiganMinnesotaMissouriMississippiMontanaNorth Carolina
Questions about this recall
Why was it recalled?
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Which lots are affected?
DYNJ909116D UDI-DI 40198459055356 (case), 10198459055355 (ea) Lots 25FBM604 25GBP604; DYNJ909124F UDI-DI 40198459130619 (case), 10198459130618 (ea) Lots 25FBB831 25GBI833
Where was it sold?
International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
9 other products · FDA event 97882More recalls from Medline Industries
All 707| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Medline Basic Homecare Beds and Medlite Homecare Beds recalled | Medline Industries | Device Malfunction | Nationwide |
| Device | CLASS I | Medline Basic Homecare Beds and Medlite Homecare Beds recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Major Pack recalled by Medline Industries | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Peds Sinus Pack recalled by Medline Industries | Medline Industries | Other | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Mfc Craniotomy Pack recalled | Medline Industries | Other | Nationwide |
Other Heart & Vascular Devices recalls
All 6,217| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medical convenience kits containing BD CliniClean Pericare recalled | American Contract Systems | Device Malfunction | Nationwide |
| Device | CLASS II | Medical convenience kits containing BD CliniClean Pericare recalled | American Contract Systems | Device Malfunction | Nationwide |
| Device | CLASS II | Pipeline Flex Embolization Device with Shield Technology recalled | Micro Therapeutics | Labeling Errors | Nationwide |
| Device | CLASS II | System (RED 62 Reperfusion Catheter) recalled by Penumbra | Penumbra | Device Malfunction | Nationwide |
| Device | CLASS II | Medical convenience kits containing BD CliniClean Pericare recalled | American Contract Systems | Device Malfunction | Nationwide |