CLASS I ONGOING Device recall · Reported 19 Nov 2025 · FDA Enforcement Report

Convenience kits recalled by Medline Industries

Medline Industries is recalling Convenience kits, distributed nationwide in the US. The recall was initiated on 10 Oct 2025 and the FDA has classified it as Class I.

Check your package

Lot, batch or date codesDR SALMA'S SPINE DYNJ909111F UDI-DI 40198459130794 (case), 10198459130793 (ea) Lots 25FBE036 25GBV759; ORTHO SPINE FUMICH DYNJ901047R UDI-DI 40198459130466 (case), 10198459130465 (ea) Lots 25FBR230 25GBU693; PATTERSON TOTAL JOINT DYNJ909117G UDI-DI 40198459130480 (case), 10198459130489 (ea) Lots 25FBO307 25GBN625; TOTAL JOINT DYNJ902308K UDI-DI 40198459130596 (case), 10198459130595 (ea) Lots 25FBB321 25GBU688 KYPHOPLASTY DYNJ909113F UDI-DI 40198459130459 (case), 10198459130458 (ea) Lot 25IBM788
Quantity recalled96,406 total units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0451-2026
FDA event ID
97882
Recalling firm
Medline Industries, Northfield, IL
Classification
Class I
Status
Ongoing
Recall initiated
10 Oct 2025
FDA classified
13 Nov 2025
Reported
19 Nov 2025
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Packaging Defects
Device type
Spinal Implants

Timeline

10 Oct 2025Recall initiated by company
13 Nov 2025Classified by FDA+34 days
19 Nov 2025Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.

Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.

Product as listed by the FDA

Convenience kits; DYNJ901047R ORTHO SPINE FUMICH DYNJ902308K TOTAL JOINT DYNJ909111F DR SALMA'S SPINE DYNJ909117G PATTERSON TOTAL JOINT DYNJ909113F KYPHOPLASTY

Where it was distributed

International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;

Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutFloridaGeorgiaIowaIllinoisKansasKentuckyMassachusettsMichiganMinnesotaMissouriMississippiMontanaNorth Carolina

Questions about this recall

Why was it recalled?

Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.

Which lots are affected?

DR SALMA'S SPINE DYNJ909111F UDI-DI 40198459130794 (case), 10198459130793 (ea) Lots 25FBE036 25GBV759; ORTHO SPINE FUMICH DYNJ901047R UDI-DI 40198459130466 (case), 10198459130465 (ea) Lots 25FBR230 25GBU693; PATTERSON TOTAL JOINT DYNJ909117G UDI-DI 40198459130480 (case), 10198459130489 (ea) Lots 25FBO307 25GBN625; TOTAL JOINT DYNJ902308K UDI-DI 40198459130596 (case), 10198459130595 (ea) Lots 25FBB321 25GBU688 KYPHOPLASTY DYNJ909113F UDI-DI 40198459130459 (case), 10198459130458 (ea) Lot 25IBM788

Where was it sold?

International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class I mean?

Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

9 other products · FDA event 97882

This product is part of a recall event; the main page for the event is Anesthesia Circuit Kits and other various kits that contain recalled.