CLASS I ONGOING Device recall · Reported 19 Nov 2025 · FDA Enforcement Report
Convenience Kits: DYNJ901048N Ortho Spine ST Clair recalled
Medline Industries is recalling Convenience Kits: DYNJ901048N Ortho Spine ST Clair DYNJ909123F Urology Minor DYNJ909132F, distributed nationwide in the US. The recall was initiated on 10 Oct 2025 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0448-2026
- FDA event ID
- 97882
- Recalling firm
- Medline Industries, Northfield, IL
- Classification
- Class I
- Status
- Ongoing
- Recall initiated
- 10 Oct 2025
- FDA classified
- 13 Nov 2025
- Reported
- 19 Nov 2025
- Type
- Voluntary: Firm initiated
- Customers notified by
- Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
- Reason category
- Packaging Defects
- Device type
- Hip & Knee Implants
Timeline
| 10 Oct 2025 | Recall initiated by company | |
| 13 Nov 2025 | Classified by FDA | +34 days |
| 19 Nov 2025 | Published in enforcement report | +6 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Packaging recalls include wrong labels, wrong products in the wrong bottle, missing leaflets, short-filled or empty capsules and seals that fail.
Product as listed by the FDA
Convenience Kits: DYNJ901048N ORTHO SPINE ST CLAIR DYNJ909123F UROLOGY MINOR DYNJ909132F SET UP DYNJ909308D HIP PINNING
Where it was distributed
International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;
Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutFloridaGeorgiaIowaIllinoisKansasKentuckyMassachusettsMichiganMinnesotaMissouriMississippiMontanaNorth Carolina
Questions about this recall
Why was it recalled?
Firm received multiple complaints of anesthesia circuit leaks due to cracks in the 120" expandable tubing. Cracks and resulting leaks may lead to hypoxia, hypoventilation, and inadequate delivery of inhaled anesthetic agents, which may result in serious injury or death. Volatile anesthetics may also escape into the operating room environment.
Which lots are affected?
HIP PINNING DYNJ909308D UDI-DI 40198459055288 (case), 10198459055287 (ea) Lots 25FBB311 25GBC282 25GBW486; ORTHO SPINE ST CLAIR DYNJ901048N UDI-DI 40198459055554 (case), 10198459055553 (ea) Lots 25GBQ070; SET UP DYNJ909132F UDI-DI 40198459130572 (case), 10198459130571 (ea) Lots 25GBI845 25IBF708; UROLOGY MINOR DYNJ909123F UDI-DI 40198459130602 (case), 10198459130601 (ea) Lots 25GBT826 25IBF707
Where was it sold?
International distribution to Mexico and Panama. US nationwide distribution to the following: AK AL AR AZ CA CO CT FL GA IA IL IN KS KY MA ME MI MN MO MS MT NC NJ NY OH OR PA PR RI SC TN TX WA WI;
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
Other products in this recall event
9 other products · FDA event 97882This product is part of a recall event; the main page for the event is Anesthesia Circuit Kits and other various kits that contain recalled.
More recalls from Medline Industries
All 686| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline Convenience Kits: Blood Draw Kit recalled by Medline Industries | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep Clear recalled | Medline Industries | Packaging Defects | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Ortho Aspiration Kit recalled | Medline Industries | Sterility | Nationwide |
| Device | CLASS II | Medline Convenience Kits containing BD ChloraPrep DYKA1054F recalled | Medline Industries | Packaging Defects | Nationwide |
| Device | CLASS II | Medline Convenience Kits: Arterial Line Insertion Kit recalled | Medline Industries | Sterility | Nationwide |
Other Hip & Knee Implants recalls
All 3,097| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Medline convenience kits labeled as: Femoral Pack, Medline Kit recalled | Medline Industries | Other | Nationwide |
| Device | CLASS II | Arvis Surgical Navigation System Gen recalled by Kico Knee Innovation | Kico Knee Innovation | Device Malfunction | 19 states |
| Device | CLASS II | Veritas Standard Reverse Total Shoulder Veritas Standard recalled | restor3d | Device Malfunction | Nationwide |
| Device | CLASS II | Endo-Model, Replacement Set, Tibial Plateau, right/left, L, Uhmwpe recalled | Waldemar Link GmbH & Co. KG (Mfg Site) | Other | IN |
| Device | CLASS II | U2 Total Knee System Femoral offset bushing recalled | United Orthopedic | Device Malfunction | Nationwide |