CLASS II Device recall · Reported 4 Jul 2012 · FDA Enforcement Report

Architect CA 19-9XR Reagent Kit recalled by Abbott Laboratories

Abbott Laboratories is recalling Architect CA 19-9XR Reagent Kit, distributed nationwide in the US. The recall was initiated on 30 May 2012 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codeslot 08849M500 exp. 10/13/2012; lot 08851M500 exp. 10/13/2012; lot 08852M500 exp. 11/30/2012; lot 08853M500 exp. 11/30/2012; lot 10040M500 exp. 11/30/2012; and lot 10122M500 exp. 11/30/2012
Quantity recalled316 kits - USA, 8,356 kits ROW

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1904-2012
FDA event ID
61979
Recalling firm
Abbott Laboratories, Abbott Park, IL
Classification
Class II
Status
Terminated
Recall initiated
30 May 2012
FDA classified
28 Jun 2012
Reported
4 Jul 2012
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Potency & Assay

Timeline

30 May 2012Recall initiated by company
28 Jun 2012Classified by FDA+29 days
4 Jul 2012Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

The six ARCHITECT CA 19-9XR reagent lots are demonstrating a shift up in patient results. The Abbott Controls however do not detect the upward shift and will test within range.

Potency and specification recalls happen when a batch tests outside its approved limits for strength, purity or dissolution, which can mean the dose delivered is higher or lower than labelled.

Product as listed by the FDA

ARCHITECT CA 19-9XR Reagent Kit; ARCHITECT CA 19-9 Reagent Kit, list 291-20 (4 x 100 tests), 2K91-25 and 2K91-27 (1 x 100 tests); Product Usage: The ARCHITECT CA 19-9XR assay is a chemiluminescent microparticle immunoassay (CMIA) for the quantitative determination of 1116-NS-19-9 reactive determinants in human serum or plasma on the ARCHITECT i System. The ARCHITECT CA-19-9XR assay is to be used as an aid in the management of pancreatic cancer patients in conjunction with other clinical methods.

Where it was distributed

Worldwide Distribution - US (nationwide) the states of: Alabama, Arkansas, Arizona, California, Florida, Iowa, Indiana, Louisiana, Michigan, Minnesota, New Jersey, Nevada, New York, Ohio, Oklahoma, Pennsylvania, South Dakota, Texas, Virginia, Washington, including Puerto Rico, and the countries of Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Barbados, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Ecuador, Egypt, Estonia, Finland, France, Gaza & Jericho, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kyrgyzstan, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Mexico, Montenegro, Morocco, Netherlands, New Zealand, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Republic of Yemen, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Togo, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela and Vietnam.

Nationwide AlabamaArkansasArizonaCaliforniaFloridaIowaIndianaLouisianaMichiganMinnesotaNew JerseyNevadaNew YorkOhioOklahomaPennsylvaniaPuerto RicoSouth DakotaTexasVirginia

Questions about this recall

Is Architect CA 19-9XR Reagent Kit recalled?

Yes. Abbott Laboratories recalled Architect CA 19-9XR Reagent Kit on 30 May 2012. The recall is Class II and its status is Terminated. Recall number Z-1904-2012.

Why was it recalled?

The six ARCHITECT CA 19-9XR reagent lots are demonstrating a shift up in patient results. The Abbott Controls however do not detect the upward shift and will test within range.

Which lots are affected?

lot 08849M500 exp. 10/13/2012; lot 08851M500 exp. 10/13/2012; lot 08852M500 exp. 11/30/2012; lot 08853M500 exp. 11/30/2012; lot 10040M500 exp. 11/30/2012; and lot 10122M500 exp. 11/30/2012

Where was it sold?

Worldwide Distribution - US (nationwide) the states of: Alabama, Arkansas, Arizona, California, Florida, Iowa, Indiana, Louisiana, Michigan, Minnesota, New Jersey, Nevada, New York, Ohio, Oklahoma, Pennsylvania, South Dakota, Texas, Virginia, Washington, including Puerto Rico, and the countries of Algeria, Argentina, Australia, Austria, Azerbaijan, Bahrain, Barbados, Belarus, Belgium, Bosnia and Herzegovina, Brazil, Bulgaria, Canada, Chile, China, Colombia, Costa Rica, Croatia, Czech Republic, Ecuador, Egypt, Estonia, Finland, France, Gaza & Jericho, Germany, Greece, Guatemala, Hong Kong, Hungary, India, Indonesia, Iran, Ireland, Israel, Italy, Japan, Jordan, Kazakhstan, Kenya, Kyrgyzstan, Latvia, Lebanon, Lithuania, Luxembourg, Malaysia, Mexico, Montenegro, Morocco, Netherlands, New Zealand, Oman, Pakistan, Panama, Peru, Philippines, Poland, Portugal, Republic of Yemen, Romania, Russia, Saudi Arabia, Senegal, Serbia, Singapore, Slovakia, Slovenia, South Africa, South Korea, Spain, Sweden, Switzerland, Thailand, Togo, Tunisia, Turkey, United Arab Emirates, United Kingdom, Uruguay, Venezuela and Vietnam.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 21 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.