CLASS I Device recall · Reported 29 Nov 2017 · FDA Enforcement Report
Arkon Anesthesia Delivery System recalled by Spacelabs Healthcare
Spacelabs Healthcare is recalling Arkon Anesthesia Delivery System, distributed nationwide in the US. The recall was initiated on 11 Oct 2017 and the FDA has classified it as Class I.
Check your package
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
Recall details
- Recall number
- Z-0072-2018
- FDA event ID
- 78330
- Recalling firm
- Spacelabs Healthcare, Hertford, N/A
- Classification
- Class I
- Status
- Terminated
- Recall initiated
- 11 Oct 2017
- FDA classified
- 21 Nov 2017
- Reported
- 29 Nov 2017
- Type
- Voluntary: Firm initiated
- Customers notified by
- Letter
- Reason category
- Device Malfunction
- Device type
- CPAP & Ventilators
Timeline
| 11 Oct 2017 | Recall initiated by company | |
| 21 Nov 2017 | Classified by FDA | +41 days |
| 29 Nov 2017 | Published in enforcement report | +8 days |
Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.
Reason for recall
Arkon Anesthesia Workstation, with software version 2.61, experienced failure in mechanical ventilation, oxygen and anesthetic gas delivery, with concurrent failure of the display unit that resulted in a blank screen without audible or visible alarms.
Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.
Product as listed by the FDA
Arkon Anesthesia Delivery System, Model 99999, and software version 2.61 and upgrade kit 050-9043-00 Rev G. The Spacelabs Arkon Anesthesia Workstation is intended for use in the hospital environment and operating room. It may be used for the delivery of oxygen, air and nitrous oxide in a controlled manner to various patient breathing circuits with or without the use of mechanical ventilator, and may be used for the delivery of anesthetic vapor by use of a dismountable vaporizer. The device is intended for use only by a suitably qualified physician
Where it was distributed
Nationwide Distribution to the states of NM, WY, FL, NC, MS, CO, CT, ME, and AL.
Nationwide AlabamaColoradoConnecticutFloridaMaineMississippiNorth CarolinaNew MexicoWyoming
Questions about this recall
Why was it recalled?
Arkon Anesthesia Workstation, with software version 2.61, experienced failure in mechanical ventilation, oxygen and anesthetic gas delivery, with concurrent failure of the display unit that resulted in a blank screen without audible or visible alarms.
Which lots are affected?
UDI: 10841522100017
Where was it sold?
Nationwide Distribution to the states of NM, WY, FL, NC, MS, CO, CT, ME, and AL.
What should I do if I have it?
Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.
What does Class I mean?
Class I is the most serious FDA recall category: there is a reasonable probability that using the product will cause serious health consequences or death.
Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.
More recalls from Spacelabs Healthcare
All 47| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS II | Xhibit Central Station recalled by Spacelabs Healthcare | Spacelabs Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Spacelabs Ultraview SL Command Module SL Command Module recalled | Spacelabs Healthcare | Packaging Defects | Nationwide |
| Device | CLASS II | Multi-parameter Command Module recalled by Spacelabs Healthcare | Spacelabs Healthcare | Labeling Errors | Nationwide |
| Device | CLASS II | Sentinel V11 recalled by Spacelabs Healthcare | Spacelabs Healthcare | Device Malfunction | Nationwide |
| Device | CLASS II | Branding is not applied to this product DVI display cables recalled | Spacelabs Healthcare | Device Malfunction | Nationwide |
Other CPAP & Ventilators recalls
All 554| Reported | Class | Recall | Company | Reason | Where |
|---|---|---|---|---|---|
| Device | CLASS I | Inca (Infant Nasal Cannulae Assy) Infant Nasal CPAP, Complete Set recalled | CooperSurgical | Other | Nationwide |
| Device | CLASS I | Inca (Infant Nasal Cannulae Assy) Neonatal Nasal CPAP recalled | CooperSurgical | Other | Nationwide |
| Device | CLASS II | Medline Hudson Rci, Dbd-circuit,Vent,Inspiratory,Heated,Neo recalled | Medline Industries | Device Malfunction | Nationwide |
| Device | CLASS II | Getinge Y Sensor Module, Servo-u, Servo recalled | Maquet Critical Care AB | Device Malfunction | Nationwide |
| Device | CLASS II | Biofire Respiratory Panel 2.1 (RP2.1) recalled by BioFire Diagnostics | BioFire Diagnostics | Other | Nationwide |