CLASS II ONGOING Device recall · Reported 5 Feb 2025 · FDA Enforcement Report

Artis One Angiographic X-Ray System recalled

Siemens Medical Solutions USA is recalling Artis One Angiographic X-Ray System. The recall was initiated on 2 Jan 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesModel Number 10848600
Quantity recalled1140

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0922-2025
FDA event ID
96088
Recalling firm
Siemens Medical Solutions USA, Malvern
Classification
Class II
Status
Ongoing
Recall initiated
2 Jan 2025
FDA classified
28 Jan 2025
Reported
5 Feb 2025
Customers notified by
Letter
Reason category
Other Reasons

Timeline

2 Jan 2025Recall initiated by company
28 Jan 2025Classified by FDA+26 days
5 Feb 2025Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

A potential issue with ARTIS One systems was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose-rate may be higher than intended by the user.

Product as listed by the FDA

ARTIS One Angiographic X-Ray System

Where it was distributed

23 Distributed in the US

Questions about this recall

Why was it recalled?

A potential issue with ARTIS One systems was identified. In very rare situations, the first x-ray release following patient registration may be performed with incorrect copper prefiltration. If the described issue occurs, the applied radiation dose-rate may be higher than intended by the user.

Which lots are affected?

Model Number 10848600

Where was it sold?

23 Distributed in the US

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.