CLASS II Device recall · Reported 1 Aug 2018 · FDA Enforcement Report

Artis Pheno, Fluoroscopic X-Ray system recalled

Siemens Medical Solutions USA is recalling Artis Pheno, Fluoroscopic X-Ray system, distributed in 18 states. The recall was initiated on 30 May 2018 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes164033, 164071, 164034, 164095, 164057, 164054, 164055, 164080, 164072, 164088, 164048, 164094, 164068, 164019, 164013, 164073, 164025, 164026, 164027, 164060, 164083, 164021, 164066, 164075, 164074, 164084, 164086, 164098, 164028, 164053, 164063, 164064, 164024, 164049, 164018, 164022, 164061, 164100, 164097, 164011
Quantity recalled40 in the United States

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2524-2018
FDA event ID
80417
Recalling firm
Siemens Medical Solutions USA, Malvern, PA
Classification
Class II
Status
Terminated
Recall initiated
30 May 2018
FDA classified
20 Jul 2018
Reported
1 Aug 2018
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

30 May 2018Recall initiated by company
20 Jul 2018Classified by FDA+51 days
1 Aug 2018Published in enforcement report+12 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

During manual LAO/RAO rotation at maximum speed or a combination of LAO/RAO and Cran/Caud rotation at maximum speed, the possibility exists for the safety monitoring within the system control to become activated. As a result, the C-arm stops the rotation and can drop up to 40 mm (1.6 ).

Product as listed by the FDA

ARTIS Pheno, Fluoroscopic X-Ray system, Model Number 10849000

Where it was distributed

The devices were distributed to the following US states: AL, CA, CO, FL, IL, KY, MA, MI, MN, MS, ME, NJ, NY, OH, TN, TX, UT, and VA.

AlabamaCaliforniaColoradoFloridaIllinoisKentuckyMassachusettsMaineMichiganMinnesotaMississippiNew JerseyNew YorkOhioTennesseeTexasUtahVirginia

Questions about this recall

Why was it recalled?

During manual LAO/RAO rotation at maximum speed or a combination of LAO/RAO and Cran/Caud rotation at maximum speed, the possibility exists for the safety monitoring within the system control to become activated. As a result, the C-arm stops the rotation and can drop up to 40 mm (1.6 ).

Which lots are affected?

164033, 164071, 164034, 164095, 164057, 164054, 164055, 164080, 164072, 164088, 164048, 164094, 164068, 164019, 164013, 164073, 164025, 164026, 164027, 164060, 164083, 164021, 164066, 164075, 164074, 164084, 164086, 164098, 164028, 164053, 164063, 164064, 164024, 164049, 164018, 164022, 164061, 164100, 164097, 164011

Where was it sold?

The devices were distributed to the following US states: AL, CA, CO, FL, IL, KY, MA, MI, MN, MS, ME, NJ, NY, OH, TN, TX, UT, and VA.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Siemens Medical Solutions USA

All 796

Other Imaging Equipment recalls

All 2,336