CLASS II Device recall · Reported 21 Jan 2015 · FDA Enforcement Report

The Attune Intuition Impaction Handle is a re-useable recalled

DePuy Orthopaedics is recalling The Attune Intuition Impaction Handle is a re-useable instrument utilized in knee, distributed nationwide in the US. The recall was initiated on 10 Nov 2014 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCATALOG NO. : 2544-01-017 LOT NUMBER: All Lots Barcode GTIN: 10603295423324
Quantity recalled7,944

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0953-2015
FDA event ID
70014
Recalling firm
DePuy Orthopaedics, Warsaw, IN
Classification
Class II
Status
Terminated
Recall initiated
10 Nov 2014
FDA classified
12 Jan 2015
Reported
21 Jan 2015
Type
Voluntary: Firm initiated
Customers notified by
Two or more of the following: Email, Fax, Letter, Press Release, Telephone, Visit
Reason category
Other Reasons

Timeline

10 Nov 2014Recall initiated by company
12 Jan 2015Classified by FDA+63 days
21 Jan 2015Published in enforcement report+9 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Complaints have indicated that the ATTUNE INTUITION Impaction Handle lever has fractured during impaction. If the lever fractures, there is the potential for fractured pieces of the device, or the inner spring, if released, to be left in the patient if not observed during surgery.

Product as listed by the FDA

The ATTUNE INTUITION Impaction Handle is a re-useable instrument utilized in knee replacement procedures. The instrument interfaces with several related ATTUNE instruments: keel punch, tibial tower, evaluation bullet, fixed bearing tibial impactor, rotating platform tibial impactor, and femoral impactor, for the purpose of impaction, extraction, and insertion applications at steps within the surgical procedure. Due to the frequency of its use, two impaction handles are supplied in each kit.

Where it was distributed

Worldwide Distribution-US (nationwide) except AK and RI, including PR and the countries of Chile, Canada, Australia, Austria, Belgium, France, Hong Kong, Ireland, India, Israel, Italy, Japan, Malaysia, Netherlands, New Zealand, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Switzerland, Thailand, United Arab Emirates (UAE), and United Kingdom (UK

Nationwide AlaskaPuerto RicoRhode Island

Questions about this recall

Why was it recalled?

Complaints have indicated that the ATTUNE INTUITION Impaction Handle lever has fractured during impaction. If the lever fractures, there is the potential for fractured pieces of the device, or the inner spring, if released, to be left in the patient if not observed during surgery.

Which lots are affected?

CATALOG NO. : 2544-01-017 LOT NUMBER: All Lots Barcode GTIN: 10603295423324

Where was it sold?

Worldwide Distribution-US (nationwide) except AK and RI, including PR and the countries of Chile, Canada, Australia, Austria, Belgium, France, Hong Kong, Ireland, India, Israel, Italy, Japan, Malaysia, Netherlands, New Zealand, Saudi Arabia, Singapore, Slovakia, South Africa, South Korea, Spain, Switzerland, Thailand, United Arab Emirates (UAE), and United Kingdom (UK

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.