CLASS II ONGOING Device recall · Reported 12 Mar 2025 · FDA Enforcement Report

Aurora Surgiscope System, Sterile recalled by Integra LifeSciences

Integra LifeSciences is recalling Aurora Surgiscope System, Sterile, single use device that contains a Sheath, Obturator, distributed nationwide in the US. The recall was initiated on 6 Feb 2025 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesCatalog Number: (1) ASX15/60 (2) ASX15/80; UDI-DI: (1) 00850002332254, (2) 008500023332247; Lot No. (1) All unexpired lots, (2) All unexpired lots.
Quantity recalled623 units

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-1295-2025
FDA event ID
96216
Recalling firm
Integra LifeSciences, Princeton, NJ
Classification
Class II
Status
Ongoing
Recall initiated
6 Feb 2025
FDA classified
6 Mar 2025
Reported
12 Mar 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

6 Feb 2025Recall initiated by company
6 Mar 2025Classified by FDA+28 days
12 Mar 2025Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Possibility for the obturator to break (separate).

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

AURORA Surgiscope System, Sterile, single use device that contains a Sheath, Obturator, and Imager. Manufacturer's Product Numbers (Catalog Number): (1) ASX15/60 and (2) ASX15/80

Where it was distributed

US Nationwide distribution in the states of MO, FL, TX, NY, IL, OH, NJ, CA, MN, WA, TN, MD, LA, OK, SC, PA, MI, KY, NC, DC.

Nationwide CaliforniaDistrict of ColumbiaFloridaIllinoisKentuckyLouisianaMarylandMichiganMinnesotaMissouriNorth CarolinaNew JerseyNew YorkOhioOklahomaPennsylvaniaSouth CarolinaTennesseeTexasWashington

Questions about this recall

Why was it recalled?

Possibility for the obturator to break (separate).

Which lots are affected?

Catalog Number: (1) ASX15/60 (2) ASX15/80; UDI-DI: (1) 00850002332254, (2) 008500023332247; Lot No. (1) All unexpired lots, (2) All unexpired lots.

Where was it sold?

US Nationwide distribution in the states of MO, FL, TX, NY, IL, OH, NJ, CA, MN, WA, TN, MD, LA, OK, SC, PA, MI, KY, NC, DC.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from Integra LifeSciences

All 246

Other Other Medical Devices recalls

All 13,908