CLASS II ONGOING Device recall · Reported 22 Jan 2025 · FDA Enforcement Report

BD Pyxis MedStation ES (Med ES Main) recalled by CareFusion 303

CareFusion 303 is recalling BD Pyxis MedStation ES (Med ES Main), distributed nationwide in the US. The recall was initiated on 5 Apr 2023 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesSoftware v1.7.3. REF/UDI-DI: 323/10885403512667, 352/10885403512674
Quantity recalled622

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-0906-2025
FDA event ID
96002
Recalling firm
CareFusion 303, San Diego, CA
Classification
Class II
Status
Ongoing
Recall initiated
5 Apr 2023
FDA classified
14 Jan 2025
Reported
22 Jan 2025
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Device Malfunction

Timeline

5 Apr 2023Recall initiated by company
14 Jan 2025Classified by FDA+650 days
22 Jan 2025Published in enforcement report+8 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

Due to an automated dispensing cabinet software issue, when a medication single order contains more than one component and at least one is unavailable for a multi-item or combination order, the order may display as grayed out; alternate location tooltip will display "Server connection failed," not listing other area devices where unavailable item(s) may be found, which may lead to delayed therapy.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

BD Pyxis MedStation ES (Med ES Main), REF: 323; and BD Pyxis MedStation ES (Med ES Main Tower), REF: 352

Where it was distributed

US Nationwide distribution in the states of NY, NC, LA.

Nationwide LouisianaNorth CarolinaNew York

Questions about this recall

Why was it recalled?

Due to an automated dispensing cabinet software issue, when a medication single order contains more than one component and at least one is unavailable for a multi-item or combination order, the order may display as grayed out; alternate location tooltip will display "Server connection failed," not listing other area devices where unavailable item(s) may be found, which may lead to delayed therapy.

Which lots are affected?

Software v1.7.3. REF/UDI-DI: 323/10885403512667, 352/10885403512674

Where was it sold?

US Nationwide distribution in the states of NY, NC, LA.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.