CLASS II ONGOING Device recall · Reported 7 Aug 2024 · FDA Enforcement Report

Biofire Blood Culture Identification 2 (BCID2) Panel recalled

BioFire Diagnostics is recalling Biofire Blood Culture Identification 2 (BCID2) Panel, distributed nationwide in the US. The recall was initiated on 19 Jun 2024 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codesBCID2 Panel REF: RFIT-ASY-0147, UDI-DI: 00815381020338, All unexpired lots. BACT/ALERT REF/UDI-DI: Lot/Expiration: 410851/03573026596057: 0004101638/3/20/2024, 0004101930/7/28/2024, 0004102026/9/4/2024, 0004102408/2/16/2025, 0004102996/9/24/2025, 0004102946/8/29/2025, 0004102956/9/17/2025, 0004102998/9/24/2025. 410853/03573026596095: 0004101718/4/24/2024, 0004101790/6/1/2024, 0004101958/8/3/2024, 0004102452/3/3/2025, 0004102964/9/19/2025
Quantity recalled99,822

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2456-2024
FDA event ID
94886
Recalling firm
BioFire Diagnostics, Salt Lake City, UT
Classification
Class II
Status
Ongoing
Recall initiated
19 Jun 2024
FDA classified
31 Jul 2024
Reported
7 Aug 2024
Type
Voluntary: Firm initiated
Customers notified by
Letter
Reason category
Other Reasons

Timeline

19 Jun 2024Recall initiated by company
31 Jul 2024Classified by FDA+42 days
7 Aug 2024Published in enforcement report+7 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

If blood culture identification panel is used in conjunction with specific lots of culture media bottles, then false positive Serratia marcescens results may occur, due to an increased level of non-viable organism from serratia marcescens targets in culture media bottles, false positive result may lead to an inappropriate change in patient therapy.

Product as listed by the FDA

BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, for FILMARRAY Systems, when used in conjunction with Culture Media Bottles: BACT/ALERT FA PLUS, REF: 410851, and BACT/ALERT PF PLUS, REF: 410853

Where it was distributed

Worldwide - US Nationwide distribution including in the states of FL, IL, KY, CO, KS, NC, AK, NY, ND, CA, TX, MD, GA, OH, WI, SD, LA, MT, VA, IA, MO, MA, WV, NE, OK, AZ, OR, SC, MI, IN, ID, TN, WA, NM, NJ, AR, AL, MN, NH, NV, PA, WY, MS, CT and the countries of AE, AM, AR, AT, BA, BG, BH, BN, BR, CA, CH, CI, CK, CN, CO, CZ, DE, DZ, EC, EE, EG, ES, FI, FR, GA, GB, GE, GR, GT, HR, HU, IL, IN, IQ, IS, IT, JO, JP, KK, KR, KW, LB, LT, LV, MK, MM, MX, MY, NG, NO, NW, OM, PA, PH, PK, PL, PT, QA, RO, RS, SA, SI, TN, TR, TW, UG, UY, VN, WF.

Nationwide AlaskaAlabamaArkansasArizonaCaliforniaColoradoConnecticutDelawareFloridaGeorgiaIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMarylandMichigan

Questions about this recall

Why was it recalled?

If blood culture identification panel is used in conjunction with specific lots of culture media bottles, then false positive Serratia marcescens results may occur, due to an increased level of non-viable organism from serratia marcescens targets in culture media bottles, false positive result may lead to an inappropriate change in patient therapy.

Which lots are affected?

BCID2 Panel REF: RFIT-ASY-0147, UDI-DI: 00815381020338, All unexpired lots. BACT/ALERT REF/UDI-DI: Lot/Expiration: 410851/03573026596057: 0004101638/3/20/2024, 0004101930/7/28/2024, 0004102026/9/4/2024, 0004102408/2/16/2025, 0004102996/9/24/2025, 0004102946/8/29/2025, 0004102956/9/17/2025, 0004102998/9/24/2025. 410853/03573026596095: 0004101718/4/24/2024, 0004101790/6/1/2024, 0004101958/8/3/2024, 0004102452/3/3/2025, 0004102964/9/19/2025

Where was it sold?

Worldwide - US Nationwide distribution including in the states of FL, IL, KY, CO, KS, NC, AK, NY, ND, CA, TX, MD, GA, OH, WI, SD, LA, MT, VA, IA, MO, MA, WV, NE, OK, AZ, OR, SC, MI, IN, ID, TN, WA, NM, NJ, AR, AL, MN, NH, NV, PA, WY, MS, CT and the countries of AE, AM, AR, AT, BA, BG, BH, BN, BR, CA, CH, CI, CK, CN, CO, CZ, DE, DZ, EC, EE, EG, ES, FI, FR, GA, GB, GE, GR, GT, HR, HU, IL, IN, IQ, IS, IT, JO, JP, KK, KR, KW, LB, LT, LV, MK, MM, MX, MY, NG, NO, NW, OM, PA, PH, PK, PL, PT, QA, RO, RS, SA, SI, TN, TR, TW, UG, UY, VN, WF.

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

More recalls from BioFire Diagnostics

All 47

Other Other Medical Devices recalls

All 13,929