CLASS II Device recall · Reported 31 Aug 2016 · FDA Enforcement Report

Brilliance CT Big Bore Oncology recalled by Philips Medical Systems

Philips Medical Systems (Cleveland) is recalling Brilliance CT Big Bore Oncology, distributed nationwide in the US. The recall was initiated on 20 Jul 2016 and the FDA has classified it as Class II.

Check your package

Lot, batch or date codes728243  Brilliance BigBore Oncology CT 7116 7611 7354 7063 7309 7169 7205 7016 7019 7043 7069 7073 7091 7112 7132 7163 7141 7194 7226 7325 7294 7350 7316 7480 7114 7210 7310 7438 7530 7186 7119 7042
Quantity recalled33

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

Recall details

Recall number
Z-2653-2016
FDA event ID
74820
Recalling firm
Philips Medical Systems (Cleveland), Cleveland, OH
Classification
Class II
Status
Terminated
Recall initiated
20 Jul 2016
FDA classified
25 Aug 2016
Reported
31 Aug 2016
Type
Voluntary: Firm initiated
Customers notified by
E-Mail
Reason category
Device Malfunction

Timeline

20 Jul 2016Recall initiated by company
25 Aug 2016Classified by FDA+36 days
31 Aug 2016Published in enforcement report+6 days

Companies usually start a recall weeks before the FDA classifies and publishes it; the gap above is that lag.

Reason for recall

After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.

Device recalls are issued when a component, software or design fault could cause the device to fail or give wrong readings. Many are corrections applied in the field rather than physical returns.

Product as listed by the FDA

Brilliance CT Big Bore Oncology, Model number 728243 Product Usage: Computed Tomography X-ray systems intended to produce cross-sectional images of the body by computer reconstruction of X-ray transmission data taken at different angles and planes.

Where it was distributed

US Nationwide Distribution in the states of: AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA, MD,ME,MI,MN,MO,MS,NC,ND,NE,NJ,NM, NV,NY,OH,OK,OR,PA,PR,SC,TN,TX,UT,VA,VT,WA,WI,WV,WY

Nationwide AlabamaArkansasArizonaCaliforniaColoradoConnecticutDistrict of ColumbiaDelawareFloridaGeorgiaHawaiiIowaIdahoIllinoisIndianaKansasKentuckyLouisianaMassachusettsMaryland

Questions about this recall

Why was it recalled?

After upgrading to 3.6.7 software version via FCO72800643, during reconstruction of gated helical scans, the planned anatomy can be cut off on the CT images. This issue can occur on gated CT helical reconstruction that is not planned at iso-center (0,0). If the region of interest is not visualized in the images, a CT rescan may be performed.

Which lots are affected?

728243  Brilliance BigBore Oncology CT 7116 7611 7354 7063 7309 7169 7205 7016 7019 7043 7069 7073 7091 7112 7132 7163 7141 7194 7226 7325 7294 7350 7316 7480 7114 7210 7310 7438 7530 7186 7119 7042

Where was it sold?

US Nationwide Distribution in the states of: AL,AR,AZ,CA,CO,CT,DC,DE,FL,GA,HI,IA,ID,IL,IN,KS,KY,LA,MA, MD,ME,MI,MN,MO,MS,NC,ND,NE,NJ,NM, NV,NY,OH,OK,OR,PA,PR,SC,TN,TX,UT,VA,VT,WA,WI,WV,WY

What should I do if I have it?

Stop using the device if the notice says to, and contact the manufacturer or your healthcare provider for the correction or replacement.

What does Class II mean?

Class II means the product may cause temporary or medically reversible health problems, or the probability of serious harm is remote.

Read the official FDA notice · Source record retrieved 26 Sep 2026. The FDA describes openFDA data as unvalidated; the official notice takes precedence. This page reports the record and gives no medical advice.

Other products in this recall event

6 other products · FDA event 74820

This product is part of a recall event; the main page for the event is Brilliance CT 16 Power recalled by Philips Medical Systems.

More recalls from Philips Medical Systems (Cleveland)

All 315

Other Imaging Equipment recalls

All 2,336